Why it was recalled

An incomplete seal on the inner pouch allows exposure to oxygen, which may result in oxidation and would affect long-term performance of the implant.

FDA root cause category: Other

Recall details

Recall number
Z-1317-05
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
May 26, 2005
Posted by the FDA
Aug 12, 2005
Recalling firm
Depuy Orthopaedics, Inc.
Product code
JWH
FDA submissions
K961685
Distribution
Nationwide, Argentina, Brazil, Canada, Caguas, Colombia, Ireland, Malaysia, Mexico, Panama, Peru, Singapore, South Africa, Sweden, Uruguay and Venezuela.

Product description

Depuy SIGMA curved plus insert; polyethylene tibial insert, Sz 3, 15.0 mm; Ref 97-0353.

Code information (models, lots, serial numbers)

Lots X23AD4 and X23AE4.

Other devices under the same FDA submission

34 devices cleared under K961685 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 32192).

Sources and updates