EASYTRAK® 2 IS-1 4543 (UDI 00802526410994)Recalls & reported problems
Boston Scientific Corporation · Coronary Venous Steroid-Eluting Bipolar Pace/Sense Lead · Class 3 · Implant
All models of EASYTRAK 2 IS-1 (6)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Boston Scientific EASYTRAK 2 lead is a wire placed in a heart vein to pace the left side of the heart. Its 724 reports often describe high electrical resistance, failure to trigger heartbeats and fractures in the lead, and many mention infections. The reports mention 5 deaths and 595 injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 727
- Deaths reported
- 5
- Injuries reported
- 598
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 727 adverse event reports that list this exact device (DI 00802526410994), received between Apr 9, 2012 and Sep 24, 2026 — classified as 5 deaths, 598 injuries, 124 malfunctions.
The most frequently reported specific device problem is “High impedance” (255 reports).
19 other recalls cover devices cleared under the same FDA submission (P010012); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- High impedance255
- Adverse Event Without Identified Device or Use Problem201
- Failure to Capture196
- High Capture Threshold120
- Fracture108
- Device Dislodged or Dislocated60
- Material Integrity Problem52
- Defective Device41
Patient problems
- No Clinical Signs, Symptoms or Conditions207
- Unspecified Infection181
- No Known Impact Or Consequence To Patient144
- No Consequences Or Impact To Patient79
- Sepsis64
- No Code Available63
- Device Overstimulation of Tissue21
- Muscle Stimulation18
Recent adverse event reports
The 25 most recent of 727 reports, in the reporter's own words as published by the FDA.
InjurySep 24, 2026Insufficient Device Problem Information; High Capture Threshold
It was reported that this left ventricular (LV) lead was surgically abandoned due to an unknown product anomaly. A new lead was successfully implanted. No additional adverse patient effects were reported. This LV lead is no longer in service.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50885206 ↗InjurySep 14, 2026High impedance; Material Integrity Problem
It was reported that this left ventricular (LV) lead exhibited high out of range pacing impedance measurements. It was later discovered this LV lead was fractured. The cause and location of the fracture on the lead are unknown. Subsequently, this LV lead was surgically abandoned and successfully replaced. No additional adverse patient effects were reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50756331 ↗InjurySep 4, 2026Insufficient Device Problem Information; High Capture Threshold
It was reported that this left ventricular (LV) lead was surgically abandoned due to an unknown product performance issue. This LV lead was successfully replaced. No additional adverse patient effects were reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50602093 ↗MalfunctionAug 14, 2026Failure to Capture
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50246164 ↗InjuryJul 8, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THIS LEFT VENTRICULAR (LV) LEAD WAS PART OF A SYSTEM REVISION DUE TO INFECTION. THERE WERE NO ADDITIONAL ADVERSE PATIENT EFFECTS REPORTED. THE LV LEAD WAS EXPLANTED.
Patient problems: Unspecified Infection
MAUDE report 25756473 ↗InjuryMay 18, 2026Use of Device Problem; Material Integrity Problem
IT WAS REPORTED THAT THIS LEFT VENTRICULAR (LV) LEAD WAS CUT UPON REMOVAL. THIS LV LEAD WAS EXPLANTED, REPLACED WITH DIFFERENT MODEL AND WILL NOT BE RETURNED FOR ANALYSIS. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25200035 ↗InjuryMay 12, 2026Fracture; High impedance
