Skip to content

FastCath Transseptal Catheter Introducers with Hemostasis… recallZ-0535-2007 · St. Jude Medical / Daig Division · initiated Oct 5, 2006

Terminated All recalls posted in March 2007

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from St. Jude Medical / Daig Division or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Limited number of FastCath Transseptal Catheter Introducers (REF #406850) are susceptible to the irrigation extension tube on the side port becoming disconnected when a pull force is applied.

FDA root cause category: Other

What the company is doing

The affected customers were notified on 10/5/06 by letter.

Recall details

Recall number
Z-0535-2007
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Oct 5, 2006
Posted by the FDA
Mar 1, 2007
Recalling firm
St. Jude Medical / Daig Division
Quantity
40 Units
Product code
DYB
FDA submissions
K964518
Distribution
Nationwide.

Product description

FastCath Transseptal Catheter Introducers with Hemostasis Valve, DAIG division of St. Jude Medical, Minnetonka, MN, consists of a radiopaque sheath and dilator; Product # 406850.

Code information (models, lots, serial numbers)

Lot number 1239904

Other devices under the same FDA submission

5 devices cleared under K964518 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.