CATH, GUIDE 6F JR4.0 SH 110CM Coronary Guiding catheter… recallZ-1019-06 · Medtronic, Inc. · initiated Apr 5, 2006
Terminated All recalls posted in June 2006
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential for non-sterility due to loss of package integrity
FDA root cause category: Other
What the company is doing
Medtronic notified accounts and sales representatives on 4/05/06 advising the recalled product be segregated and held for return to the sales rep. and documented on the reconciliation form.
Recall details
- Recall number
- Z-1019-06
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Apr 5, 2006
- Posted by the FDA
- Jun 1, 2006
- Recalling firm
- Medtronic, Inc.
- Quantity
- 38 units
- Product code
- DQY
- FDA submissions
- K021256
- Distribution
- Nationwide Foreign: Austria, Australia, Bahrain, Bulgaria, Canada, Czech Republic, Denmark, Germany, Hungary, Italy, Jordan, France, Greece, Japan, Hong Kong, India, Italy, Norway, Poland, Portugal, Saudi Arabia, Spain, Sweden, Switzerland, UK.
Product description
CATH, GUIDE 6F JR4.0 SH 110CM Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are introduced into the coronary or peripheral vascular system. Product Code: Z26JR40SHA
Code information (models, lots, serial numbers)
Lot Number: 224554
Other devices under the same FDA submission
50 devices cleared under K021256 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Launcher LA6JR40Medtronic, Inc. · CATHETER, PERCUTANEOUS10 reports
- Launcher LA6IMAKMedtronic, Inc. · CATHETER, PERCUTANEOUS9 reports
- Launcher LA6EBU35Medtronic, Inc. · Catheter, percutaneous9 reports
- Launcher LA6EBU35Medtronic, Inc. · CATHETER, PERCUTANEOUS1 death5 reports
- LAUNCHER LA6EBU35Medtronic, Inc. · Catheter, percutaneous5 reports
- Launcher LA6JR35Medtronic, Inc. · CATHETER, PERCUTANEOUS4 reports
- Launcher LA6JL35Medtronic, Inc. · CATHETER, PERCUTANEOUS4 reports
- Launcher LA6AL75Medtronic, Inc. · CATHETER, PERCUTANEOUS4 reports
- Launcher LA6JL40Medtronic, Inc. · CATHETER, PERCUTANEOUS3 reports
- Launcher LA6AR10Medtronic, Inc. · CATHETER, PERCUTANEOUS3 reports
- Launcher LA6JR35SHMedtronic, Inc. · CATHETER, PERCUTANEOUS2 reports
- Launcher LA6MP1SHMedtronic, Inc. · CATHETER, PERCUTANEOUS2 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 35236).
- Z-0937-06 · TerminatedCATH. GUIDE 6F 3DRC LAUNCHER CATHETER LA63DRC LA 6F 100CM 3DRC Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are intro…
- Z-0938-06 · TerminatedCATH. GUIDE 6F 3DRC W/SH LAUNC CATHETER LA63DRCSH LA 6F 100CM 3DRC Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are …
- Z-0939-06 · TerminatedCATH. GUIDE 6F AL 1.0 110CM LA CATHETER LA6AL10A LA 6F 110CM AL10 Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are i…
- Z-0940-06 · TerminatedCATH. GUIDE 6F AL 2.0 110CM LA CATHETER LA6AL20A LA 6F 110CM AL20 Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are i…
- Z-0941-06 · TerminatedCATH GUIDE 6F LAUNCHER AL 3.0 CATHETER LA6AL30A LA 6F 110CM AL30 Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are in…
- Z-0942-06 · TerminatedCATH. GUIDE 6F EBU3.5 110CM LA CATHETER LA6EBU35A LA 6F 110CM EB35 Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are …
- Z-0943-06 · TerminatedCATH. GUIDE 6F EBU4.0 110CM LA CATHETER LA6EBU40A LA 6F 110CM EB40 Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are …