IT WAS REPORTED THAT THIS LEFT VENTRICULAR (LV) LEAD EXHIBITED A RISE AND THEN A SUDDEN JUMP OF PACING IMPEDANCE MEASUREMENTS REACHING HIGHER THAN 3000 OHMS. ADDITIONALLY, IT WAS NOTED THIS LEAD WAS FRACTURED. THIS LV LEAD WAS SUBSEQUENTLY EXPLANTED AND SUCCESSFULLY REPLACED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25156228 ↗MalfunctionMay 1, 2026High impedance
IT WAS REPORTED THAT THIS LEFT VENTRICULAR (LV) LEAD EXHIBITED HIGH OUT OF RANGE PACING IMPEDANCE MEASUREMENTS. IT WAS ALSO NOTED THAT IT APPEARED THE THRESHOLD HAD DROPPED WHEN THE IMPEDANCE WENT UP. TECHNICAL SERVICES (TS) WAS CONSULTED AND DISCUSSED CONDUCTING TESTING IN PERSON. THIS LV LEAD REMAINS IN SERVICE AND NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25056729 ↗InjuryApr 27, 2026High impedance
IT WAS REPORTED THAT THIS SYSTEM WITH A LEFT VENTRICULAR (LV) LEAD AND CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) RECORDED UNSTABLE PACING IMPEDANCES (PI) THAT HAVE REACHED HIGH, OUT OF RANGE VALUES. PATIENT FELT SHORTNESS OF BREATH AND SYNCOPAL WHEN PACING IMPEDANCE VALUES WERE OUT OF RANGE. THRESHOLDS WERE STABLE ON DAYS WHEN PI WAS OOR AND CHANNEL WAS ABSENT OF NOISE. PRESENTING ELECTROGRAMS UNCHANGED. IT WAS RECOMMENDED TO CONSIDER SERIAL IMPEDANCES WITH ISOMETRIC MOVEMENTS IF NOT PERFORMED YET AND EVALUATE RING CONFIGURATIONS. AS PATIENT IS NOT DEPENDENT AND RV STABLE, SO CAN CONTINUE TO MONITOR THE LV AND CRT-D THAT REMAIN IN SERVICE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Dyspnea; Syncope/Fainting
MAUDE report 25010333 ↗InjuryApr 22, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THIS PREVIOUSLY CAPPED LEFT VENTRICULAR (LV) LEAD WAS PART OF A SYSTEM REVISION DUE TO INFECTION. THERE WERE NO ADDITIONAL ADVERSE PATIENT EFFECTS REPORTED. THIS PREVIOUSLY CAPPED LV LEAD WAS EXPLANTED.
Patient problems: Unspecified Infection
MAUDE report 24960627 ↗MalfunctionApr 20, 2026Failure to Capture; High impedance
IT WAS REPORTED THAT THIS CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM EXHIBITED HIGH PACING IMPEDANCE MEASUREMENTS OUT OF RANGE AND LOSS OF CAPTURE (LOC) ON THE LEFT VENTRICULAR (LV) LEAD. IT WAS ALSO NOTED THAT THE PATIENT EXPERIENCED PALPITATIONS. TECHNICAL SERVICES (TS) REVIEWED AND PROVIDED ASSISTANCE. THIS PRODUCT REMAINS IN SERVICE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Palpitations; Unspecified Heart Problem
MAUDE report 24925717 ↗InjuryApr 4, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THIS IMPLANTABLE LEAD WAS PART OF THE ALLERGIC REACTION AND ITCHING ISSUE. REPORTEDLY, THE PATIENT HAD A SEVERE ITCHING AND ALLERGY PROBLEM DUE TO METAL MATERIALS OF THE PACEMAKER. THE PATIENT WAS THEN TRANSFERRED TO THE BOSTON SCIENTIFIC TECHNICAL SERVICES (TS) GROUP. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THE IMPLANTABLE LEAD REMAINS IMPLANTED.
Patient problems: Hypersensitivity/Allergic reaction; Itching Sensation
MAUDE report 24785515 ↗InjuryMar 20, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THIS IMPLANTABLE LEAD WAS PART OF THE SYSTEM REVISION DUE TO INFECTION AND POCKET EROSION. REPORTEDLY, THE PATIENT HAD A HOLE IN THE CHEST AND WAS HOSPITALIZED FOR A YEAR. THE PATIENT WAS THEN REFERRED TO WARRANTY DEPARTMENT AFTER THE FIRST CALL DUE TO THE ALLEGATION OF PAIN AND DISCOMFORT BUT WAS INFORMED THAT THE INFECTION, EROSION OR MIGRATION ISSUES WERE A KNOWN MEDICAL RISK OF IMPLANTED PRODUCTS, THEREFORE THE EXPLANTED PRODUCTS DID NOT MEET THE WARRANTY ELIGIBILITY. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THE IMPLANTABLE LEAD WAS EXPLANTED.
Patient problems: Discomfort; Pain; Pocket Erosion; Unspecified Infection
MAUDE report 24646500 ↗InjuryFeb 27, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THIS LEFT VENTRICULAR (LV) LEAD WAS PART OF A SYSTEM REVISION DUE TO INFECTION. THERE WERE NO ADDITIONAL ADVERSE PATIENT EFFECTS REPORTED. THIS LV LEAD WAS EXPLANTED.
