SunMed Tracheal Tube, Size 2.5 recallZ-3172-2026 · Well Lead Medical CO. Ltd · initiated Jul 31, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Well Lead Medical CO. Ltd or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The 6Fr suction catheter could not pass through the samples smoothly.
FDA root cause category: Process control
What the company is doing
On July 31, 2026 URGENT: MEDICAL DEVICE RECALL letters were emailed to customers. Actions to be taken: 1. We request that you check your inventory for product within the scope of this recall. Quarantine all impacted product immediately. Cease use and distribution immediately. 2. If the affected product is currently in use on a patient, do not perform a prophylactic endotracheal tube exchange solely due to this notification. Continue to manage the patient according to institutional protocols and clinical judgment. 3. If you do have stock in scope of this recall, mark the applicable checkbox on the Affected Product Information Table and contact Well Lead by E-mail . 4. If you do not have stock in scope of this recall, mark the applicable checkbox on the Affected Product Information Table and return the form to the E-mail address . 5. For affected products, please contact your distributor to arrange product replacement.
Recall details
- Recall number
- Z-3172-2026
- Status
- Open, Classified
- Initiated by the company
- Jul 31, 2026
- Posted by the FDA
- Sep 15, 2026
- Recalling firm
- Well Lead Medical CO. Ltd
- Quantity
- 27200
- Product code
- BTR
- FDA submissions
- K042683
- Distribution
- US Nationwide distribution in the state of MI.
Product description
SunMed Tracheal Tube, Size 2.5; Model Number: 1-7333-25
Code information (models, lots, serial numbers)
Model Number: 1-7333-25; UDI-DI:10889483048108; Lot No.: 2108011462, 2108011506, 2109011963, 2111012039, 2112012040, 2204010568, 2204010569, 2204010734, 2205010833, 2207011263, 2208011729, 2208011789, 2304010830, 2306011372, 2306011425, 2308011907, 2308012011, 2308012012, 2309012128, 2310012358, 2310012599, 2311013016, 2311012924, 2312013223, 2401010069, 2401010197, 2402010378, 2403010466, 2403010889, 2404010999, 2406011822, 2407012282, 2408012389, 2409012800, 2409012882, 2508013006, 2510013653, 2511014081, 2512014080, 2602010272, 2603010274, 2603010900, 2604011014, 2604011416, 2605011923, 2605011924;
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Sunmed 1-7333-25Sunmed, LLC · Tube, Tracheal (W/Wo Connector)1 recall2 reports
Other devices under the same FDA submission
25 devices cleared under K042683 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Medline Industries, Inc. DYND43075Medline Industries, Inc. · Tube, tracheal (w/wo connector)1 death11 reports
- Medline DYNJAETC75Medline Industries, Inc. · Tube, tracheal (w/wo connector)5 reports
- Medline Industries, Inc. DYND43070Medline Industries, Inc. · Tube, tracheal (w/wo connector)4 reports
- Medline Industries, Inc. DYND43080Medline Industries, Inc. · Tube, tracheal (w/wo connector)2 reports
- RediTube™ 8070RTCAirlife · Tube, tracheal (w/wo connector)2 reports
- Medline DYNJAETC30Medline Industries, Inc. · Tube, tracheal (w/wo connector)2 reports
- Medline DYNJAETC70Medline Industries, Inc. · Tube, tracheal (w/wo connector)2 reports
- Medline DYNJAETC80Medline Industries, Inc. · Tube, tracheal (w/wo connector)2 reports
- Medline DYNJAETC70SMedline Industries, Inc. · Tube, tracheal (w/wo connector)2 reports
- Sunmed 1-7333-75Sunmed, LLC · Tube, Tracheal (W/Wo Connector)2 reports
- Sunmed 1-7321-80Sunmed, LLC · Tube, Tracheal (W/Wo Connector)2 reports
- Medline Industries, Inc. DYND43040Medline Industries, Inc. · Tube, tracheal (w/wo connector)1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page