Impulsator, Product Number: S00010, Phasitron 5, Product… recallZ-3323-2026 · Percussionaire Corporation · initiated Aug 18, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Percussionaire Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents, resulting in deeper/lighter anesthetic effects that may result in hypertonia/hypotonia, tachycardia/ bradycardia, elevated intracranial pressure, awareness during anesthesia, oversedation
FDA root cause category: Device Design
What the company is doing
On 8/18/2026, correction notices were emailed to customers and distributors who were informed of the following: The following warning will be added to associated Instructions for Use: Warning: Do not use IPV therapy in conjunction with volatile anesthetic delivery systems. The high-frequency percussive flow generated during IPV therapy may alter the controlled delivery of volatile anesthetic agents within the breathing circuit, resulting in unpredictable anesthetic agent concentrations that may result in serious patient harm. Note: Consult the anesthesia healthcare provider before disconnecting from the system if considering delivering IPV directly to the patient's airway. Customers were asked to do the following: 1) Print and post the new warning for applicable IPV devices throughout your facility to ensure adequate distribution of new information to users. 2) Reinforce adherence to the IPV User Guide and IPV Quick Reference Guide instructions, including ensuring communication and training primary users of IPV therapy, specifically those using it in-line with a ventilator, on the instructions and new warning. Distributors were asked to do the following: - Translate (if applicable) and send the Urgent Medical Device Field Safety Notice to affected end customers. -If the Distributor receives the signed acknowledgement letters from affected customers, please forward them and all related communication to [email protected]. If there are any additional questions related to the updated instructions or other questions, please contact [email protected]
Recall details
- Recall number
- Z-3323-2026
- Status
- Open, Classified
- Initiated by the company
- Aug 18, 2026
- Posted by the FDA
- Sep 28, 2026
- Recalling firm
- Percussionaire Corporation
- Quantity
- 228,277
- Product code
- NHJ
- Distribution
- Worldwide - US Nationwide distribution including in the states of CA, PA, OK, GA, TX, OH, WI, WA, NY, MO, FL, DE, LA, SC, WV, NJ, NE, KS, NC, AL, CO, VA, OR, AZ, IL, CT, ID, NH, UT, TN, MD, VT, MS, NV, DC, ME, IA, AR, HI, ND, IN, MT, NM, MN, MI, SD, MA, AK, RI and the countries of Canada, Belgium, Chile, France, Germany, Israel, Italy, Latvia, Netherlands, Norway, Oman, Philippines, Portugal, Qatar, Russia, Saudi Arabia, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom, Kuwait, Australia, South Africa.
Product description
Impulsator, Product Number: S00010, Phasitron 5, Product Number: P5 used with Instruction Manual
Code information (models, lots, serial numbers)
Product Number/UDI-DI: S00010/00849436000389, P5/00849436000501, Instruction Manual: P20047 Rev E
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Percussionaire P5Percussionaire Corporation · Device, Positive Pressure Breathing, Intermittent4 recalls2 reports
- Percussionaire F00012Percussionaire Corporation · Device, Positive Pressure Breathing, Intermittent1 recall3 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 99617).
- Z-3319-2026 · Open, ClassifiedIntrapulmonary Percussive Ventilation, IPV 1, Product Number: S00065; Phasitron 5 Breathing Circuit with Unified Connector, Product Number: P5-UC, used with In-Line Valve Instruct…
- Z-3320-2026 · Open, ClassifiedIn-Line Valve, Product Number: P5-TEE
- Z-3321-2026 · Open, ClassifiedIntrapulmonary Percussive Ventilation, IPV-1C, Product Number: S00002-C; Phasitron 5, Product Number: P5 used with Instruction Manual
- Z-3322-2026 · Open, ClassifiedIntrapulmonary Percussive Ventilation, IPV-2C, Product Number: S00002-C; Phasitron 5, Product Number: P5 used with Instruction Manual
- Z-3324-2026 · Open, ClassifiedTravel Air, Product Number: S00055; Phasitron 5, Product Number: P5 used with Instruction Manual
- Z-3325-2026 · Open, ClassifiedIntrapulmonary Percussive Ventilation, IPV HC, Product Number: S00009; Phasitron, P5-HC
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page