Patient problems: Unspecified Infection
MAUDE report 24470294 ↗InjuryFeb 23, 2026Failure to Capture; High Capture Threshold
BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS LEFT VENTRICULAR (LV) LEAD WAS SURGICALLY ABANDONED THREE MONTHS POST IMPLANT DUE TO DIAPHRAGMATIC STIMULATION. AN EPICARDIAL LV LEAD WAS SUCCESSFULLY IMPLANTED. ADDITIONAL INFORMATION RECEIVED APPROXIMATELY 17 YEARS LATER REPORTED THAT LV LEAD WAS ACTIVE IN THE PATIENT'S CURRENT CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P) SYSTEM. THE PATIENT PRESENTED TO THE HOSPITAL DUE TO SYMPTOMS CONSISTENT WITH PHRENIC NERVE STIMULATION (PNS) DESCRIBED AS SMALL SHOCKS AND HICCUP TYPE SENSATIONS IN THE CHEST. A LOCAL FIELD REPRESENTATIVE WAS PAGED FOR FURTHER EVALUATION. THE FOLLOWING DAY, RIGHT ATRIAL (RA) LEAD LOSS OF CAPTURE (LOC) WAS NOTED. SINUS ARREST WAS SUSPECTED DUE TO NO SENSING OF P-WAVES WITH RA SENSITIVITY AT 0.15 MV. ADDITIONALLY, THERE WAS INTERMITTENT LEFT VENTRICULAR (LV) LEAD LOC WITH A LV THRESHOLD MEASUREMENT OF 3.5 V @ 2.0 MS. LV OUTPUT WAS 3.5 V @ 2.0 MS, AND ALL OTHER VECTORS SHOWED THRESHOLD MEASUREMENTS GREATER THAN 5 V. ULTIMATELY, THE PHYSICIAN CHOSE TO PROGRAM THE CRT-P TO VVIR 60 BEATS PER MINUTE (BPM) AND RIGHT VENTRICULAR (RV) ONLY. ADDITIONAL INFORMATION RECEIVED REPORTED THAT SINUS ARREST WAS CONFIRMED
Patient problems: Asystole; Device Overstimulation of Tissue
MAUDE report 24420875 ↗InjuryFeb 19, 2026Device Dislodged or Dislocated
IT WAS REPORTED THAT THE LEFT VENTRICULAR (LV) LEAD WAS MISPLACED. THE LV LEAD WAS EXPLANTED, AND A NEW LEAD OF A DIFFERENT MODEL WAS SUCCESSFULLY IMPLANTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24389227 ↗InjuryFeb 11, 2026Fracture
IT WAS REPORTED THAT THE LEFT VENTRICULAR (LV) LEAD WAS EXPLANTED DUE TO FRACTURED. A NEW LEAD WAS IMPLANTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24319040 ↗MalfunctionJan 26, 2026High impedance
IT WAS REPORTED THAT WHEN THIS PATIENT WAS GOING TO BE PROGRAMMED INTO MRI PROTECTION MODE THE LEFT VENTRICULAR (LV) LEAD PACE IMPEDANCES WERE FOUND TO BE HIGH OUT OF RANGE. THIS LV LEAD REMAINS IN SERVICE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24185270 ↗InjuryJan 23, 2026Insufficient Device Problem Information; High Capture Threshold
IT WAS REPORTED THAT THIS LEFT VENTRICULAR (LV) LEAD WAS CAPPED DUE TO AN UNKNOWN PRODUCT PERFORMANCE ANOMALY. A NEW LEAD WAS IMPLANTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Device Overstimulation of Tissue; No Clinical Signs, Symptoms or Conditions
MAUDE report 24166444 ↗InjuryJan 21, 2026Insufficient Device Problem Information
IT WAS REPORTED THAT THIS LEFT VENTRICULAR (LV) LEAD WAS EXPLANTED AND REPLACED DUE TO UNKNOWN REASONS. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24131883 ↗MalfunctionJan 15, 2026High impedance
IT WAS REPORTED THAT THIS CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P) EXHIBITED HIGH PACING IMPEDANCE OUT OF RANGE (OOR) MEASUREMENTS ON THE LEFT VENTRICULAR (LV) LEAD, WHICH TRIGGERED TO LEAD SAFETY SWITCH (LSS) FROM BIPOLAR TO UNIPOLAR MODE. NO EVIDENCE OF LEAD NOISE AND IT WAS NOTED THAT THE MEASUREMENTS HAD BEEN STABLE IN THE RANGE. CLINICIAN WILL BRING THE PATIENT IN FOR FURTHER EVALUATION. CURRENTLY, THIS PRODUCT REMAINS IN SERVICE AND NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24086273 ↗InjuryJan 5, 2026Signal Artifact/Noise; Failure to Capture; High impedance; High Capture Threshold
