Guidant VITALITY HE Implantable Cardioverter Defibrillator… recallZ-0828-06 · Guidant Corporation · initiated Apr 7, 2006
Terminated All recalls posted in May 2006
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Guidant Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
77 distributed devices may have a cracked layer of insulation in a flexible hybrid circuit, leading to internal shorting during charging such as during routine capacitor formation or for defibrillation. This would likely render the device non-functional, resulting in no output and/or telemetry communications.
FDA root cause category: Other
What the company is doing
Guidant Sales Representatives are retrieving non-implanted devices April 10-12, 2006. If a device is implanted the physician is sent an Urgent Medical Device Safety Information & Corrective Action letter, April 11-14, 2006. This letter describes the issue, states that Guidant is retrieving all non-implanted devices and that Guidant is currently conducting tests on retrieved devices. Physicians will be advised once more information is available.
Recall details
- Recall number
- Z-0828-06
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Apr 7, 2006
- Posted by the FDA
- May 2, 2006
- Recalling firm
- Guidant Corporation
- Quantity
- 20 devices
- Product code
- NIK
- FDA submissions
- P960040
- Distribution
- Nationwide. AZ, FL, GA, IL, IN, KY, MA, ME, MI, MN, NC, NE, NY, OH, PA, TN, TX, UT, VA, WI. OUS to include: Australia, Canada, Germany, Italy, Netherlands, Portugal, Slovenia, Spain.
Product description
Guidant VITALITY HE Implantable Cardioverter Defibrillator (model T180). Sterile EO. Guidant Corporation, Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798 USA. This ICD is designed to detect and terminate ventricular tachycardia (VT) and ventricular fibrillation (VF) and provide bradycardia therapy (atrial and ventricular pacing). Therapies include both low- and high-eneergy shocks using either a biphasic or monophasic waveform. Bradycardia pacing, including adaptive-rate features, is available to detect and treat bradyarrhythmias and to support the cardiac rhythm after defibrillation therapy. The device offers dual-chamber bradycardia features (atrial and/or ventricular pacing and sensing).
Code information (models, lots, serial numbers)
model T180 serial numbers: 204707, 204711, 204751, 204754, 204908.
Other devices under the same FDA submission
45 devices cleared under P960040 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- VIGILANT™ EL ICD DR D233Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)9 deaths1,563 reports
- DYNAGEN™ EL ICD DR D152Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)12 deaths1,208 reports
- INOGEN™ EL ICD DR D142Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)6 deaths1,017 reports
- RESONATE™ EL ICD DR D433Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)4 deaths965 reports
- DYNAGEN™ EL ICD DR D153Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)4 deaths886 reports
- MOMENTUM™ EL ICD DR D121Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)2 deaths850 reports
- DYNAGEN™ EL ICD VR D150Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)8 deaths831 reports
- INOGEN™ EL ICD VR D140Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)4 deaths823 reports
- ENERGEN™ ICD E140Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)3 deaths819 reports
- VIGILANT™ EL ICD VR D232Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)8 deaths750 reports
- ENERGEN™ ICD E142Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)1 death705 reports
- INCEPTA™ ICD E160Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)2 deaths601 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 35166).
- Z-0825-06 · TerminatedGuidant CONTAK RENEWAL 3 (models H170,H175) CONTAK RENEWAL 3 HE (models H177, H179) Cardiac Resynchronization Therapy Defibrillator (CRT-D). Sterile EO. Guidant Corporation, 4100…
- Z-0826-06 · TerminatedGuidant CONTAK RENEWAL 4 (models H190, H195) CONTAK RENEWAL; 4 HE (model H197). Cardiac Resynchronization Therapy Defibrillator (CRT-D). Sterile EO. Guidant Corporation, 4100 Ha…
- Z-0827-06 · TerminatedGuidant CONTAK RENEWAL 4 AVT (model M170). Cardiac Resynchronization Therapy Defibrillator (CRT-D). Sterile EO. Guidant Corporation, 4100 Hamline Avenue North, St. Paul, MN 55…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page