Guidant CONTAK RENEWAL 3 (models H170,H175) CONTAK RENEWAL… recallZ-0825-06 · Guidant Corporation · initiated Apr 7, 2006
Terminated All recalls posted in May 2006
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Guidant Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
77 distributed devices may have a cracked layer of insulation in a flexible hybrid circuit, leading to internal shorting during charging such as during routine capacitor formation or for defibrillation. This would likely render the device non-functional, resulting in no output and/or telemetry communications.
FDA root cause category: Other
What the company is doing
Guidant Sales Representatives are retrieving non-implanted devices April 10-12, 2006. If a device is implanted the physician is sent an Urgent Medical Device Safety Information & Corrective Action letter, April 11-14, 2006. This letter describes the issue, states that Guidant is retrieving all non-implanted devices and that Guidant is currently conducting tests on retrieved devices. Physicians will be advised once more information is available.
Recall details
- Recall number
- Z-0825-06
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Apr 7, 2006
- Posted by the FDA
- May 2, 2006
- Recalling firm
- Guidant Corporation
- Quantity
- 65 devices
- Product code
- NIK
- FDA submissions
- P010012
- Distribution
- Nationwide. AZ, FL, GA, IL, IN, KY, MA, ME, MI, MN, NC, NE, NY, OH, PA, TN, TX, UT, VA, WI. OUS to include: Australia, Canada, Germany, Italy, Netherlands, Portugal, Slovenia, Spain.
Product description
Guidant CONTAK RENEWAL 3 (models H170,H175) CONTAK RENEWAL 3 HE (models H177, H179) Cardiac Resynchronization Therapy Defibrillator (CRT-D). Sterile EO. Guidant Corporation, 4100 Hamline Avenue North, St. Paul, MN 55112-5798 USA. Cardiac resynchronization therapy defibrillators (CRT-D) provide ventricular tachyarrhythmia and cardiac resynchronization therapies. Ventricular tachyarrhythmia therapy is for the treatment of ventricular tachycardia (VT) and ventricular fibrillation (VF), rhythms that are associated with sudden cardiac death (SCD). Cardiac resynchronization therapy is for the treatment of heart failure (HF) and uses biventricular electrical stimulation to synchronize ventricular contractions. Cardioversion/defibrillation therapies include a range of low-and high-energy shocks using either a biphasic or monophasic waveform.
Code information (models, lots, serial numbers)
model H170 serial numbers: 364406, 364446, 364542, 364576, 364582, 364586, 364587, 364590, 364591, 364593, 364595, 364596, 364597, 364599, 364600, 364601, 364602, 364604, 364605, 364606, 364607, 364608, 364609, 364611, 364613, 364615, 364623, 364625, 364632, 364633, 364634, 364636, 364643, 364680, 364732, 364735, 364752. model H175 serial numbers: 315437, 315438, 315452, 315454, 315455, 315456, 315461, 315462, 315464, 315470, 315473, 315480, 315482, 315489, 315496, 315497, 315500, 315580, 315588, 315605. model H177 serial number: 285055 model H179 serial numbers: 114267, 114268, 114269, 114270, 114271, 114280, 114470.
Other devices under the same FDA submission
50 devices cleared under P010012 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- VIGILANT™ X4 CRT-D G247Boston Scientific Corporation · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)17 deaths2,066 reports
- ACUITY™ X4 Spiral S 4674Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)16 deaths1,975 reports
- ACUITY™ X4 Straight 4671Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)5 deaths1,875 reports
- INOGEN™ X4 CRT-D G148Boston Scientific Corporation · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)8 deaths1,252 reports
- RESONATE™ X4 CRT-D G447Boston Scientific Corporation · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)13 deaths1,218 reports
- DYNAGEN™ X4 CRT-D G158Boston Scientific Corporation · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)13 deaths1,166 reports
- MOMENTUM™ CRT-D G125Boston Scientific Corporation · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)4 deaths792 reports
- MOMENTUM™ CRT-D G124Boston Scientific Corporation · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)5 deaths790 reports
- ACUITY™ X4 Spiral L 4677Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)2 deaths729 reports
- EASYTRAK® 2 IS-1 4543Boston Scientific Corporation · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)5 deaths727 reports
- RESONATE™ HF CRT-D G547Boston Scientific Corporation · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)1 death542 reports
- EASYTRAK® 2 IS-1 4542Boston Scientific Corporation · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)3 deaths536 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 35166).
- Z-0826-06 · TerminatedGuidant CONTAK RENEWAL 4 (models H190, H195) CONTAK RENEWAL; 4 HE (model H197). Cardiac Resynchronization Therapy Defibrillator (CRT-D). Sterile EO. Guidant Corporation, 4100 Ha…
- Z-0827-06 · TerminatedGuidant CONTAK RENEWAL 4 AVT (model M170). Cardiac Resynchronization Therapy Defibrillator (CRT-D). Sterile EO. Guidant Corporation, 4100 Hamline Avenue North, St. Paul, MN 55…
- Z-0828-06 · TerminatedGuidant VITALITY HE Implantable Cardioverter Defibrillator (model T180). Sterile EO. Guidant Corporation, Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 551…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page