INTERVENTIONAL ANGIOGRAPHY SYSTEM recallZ-0289-2027 · Canon Medical System, USA, INC. · initiated Oct 12, 2026
Class II Open, Classified All recalls posted in October 2026
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Canon Medical System, USA, INC. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to software issue, X-ray interventional system may experience internal error if F-Store or imagine acquisition is performed while clinical analysis function is use.
FDA root cause category: Software design
What the company is doing
On October 12, 2026, Canon Medical Systems USA, Inc. issued a Urgent: Medical Device Correction to affected consignees. Canon provided the following to consignees: 1. A representative of Canon Medical Systems USA will contact you to schedule a time to conduct installation of modified software to prevent internal errors from occuring if F-STORE or image acquisition is performed while CAAS clinical analysis function is used. 2. If you have any questions regarding this matter, please contact your service representative or contact our InTouch team at (800) 521-1968 or by email [email protected]. 3. In order to comply with the FDA record keeping requirement, Canon respectfully requests that the attached form be completed and returned via email to [email protected]
Recall details
- Recall number
- Z-0289-2027
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Oct 12, 2026
- Posted by the FDA
- Oct 27, 2026
- Recalling firm
- Canon Medical System, USA, INC.
- Quantity
- 15 SYSTEMS
- Product code
- OWB
- FDA submissions
- K210900
- Distribution
- US Nationwide distribution in the states of AL, AR, AZ, CA, FL, GA, HI, IL, LA, MA, MI, MN, MS, NC, NV, OH, OR, TX.
Product description
ALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM REF: INFX-8000C
Code information (models, lots, serial numbers)
DI-04987670100192
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- INTERVENTIONAL ANGIOGRAPHY SYSTEM INFX-8000CCanon Inc. · System, x-ray, fluoroscopic, image-intensified5 recalls5 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 99889).
- Z-0290-2027 · Open, ClassifiedALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM REF: INFX-8000F
- Z-0291-2027 · Open, ClassifiedALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM REF: INFX-8000V
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Nov 27, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Nov 27, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Nov 27, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page