- Z-0944-06 · TerminatedCATH GUIDE 6F LAUNCHER EBU 4.5 CATHETER LA6EBU45A LA 6F 110CM EB45 Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are …
- Z-0945-06 · TerminatedCATHETER LA6HSIA 6F 110CM HSI Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are introduced into the coronary or perip…
- Z-0946-06 · TerminatedCATH. GUIDE 6F JL3.5 110CM LAU CATHETER LA6JL35A LA 6F 110CM JL35 Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are i…
- Z-0947-06 · TerminatedCATH. GUIDE 6F JL4.0 5 PACK LA Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are introduced into the coronary or peri…
- Z-0948-06 · TerminatedCATH. GUIDE 6F JL4.0 110CM LAU CATHETER LA6JL40A LA 6F 110CM JL40 Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are i…
- Z-0949-06 · TerminatedCATH.GUIDE 6F JL 4.0 SH 110CM CATHETER LA6JL40SHA LA 6F 110CM JL40 Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are …
- Z-0950-06 · TerminatedCATHETER LA6JL45A 6F 110CM JL4 CATHETER LA6JL45A 6F 110CM JL45 Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are intr…
- Z-0951-06 · TerminatedCATH. GUIDE 6F JL5.0 110CM LAU CATHETER LA6JL50A LA 6F 110CM JL50 Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are i…
- Z-0952-06 · TerminatedCATH. GUIDE 6F JR4.0 110CM LAU CATHETER LA6JR40A LA 6F 110CM JR40 Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are i…
- Z-0953-06 · TerminatedCATH.GUIDE 6F JR 4.0 SH 110CM CATHETER LA6JR40SHA LA 6F 110CM JR40 Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are …
- Z-0954-06 · TerminatedCATH. GUIDE 6F JR5.0 110CM LAU CATHETER LA6JR50A LA 6F 110CM JR50 Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are i…
- Z-0955-06 · TerminatedCATHETER LA6MB1A LA 6F 110CM M Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are introduced into the coronary or peri…
- Z-0956-06 · TerminatedCATH. GUIDE 6F SR3.0 LAUNCHER CATHETER LA6SR30 LA 6F 100CM SR30 Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are int…
- Z-0957-06 · TerminatedCATH. GUIDE 6F SR3.0 W/SH LAUN CATHETER LA6SR30SH LA 6F 100CM SR30 Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are …
- Z-0958-06 · TerminatedCATH. GUIDE 6F SR3.5 LAUNCHER CATHETER LA6SR35 LA 6F 100CM SR35 Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are int…
- Z-0959-06 · TerminatedCATH. GUIDE 6F SR3.5 W/SH LAUN CATHETER LA6SR35SH LA 6F 100CM SR35 Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are …
- Z-0960-06 · TerminatedCATH. GUIDE 6F SR4.0 LAUNCHER CATHETER LA6SR40 LA 6F 100CM SR40 Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are int…
- Z-0961-06 · TerminatedCATHETER LA6SR40A 6F 110CM SR4 CATHETER LA6SR40A 6F 110CM SR40 Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are intr…
- Z-0962-06 · TerminatedCATH. GUIDE 6F SR4.0 W/SH LAUN CATHETER LA6SR40SH LA 6F 100CM SR40 Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are …
- Z-0963-06 · TerminatedCATH. GUIDE 6F SR4.5 LAUNCHER CATHETER LA6SR45 LA 6F 100CM SR45 Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are int…
- Z-0964-06 · TerminatedCATH. GUIDE 6F SR4.5 W/SH LAUN CATHETER LA6SR45SH LA 6F 100CM SR45 Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are …
- Z-0965-06 · TerminatedCATH. GUIDE 6F SR5.0 LAUNCHER CATHETER LA6SR50 LA 6F 100CM SR50 Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are int…
- Z-0966-06 · TerminatedCATH. GUIDE 6F SR5.0 W/SH LAUN CATHETER LA6SR50SH LA 6F 100CM SR50 Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page