Midmark IQvitals Zone, automatically and noninvasively… recallZ-3219-2026 · Midmark Corporation · initiated Aug 21, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Midmark Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
A performance issue may result in inaccurate blood pressure readings when the Linear Deflation Method is used.
FDA root cause category: Process change control
What the company is doing
An Urgent Medical Device Recall notification letter dated 8/19/26 was sent to customers. Safety Instructions Until Midmark provides further instructions regarding this recall, use only the Step Deflation Method when obtaining blood pressure measurements. Do not use the Linear Deflation Method. Follow the instructions in Appendix A to verify that Step Deflation is selected before obtaining blood pressure measurements. Please complete, sign and return the Customer Response Form after completing the instructions described in Appendix A of this notification. The affected device(s) may continue to be used for blood pressure measurements provided this safety instruction is followed. Ensure all users of the affected device(s) are informed of this safety instruction. Product Correction Midmark will contact your facility with the next steps required to complete this recall. Additional instructions will be provided at that time. No immediate action is required to contact Midmark at this time. Contact Information Please be assured that maintaining a high level of safety and quality is our highest priority. If you have any questions, please contact the Technical Service Team at 1.844.856.1230, option 2, or email [email protected]. Required Distributor Actions 1. Immediately send the attached customer-facing "Urgent Medical Device Recall - IQvitals Zone Linear Deflation Blood Pressure Performance" notification to every impacted end customer to whom you supplied an affected device. Do not wait for Midmark to contact the customer before forwarding the attached safety notification. 2. If you supplied an affected device to another distributor, send that distributor both this first-consignee communication and the attached customer notification. Instruct the downstream distributor to immediately forward only the customer notification to each impacted end customer and provide the final consignee information to you for submission to Midmark. 3. Immediately provide Midmark wi
Recall details
- Recall number
- Z-3219-2026
- Status
- Open, Classified
- Initiated by the company
- Aug 21, 2026
- Posted by the FDA
- Sep 18, 2026
- Recalling firm
- Midmark Corporation
- Quantity
- 20 units
- Product code
- MWI
- FDA submissions
- K161909
- Distribution
- US Nationwide distribution.
Product description
Midmark IQvitals Zone, automatically and noninvasively measures systolic and diastolic blood pressure, pulse rate, temperature (temporal), and oxygen saturation (SpO2) for adult and pediatric patients, Model Number: 1-200-0350 with Kit Number: 4-000-0722
Code information (models, lots, serial numbers)
Model No: 1-200-0350; UDI-DI: 00841709105821; Kit No: 4-000-0722; UDI-DI: 00841709105906; Kit Serial Number: V2866143 V2866111 V2866124 V2864892 V2864113 V2864444 V2864120 V2864449 V2864894 V2866626 V2866636 V2871280 V2870561 V2869170 V2870570 V2869930 V2874364 V2897744 V2900961 V2900947; Zone Serial Number: V2864929 V2865546 V2864921 V2864495 V2796056 V2864174 V2858651 V2864142 V2873819 V2866216 V2866187 V2870249 V2870090 V2867500 V2866707 V2869394 V2873111 V2886635 V2873150 V2886126
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Midmark IQvitals Zone, w/Nellcor SPO2 and WirlessMidmark Corporation · SYSTEM, MEASUREMENT, BLOOD-PRESSURE, NON-INVASIVE1 recall0 reports
- Midmark IQvitals Zone, Wireless with Nellcor SPO2Midmark Corporation · SYSTEM, MEASUREMENT, BLOOD-PRESSURE, NON-INVASIVE1 recall0 reports
Other devices under the same FDA submission
6 devices cleared under K161909 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Midmark IQvitals Zone w/Masimo SPO2Midmark Corporation · MONITOR,PHYSIOLOGICAL,PATIENT(WITHOUT ARRHYTHMIA DETECTION OR ALARMS)1 recall1 report
- Midmark IQvitals Zone, w/Nellcor SPO2Midmark Corporation · Monitor, physiological, patient (without arrhythmia detection or alarms)1 recall0 reports
- Midmark IQvitals Zone, w/Masimo SPO2 and WirelessMidmark Corporation · Oximeter1 recall0 reports
- Midmark IQvitals Zone, Non-Wireless with Nellcor SPO2Midmark Corporation · Oximeter1 recall0 reports
- Midmark IQvitals Zone, Non-Wireless with Masimo SPO2Midmark Corporation · MONITOR,PHYSIOLOGICAL,PATIENT(WITHOUT ARRHYTHMIA DETECTION OR ALARMS)1 recall0 reports
- Midmark IQvitals Zone, Wireless with Masimo SPO2Midmark Corporation · MONITOR,PHYSIOLOGICAL,PATIENT(WITHOUT ARRHYTHMIA DETECTION OR ALARMS)1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 99641).
- Z-3217-2026 · Open, ClassifiedMidmark IQvitals Zone, automatically and noninvasively measures systolic and diastolic blood pressure, pulse rate, temperature (temporal), and oxygen saturation (SpO2) for adult an…
- Z-3218-2026 · Open, ClassifiedMidmark IQvitals Zone, automatically and noninvasively measures systolic and diastolic blood pressure, pulse rate, temperature (temporal), and oxygen saturation (SpO2) for adult an…
- Z-3220-2026 · Open, ClassifiedMidmark IQvitals Zone, automatically and noninvasively measures systolic and diastolic blood pressure, pulse rate, temperature (temporal), and oxygen saturation (SpO2) for adult an…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page