Skip to content

Percussionaire P5-HCRecalls & reported problems

Percussionaire Corporation · IPV Phasitron Breathing Circuit Kit, Single Patient, HC Phasitron 5 · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K905236); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Aug 18, 2026Class IIOpen, Classified

    Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents, resulting in deeper/lighter anesthetic effects that may result in hypertonia/hypotonia, tachycardia/ bradycardia, elevated intracranial pressure, awareness during anesthesia, oversedation

    Z-3325-2026 · Root cause: Device Design · Recall details · FDA record ↗

FDA clearance and classification

Cleared through 510(k) K905236 on Apr 18, 1991 (decision: Substantially Equivalent). Submitted by Percussionaire Corp. as “Ipv Percussionator Family & Related Accessories”. The FDA review took 162 days. FDA record ↗

The FDA regulates product code NHJ (device, positive pressure breathing, intermittent) under 21 CFR 868.5905 as a Class II device. 21 CFR 868.5905 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K905236. One submission often covers many models, so check the product and lot details in each record.

  • Z-0208-2025 · Sep 16, 2024 · Open, Classified
    Phasitron5 Breathing Circuit (P5-10, P5-HC, P5-UC-10). component of Intrapulmonary Percussive Ventilation, used to provide airway clearance therapy

Common questions about the Percussionaire P5-HC

Has the Percussionaire P5-HC been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-3325-2026 (Class II), began on Aug 18, 2026: Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents,… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

Is the Percussionaire P5-HC FDA approved?

It was cleared through 510(k) K905236 on Apr 18, 1991. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the Percussionaire P5-HC, and is it still sold?

The FDA's device database lists Percussionaire Corporation as the company responsible for it (brand Percussionaire, model P5-HC). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Device, Positive Pressure Breathing, Intermittent devices

All Device, Positive Pressure Breathing, Intermittent devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page