Medical device recalls: March 2007
The FDA posted 138 medical device recall records in March 2007, from 71 recall events by 61 companies. 138 have not been classified yet.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Boston Scientific Corporation · 3 events, 15 records
- Abbott Laboratories · 3 events, 3 records
- Baxter Healthcare Corp. · 2 events, 7 records
- Terumo Cardiovascular Systems Corp · 2 events, 6 records
- Zimmer Inc. · 2 events, 6 records
- bioMerieux, Inc. · 2 events, 2 records
- Cook, Inc. · 2 events, 2 records
- Intuitive Surgical, Inc. · 2 events, 2 records
FDA root cause
- Other · 119
- Packaging process control · 13
- Software design · 3
- Process design · 1
- Nonconforming Material/Component · 1
- Device Design · 1
Not yet classified (138)
The FDA has posted these records but has not published a class for them yet.
- Z-0685-2007 · NDO Surgical, Inc. · posted Mar 31, 2007 · Terminated
Device arm may fail to open after being deployed and require surgical intervention to remove
- Z-0686-2007 · Aspyra, Inc. · posted Mar 31, 2007 · Terminated
Medications prescribed for one patient were printed on the Medication Administration Record (MAR) of another patient.
- Z-0687-2007 · Boston Scientific Corporation · posted Mar 31, 2007 · Terminated
Defective integrated circuit board could result in loss of the system pump and patient injury (hot fluid 90 degree C into uterus) if in use during the recirculation/heating phase of treatment
- Z-0688-2007 · Boston Scientific Corporation · posted Mar 31, 2007 · Terminated
Defective integrated circuit board could result in loss of the system pump and patient injury (hot fluid 90 degree C into uterus) if in use during the recirculation/heating phase of treatment
- Z-0689-2007 · Boston Scientific Corporation · posted Mar 31, 2007 · Terminated
Defective integrated circuit board could result in loss of the system pump and patient injury (hot fluid 90 degree C into uterus) if in use during the recirculation/heating phase of treatment
- Z-0690-2007 · Boston Scientific Corporation · posted Mar 31, 2007 · Terminated
Defective integrated circuit board could result in loss of the system pump and patient injury (hot fluid 90 degree C into uterus) if in use during the recirculation/heating phase of treatment
- Z-0691-2007 · Boston Scientific Corporation · posted Mar 31, 2007 · Terminated
Defective integrated circuit board could result in loss of the system pump and patient injury (hot fluid 90 degree C into uterus) if in use during the recirculation/heating phase of treatment
- Z-0695-2007 · Gambro Renal Products, Inc. · posted Mar 31, 2007 · Terminated
Under certain conditions, * an excessive amount of anticoagulant may be infused into the patient. (* interruption of a self-test by an alarm and obstruction of the access lines)
- Z-0696-2007 · Boston Scientific Corporation · posted Mar 31, 2007 · Terminated
SpyGlass Probe and Ocular: Ocular is out of specification and during disconnect may damage probe
- Z-0697-2007 · Boston Scientific Corporation · posted Mar 31, 2007 · Terminated
SpyGlass Probe and Ocular: Ocular is out of specification and during disconnect may damage probe
- Z-0698-2007 · Boston Scientific Corporation · posted Mar 31, 2007 · Terminated
Light Degradation in transmission-A film may deposit on the front face and limit transmission of light
- Z-0699-2007 · Boston Scientific Corporation · posted Mar 31, 2007 · Terminated
A teal "bloom"/ glare on the subject image during high light conditions.
- Z-0700-2007 · Advanced Bionics Corp · posted Mar 31, 2007 · Terminated
The Precision Implantable Pulse Generator with firmware Revision 3.01 incorrectly reports an error condition for a valid output from its self internal check.
- Z-0684-2007 · E.M.S. - Electro Medical Systems SA · posted Mar 30, 2007 · Terminated
Handpiece may become dislodged causing cable damage and possible electric shock.
- Z-0704-2007 · GE OEC Medical Systems,Inc · posted Mar 30, 2007 · Terminated
Tracking inaccuracy of 5-10 mm may result when the shunt guide accessory is used with version 5.1 software.
- Z-0661-2007 · Sonosite, Inc. · posted Mar 29, 2007 · Terminated
Laparoscopic transducer exceeds the limits specified for radio-frequency emissions by up to 10db.