IT WAS REPORTED THAT THIS CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM EXHIBITED HIGH PACING IMPEDANCE MEASUREMENTS OUT OF RANGE, NOISY SIGNALS AND LOSS OF CAPTURE (LOC) ON THE LEFT VENTRICULAR (LV) LEAD. TECHNICAL SERVICES (TS) REVIEWED AND SUSPECTED THIS TO BE A LEAD ISSUE. TS WAS CONSULTED AND CLARIFIED THAT TURNING THIS LV OFF WOULD NOT IMPACT TACHYCARDIA THERAPY. THIS LEAD REMAINS IN SERVICE. NO ADVERSE PATIENT EFFECTS WERE REPORTED. ADDITIONAL INFORMATION WAS RECEIVED STATING THIS LV WAS PROGRAMMED OFF. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED. ADDITIONAL INFORMATION WAS RECEIVED STATING THIS LV WAS CAPPED AND REPLACED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23955037 ↗MalfunctionDec 10, 2025Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THAT THE PATIENT WITH THIS LEFT VENTRICULAR (LV) LEAD RECEIVED A CALL FROM THE PREVIOUS CLINIC AND WAS INFORMED ABOUT POSSIBLE CORROSION ON DEVICE WIRES. THE PATIENT EXPRESSED CONCERN BECAUSE A DEVICE CHECK WAS COMPLETED AT A DIFFERENT FACILITY THE PREVIOUS WEEK, AND ALL RESULTS WERE NORMAL. A CLARIFICATION WAS REQUESTED REGARDING WHY BOSTON SCIENTIFIC CONTACTED THE PREVIOUS CLINIC. TO DATE, THIS LV LEAD REMAINS IN SERVICE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23771262 ↗MalfunctionDec 4, 2025Failure to Capture; High impedance; Failure to Capture; High impedance
IT WAS REPORTED THAT THIS LEFT VENTRICULAR (LV) LEAD RECORDED HIGH OUT OF RANGE PACING IMPEDANCE MEASUREMENTS DURING AN IN-CLINIC VISIT. TECHNICAL SERVICES (TS) REVIEWED THE DEVICE DATA AND RECOMMENDED THE LV LEAD REPLACEMENT. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THIS LV LEAD REMAINS IN SERVICE.ADDITIONAL INFORMATION WAS PROVIDED THAT THIS LV LEAD POTENTIALLY EXHIBITED LOSS OF CAPTURE (LOC), TS INDICATED THAT ALL LV PACING VECTORS HAD FAILED CAPTURE TESTING, AND AN ATTEMPTED LEAD AND SYSTEM REVISION PROCEDURE WAS UNSUCCESSFUL DUE TO VENOUS OCCLUSION, FOLLOWED BY AN UNSUCCESSFUL ATTEMPT AT ALTERNATIVE PACING. SUBSEQUENT DEVICE INTERROGATION DEMONSTRATED NO CAPTURE IN THE LV TIP-TO-RING CONFIGURATION WITH HIGH IMPEDANCE, WHILE SUCCESSFUL LV CAPTURE WAS ACHIEVED USING AN LV TIP-TO-RIGHT VENTRICULAR (RV) CONFIGURATION AT LOW OUTPUT. THE DEVICE WAS PROGRAMMED ACCORDINGLY, AND THE PATIENT WAS REPORTED TO BE RECOVERING. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THIS LV LEAD REMAINS IN SERVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23716789 ↗InjuryNov 27, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT THIS LEAD WAS EXPLANTED AND REPLACED DUE TO UNKNOWN REASONS. THE LEAD HAS BEEN RECEIVED FOR ANALYSIS. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23669098 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P010012. One submission often covers many models, so check the product and lot details in each record.
- Z-1671-2022 · Jul 7, 2022 · Open, ClassifiedThe PRM is intended to be used as part of the ZOOM LATITUDE Programming System to communicate with Boston Scientific implantable pulse generators. The software in use controls all
- Z-0080-2018 · Oct 4, 2017 · TerminatedMOMENTUM X4 CRT-D, Model G138, Sterile.
