Skip to content

Captus 4000e Thyroid Uptake System, Models: 1. 5430-30151… recallZ-3193-2026 · Mirion Technologies (Capintec), Inc. · initiated Jul 20, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Mirion Technologies (Capintec), Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for failure of the tension bolt which would allow the spring arm to unexpectedly fall to its lowest position, which is approximately 25 inches (63.5 cm) from the ground.

FDA root cause category: Device Design

What the company is doing

On July 20, 2026 URGENT SAFETY NOTICE letters were emailed to customers. Actions to be taken: 1. Review this Alert and ensure that all affected personnel, including all operators of Captus 4000e Thyroid Uptake Systems, are aware of the contents. 2. Operate and inspect the device per the instructions provided. If your inspection shows that the collimator is not properly attached, please contact Capintec Technical Support at the phone number or email listed below. If you have any questions about this Inspection procedure of your Captus 4000 System, please contact Capintec Technical Support. If you notice any unusual change in the performance or functional response of the arm in your system (e.g. arm is making an unusual noise or arm does not move smoothly at any articulation joint), please contact Capintec Technical Support immediately. Capintec Technical Support: 1-800-631-3826 1-201-825-9500 [email protected]

Recall details

Recall number
Z-3193-2026
Status
Open, Classified
Initiated by the company
Jul 20, 2026
Posted by the FDA
Sep 16, 2026
Recalling firm
Mirion Technologies (Capintec), Inc.
Quantity
783
Product code
IZD
Distribution
Worldwide distribution - US Nationwide and the countries of ALGERIA, AUSTRIA, BANGLADESH, BRASIL, Brazil, CANADA, CHILE, Costa Rica, Ecuador, France, Hanoi Vietnam, Hong Kong, INDIA, ISRAEL, ITALY, Malaysia, Philippines, POLAND, Saudi Arabia, SOUTH KOREA, Spain, SWEDEN, TAIWAN, THAILAND, TURKEY, UNITED ARAB EMIATES, UNITED KINGDOM, VIETNAM.

Product description

Captus 4000e Thyroid Uptake System, Models: 1. 5430-30151, 2. 5430-30152, 3. 5430-30154;

Code information (models, lots, serial numbers)

Affected Units: Captus 4000: 1. Model 5430-30151; UDI-DI: 00859942006096; 2. Model 5430-30152; UDI-DI: 00859942006102; 3. Model 5430-30154; UDI-DI: 00859942006423; Devices shipped from April 2019 through February 2024, Serial Numbers 940548 through 941398, that have NOT already replaced the original arm with the redesigned arm.

Devices named in this recall

3 devices in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page