Skip to content

CONTAK RENEWAL 4 CRT-D (model no. H190, H195, H199).… recallZ-0891-06 · Guidant Corporation · initiated Sep 30, 2005

Terminated All recalls posted in May 2006

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Guidant Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Devices manufactured with batteries built on April 26 or 27, 2005, by a single operator, may have battery welds that may have been performed at settings outside of typical manufacturing specifications at the supplier.

FDA root cause category: Other

What the company is doing

Guidant issued a Physicain letter, dated 10/25/05 to implanting/following physicians. The letter states the issues and reommends patient follow-up care every three months, as stated in published labeling. An Advisory Update, dated 12/20/05, was sent to physicians that describes the issue, states that here have been zero reports of battery weld failures on implanted devices and the results of laboratory testing (accelerated testing) in which they tried to induce the failure. The results of the testing induced 2 battery weld post failures. A projected rate of occurence was also given. Patient management recommendations have not changed.

Recall details

Recall number
Z-0891-06
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Sep 30, 2005
Posted by the FDA
May 23, 2006
Recalling firm
Guidant Corporation
Quantity
70
Product code
NIK
FDA submissions
P010012
Distribution
Nationwide (AL, AZ, CA, CO, DC, FL, GA, HI, IA, IL, IN, KY, LA, MA, MI, MN, MO, MT, NC, ND, NE, NJ, NV, NY, OH, OK, PA, TN, TX, VA, WA, WI). OUS to include: Australia, France, Germany, India, Ireland, Israel, Italy, Netherlands, Norway, South Africa, Spain, Switzerland, Turkey, United Kingdom.

Product description

CONTAK RENEWAL 4 CRT-D (model no. H190, H195, H199). Cardiac Resynchronization Therapy Defibrillator. Sterilized with gaseous ethylene oxide. Guidant Corporation, Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798 USA. CONTAK RENEWAL 4 devices provide ventricular tachyarrhythmia and cardiac resynchronization therapies.

Code information (models, lots, serial numbers)

Model H190, serial numbers: 310258, 310260, 310293, 310312, 310317, 310319, 310320, 310321, 310322, 310343,  310353, 310355, 310381, 310383, 310386, 310387, 310388,  310389, 310390.   Model H195, serial numbers: 350236, 350239, 350245,  350247, 350248, 350250, 350252, 350255, 350261, 350264,  350276, 350279, 350280, 350282, 350284, 350319, 350330,  350331, 350332, 350333, 350334, 350336, 350342, 350344,  350350, 350351, 350353, 350354, 350355, 350360, 350362,  350363, 350364, 350365, 350403, 350407, 350442, 350451,  350480, 350484, 350503, 350570, 350649.  Model H199, serial numbers: 320330, 320331, 320361, 320366, 320370, 320415, 320417, 320445.

Other devices under the same FDA submission

50 devices cleared under P010012 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page