- Z-0662-2007 · Smith & Nephew, Inc. Endoscopy Division · posted Mar 29, 2007 · Terminated
Product is non-sterile but labeled incorrectly as sterile.
- Z-0663-2007 · Smith & Nephew, Inc. Endoscopy Division · posted Mar 29, 2007 · Terminated
Product is non-sterile but labeled incorrectly as sterile.
- Z-0664-2007 · Smith & Nephew, Inc. Endoscopy Division · posted Mar 29, 2007 · Terminated
Product is non-sterile but labeled incorrectly as sterile.
- Z-0665-2007 · Cook, Inc. · posted Mar 29, 2007 · Terminated
Mislabeled as to size: The outer label reads 7-4.0 balloon, when the balloon is actually 5-2.0.
- Z-0666-2007 · SuperDimension, Inc. · posted Mar 29, 2007 · Terminated
The action is being taken in response to certain incidents that have been reported to the company, one of which concerned an adverse event in which a patient incurred a pneumothorax during a procedure in which a superDimension Bronchus System was used.…
- Z-0667-2007 · SuperDimension, Inc. · posted Mar 29, 2007 · Terminated
The action is being taken in response to certain incidents that have been reported to the company, one of which concerned an adverse event in which a patient incurred a pneumothorax during a procedure in which a superDimension Bronchus System was used.…
- Z-0668-2007 · Boston Scientific Target · posted Mar 29, 2007 · Terminated
The product may be missing the temperature indicator label on the inner pouch.
- Z-0669-2007 · Becton Dickinson & Co. · posted Mar 29, 2007 · Terminated
Antibiotic susceptibility testing reagent was not manufactured according to specifications and may cause ineffective treatment regimen in patients with infections. The recalling firm received customer reports of false susceptibility results.
- Z-0670-2007 · Intuitive Surgical, Inc. · posted Mar 29, 2007 · Terminated
Da Vinci S Surgical System's spinal pin could limit mechanical motion of the arm and render the system unavailable for surgical use, and may result in system failure and the need to convert to an alternative surgical technique.
- Z-0671-2007 · Medtronic Emergency Response Systems, Inc. · posted Mar 29, 2007 · Terminated
LIFEPAK CR Plus defibrillators with software version 1.16 or lower may miscalculate the HLC battery capacity and result in the CHARGE-PAK battery charger icon/indicator to display and falsely indicate battery needs replacement.
- Z-0672-2007 · Becton Dickinson and Company · posted Mar 29, 2007 · Terminated
Mislabeled: Carton label incorrectly labeled as Ref 581618 (.40 X 16mm) , may contain product Ref 581617 (.50 x 16mm)
- Z-0673-2007 · Remington Medical Inc. · posted Mar 29, 2007 · Terminated
Pouches were not sealed prior to sterilization.
- Z-0675-2007 · Zimmer Inc. · posted Mar 29, 2007 · Terminated
The reamer may lock up on the threads of the reamer body rather than spin freely as intended by design.
- Z-0647-2007 · Zimmer Inc. · posted Mar 28, 2007 · Terminated
The device was cut to an incorrect angle, which may result in an incorrect bone cut.
- Z-0648-2007 · Zimmer Inc. · posted Mar 28, 2007 · Terminated
The device was cut to an incorrect angle, which may result in an incorrect bone cut.
- Z-0649-2007 · Zimmer Inc. · posted Mar 28, 2007 · Terminated
The device was cut to an incorrect angle, which may result in an incorrect bone cut.
- Z-0650-2007 · Zimmer Inc. · posted Mar 28, 2007 · Terminated
The device was cut to an incorrect angle, which may result in an incorrect bone cut.
- Z-0651-2007 · Zimmer Inc. · posted Mar 28, 2007 · Terminated
The device was cut to an incorrect angle, which may result in an incorrect bone cut.
- Z-0652-2007 · Siemens Medical Solutions, USA, Inc · posted Mar 28, 2007 · Terminated
The cardiac calculation feature may incorrectly calculate a derived V max value (reported out lower than it should be) and certain values imported into worksheets and reports may be incorrect due to a software defect.
- Z-0660-2007 · Intuitive Surgical, Inc. · posted Mar 28, 2007 · Terminated
The products were incorrectly programmed as training instruments, which allows for 30 uses (surgical procedures) instead of 10 uses.