- Z-0081-2018 · Oct 4, 2017 · TerminatedRESONATE X4 CRT-D, Model G447, Sterile.
- Z-0079-2018 · Oct 4, 2017 · TerminatedVIGILANT X4 CRT-D, Model G247, Sterile. Cardiac Resynchronization Therapy Defibrillators (CRT-Ds) are indicated for patients with heart failure who receive stable optimal pharma
- Z-1556-2010 · Mar 15, 2010 · TerminatedBoston Scientific LIVIAN Cardiac Resynchronization Therapy Defibrillator (CRT-D), models H220, H225, H227, H229. The LIVIAN family of cardiac resynchronization therapy defibrillat
- Z-0609-2010 · Dec 1, 2009 · TerminatedBoston Scientific COGNIS CRT-D, Cardiac Resynchronization Therapy Defibrillator, models N118, N119. the following models are not available in the US: models N106, N107, P106, P107,
- Z-1049-2007 · Mar 23, 2007 · TerminatedCONTAK RENEWAL 3 RF (model H210, H215), CONTAK RENEWAL 3 RF HE (model H217, H219) cardiac resynchronization therapy defibrillator (CRT-D).
- Z-1054-2007 · Mar 23, 2007 · TerminatedVITALITY DR HE Implantable Cardioverter Defibrillator (model T180). Sterile EO. Guidant Corporation, Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798 U
- Z-1052-2007 · Mar 23, 2007 · TerminatedVITALITY AVT (model A155). Implantable Cardioverter Defibrillator. Atrial and Ventricular Therapies. Guidant Corporation Cardiac Rhythm Management, 4100 Hamline Avenue North, St. P
- Z-1047-2007 · Mar 23, 2007 · TerminatedCONTAK RENEWAL 3 (model H170, H175) CONTAK RENEWAL 3 HE (model H177, H179) Cardiac Resynchronization Therapy Defibrillator (CRT-D). Guidant Corporation, 4100 Hamline Avenue North,
- Z-1053-2007 · Mar 23, 2007 · TerminatedVITALITY 2 (model T165, T175), VITALITY DS (model T125, T135), VITALITY EL (model T127), Implantable Cardioverter Defibrillators (ICD), are designed to detect and terminate ventric
- Z-1055-2007 · Mar 23, 2007 · TerminatedVITALITY 2 EL DR/VR (model T167, T177). contents have been sterilized with gaseous ethylene oxide. The Guidant Vitality 2 ICD, implantable cardioverter defibrillators, are design
- Z-0047-2008 · Mar 16, 2007 · TerminatedGuidant VITALITY VR/DR, Implantable Cardioverter Defibrillator, ICD, Models 1870 and 1871. Guidant VITALITY DR+, Implantable Cardioverter Defibrillator, ICD, Model 1872.
- Z-1288-06 · May 15, 2006 · TerminatedGuidant CONTAK RENEWAL 3 (models H170, H175) CONTAK RENEWAL 3 HE (models H177, H179) Cardiac Resynchronization Therapy Defibrillator (CRT-D). Guidant Corporation, 4100 Hamline Ave
- Z-0826-06 · Apr 7, 2006 · TerminatedGuidant CONTAK RENEWAL 4 (models H190, H195) CONTAK RENEWAL; 4 HE (model H197). Cardiac Resynchronization Therapy Defibrillator (CRT-D). Sterile EO. Guidant Corporation, 4100 Ha
…and 4 more.
Other Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D) devices
- Gallant™ CDHFA500QSt. Jude Medical, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)2 recalls58 deaths8,401 reports
- Quartet™ 1458Q-86St. Jude Medical, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)207 deaths6,631 reports
- QUADRA ASSURA MP™ CD3369-40QSt. Jude Medical, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)103 deaths6,084 reports
- VIGILANT™ X4 CRT-D G247Boston Scientific Corporation · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)17 deaths2,066 reports
- Quadra Assura™ CD3365-40QSt. Jude Medical, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)53 deaths2,056 reports
- Cobalt™ XT HF Quad CRT-D MRI SureScan™ DTPA2QQMedtronic, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)4 recalls40 deaths1,639 reports
- Unify Assura™ CD3357-40QSt. Jude Medical, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)54 deaths1,784 reports
- Unify Assura™ CD3357-40CSt. Jude Medical, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)56 deaths1,588 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page