Why it was recalled

Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents, resulting in deeper/lighter anesthetic effects that may result in hypertonia/hypotonia, tachycardia/ bradycardia, elevated intracranial pressure, awareness during anesthesia, oversedation

FDA root cause category: Device Design

What the company is doing

On 8/18/2026, correction notices were emailed to customers and distributors who were informed of the following: The following warning will be added to associated Instructions for Use: Warning: Do not use IPV therapy in conjunction with volatile anesthetic delivery systems. The high-frequency percussive flow generated during IPV therapy may alter the controlled delivery of volatile anesthetic agents within the breathing circuit, resulting in unpredictable anesthetic agent concentrations that may result in serious patient harm. Note: Consult the anesthesia healthcare provider before disconnecting from the system if considering delivering IPV directly to the patient's airway. Customers were asked to do the following: 1) Print and post the new warning for applicable IPV devices throughout your facility to ensure adequate distribution of new information to users. 2) Reinforce adherence to the IPV User Guide and IPV Quick Reference Guide instructions, including ensuring communication and training primary users of IPV therapy, specifically those using it in-line with a ventilator, on the instructions and new warning. Distributors were asked to do the following: - Translate (if applicable) and send the Urgent Medical Device Field Safety Notice to affected end customers. -If the Distributor receives the signed acknowledgement letters from affected customers, please forward them and all related communication to [email protected]. If there are any additional questions related to the updated instructions or other questions, please contact [email protected]

Recall details

Recall number
Z-3325-2026
Status
Open, Classified
Initiated by the company
Aug 18, 2026
Posted by the FDA
Sep 28, 2026
Recalling firm
Percussionaire Corporation
Quantity
419
Product code
NHJ
Distribution
Worldwide - US Nationwide distribution including in the states of CA, PA, OK, GA, TX, OH, WI, WA, NY, MO, FL, DE, LA, SC, WV, NJ, NE, KS, NC, AL, CO, VA, OR, AZ, IL, CT, ID, NH, UT, TN, MD, VT, MS, NV, DC, ME, IA, AR, HI, ND, IN, MT, NM, MN, MI, SD, MA, AK, RI and the countries of Canada, Belgium, Chile, France, Germany, Israel, Italy, Latvia, Netherlands, Norway, Oman, Philippines, Portugal, Qatar, Russia, Saudi Arabia, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom, Kuwait, Australia, South Africa.

Product description

Intrapulmonary Percussive Ventilation, IPV HC, Product Number: S00009; Phasitron, P5-HC

Code information (models, lots, serial numbers)

Product/Catalog Number/UDI-DI: S00009/00849436000358, P5-HC/00849436000525