Skip to content

INTERVENTIONAL ANGIOGRAPHY SYSTEM recallZ-0291-2027 · Canon Medical System, USA, INC. · initiated Oct 12, 2026

Class II Open, Classified All recalls posted in October 2026

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Canon Medical System, USA, INC. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to software issue, X-ray interventional system may experience internal error if F-Store or imagine acquisition is performed while clinical analysis function is use.

FDA root cause category: Software design

What the company is doing

On October 12, 2026, Canon Medical Systems USA, Inc. issued a Urgent: Medical Device Correction to affected consignees. Canon provided the following to consignees: 1. A representative of Canon Medical Systems USA will contact you to schedule a time to conduct installation of modified software to prevent internal errors from occuring if F-STORE or image acquisition is performed while CAAS clinical analysis function is used. 2. If you have any questions regarding this matter, please contact your service representative or contact our InTouch team at (800) 521-1968 or by email [email protected]. 3. In order to comply with the FDA record keeping requirement, Canon respectfully requests that the attached form be completed and returned via email to [email protected]

Recall details

Recall number
Z-0291-2027
Recall class
Class II
Status
Open, Classified
Initiated by the company
Oct 12, 2026
Posted by the FDA
Oct 27, 2026
Recalling firm
Canon Medical System, USA, INC.
Quantity
31 units
Product code
OWB
FDA submissions
K203551, K233107
Distribution
US Nationwide distribution in the states of AL, AR, AZ, CA, FL, GA, HI, IL, LA, MA, MI, MN, MS, NC, NV, OH, OR, TX.

Product description

ALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM REF: INFX-8000V

Code information (models, lots, serial numbers)

DI-04987670100222/

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 99889).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page