Skip to content

Depuy SIGMA curved plus insert recallZ-1320-05 · Depuy Orthopaedics, Inc. · initiated May 26, 2005

Terminated All recalls posted in August 2005

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Depuy Orthopaedics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

An incomplete seal on the inner pouch allows exposure to oxygen, which may result in oxidation and would affect long-term performance of the implant.

FDA root cause category: Other

Recall details

Recall number
Z-1320-05
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
May 26, 2005
Posted by the FDA
Aug 12, 2005
Recalling firm
Depuy Orthopaedics, Inc.
Quantity
19
Product code
JWH
FDA submissions
K961685
Distribution
Nationwide, Argentina, Brazil, Canada, Caguas, Colombia, Ireland, Malaysia, Mexico, Panama, Peru, Singapore, South Africa, Sweden, Uruguay and Venezuela.

Product description

Depuy SIGMA curved plus insert; polyethylene tibial insert, Sz 5, 15.0 mm; Ref 97-0373.

Code information (models, lots, serial numbers)

Lots Y5YA84 and Y5YA94.

Other devices under the same FDA submission

34 devices cleared under K961685 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 32192).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page