- Z-0653-2007 · Medtronic Sofamor Danek USA Inc · posted Mar 27, 2007 · Terminated
Bone screws may pull though the plate.
- Z-0654-2007 · Medtronic Sofamor Danek USA Inc · posted Mar 27, 2007 · Terminated
Bone screws may pull though the plate.
- Z-0655-2007 · Terumo Cardiovascular Systems Corp · posted Mar 27, 2007 · Terminated
The hand crank may crack in the bearing/shaft area due to a manufacturing defect. (Note: Some of the hand cranks subject to this recall were shipped as replacements for hand cranks recalled in the Summer of 2006.)
- Z-0656-2007 · Terumo Cardiovascular Systems Corp · posted Mar 27, 2007 · Terminated
The hand crank may crack in the bearing/shaft area due to a manufacturing defect. (Note: Some of the hand cranks subject to this recall were shipped as replacements for hand cranks recalled in the Summer of 2006.)
- Z-0657-2007 · Terumo Cardiovascular Systems Corp · posted Mar 27, 2007 · Terminated
The hand crank may crack in the bearing/shaft area due to a manufacturing defect. (Note: Some of the hand cranks subject to this recall were shipped as replacements for hand cranks recalled in the Summer of 2006.)
- Z-0658-2007 · Terumo Cardiovascular Systems Corp · posted Mar 27, 2007 · Terminated
The hand crank may crack in the bearing/shaft area due to a manufacturing defect. (Note: Some of the hand cranks subject to this recall were shipped as replacements for hand cranks recalled in the Summer of 2006.)
- Z-0659-2007 · Terumo Cardiovascular Systems Corp · posted Mar 27, 2007 · Terminated
The hand crank may crack in the bearing/shaft area due to a manufacturing defect. (Note: Some of the hand cranks subject to this recall were shipped as replacements for hand cranks recalled in the Summer of 2006.)
- Z-0556-2007 · GE OEC Medical Systems, Inc · posted Mar 22, 2007 · Terminated
During routine service, a cover may have been installed without the required proper lead shielding.
- Z-0557-2007 · GE OEC Medical Systems, Inc · posted Mar 22, 2007 · Terminated
During routine service, a cover may have been installed without the required proper lead shielding.
- Z-0558-2007 · GE OEC Medical Systems, Inc · posted Mar 22, 2007 · Terminated
During routine service, a cover may have been installed without the required proper lead shielding.
- Z-0559-2007 · GE OEC Medical Systems, Inc · posted Mar 22, 2007 · Terminated
During routine service, a cover may have been installed without the required proper lead shielding.
- Z-0641-2007 · Philips Medical Systems · posted Mar 22, 2007 · Terminated
Failure of tactile switches on one of two printed circuit board assemblies (PCBA) used in the FR2+ device. Potential delay or prevention of shock treatment due to PCBA board contamination, causing intermittent switch operation.
- Z-0643-2007 · bioMerieux, Inc. · posted Mar 22, 2007 · Terminated
Bacillus sp. contamination was detected in inoculated and uninoculated bottles of BacT/ALERT FA Culture bottles.
- Z-0644-2007 · bioMerieux, Inc. · posted Mar 22, 2007 · Terminated
Bottle data is not sent to BacT/ALERT 3D instrument.
- Z-0645-2007 · Siemens Healthcare Diagnostics · posted Mar 22, 2007 · Terminated
Firm received complaints for signal 2 & 4 errors, and shifts in Relative Light Units (RLU). Investigations showed that the On-Board Stability (OBS) of the acid & base reagents may be compromised after 4 days after installation onto the Centaur CP system…
- Z-0646-2007 · ThyroTec Inc · posted Mar 22, 2007 · Terminated
Positive control showed little or no positive line.
- Z-0622-2007 · HoMedics, Inc. · posted Mar 21, 2007 · Terminated
Heating pads may have a loose electrical connection, which could result in excessive heat being generated that poses a risk of patient burns, fire, and property damage.
- Z-0623-2007 · HoMedics, Inc. · posted Mar 21, 2007 · Terminated
Heating pads may have a loose electrical connection, which could result in excessive heat being generated that poses a risk of patient burns, fire, and property damage.
- Z-0624-2007 · HoMedics, Inc. · posted Mar 21, 2007 · Terminated
Heating pads may have a loose electrical connection, which could result in excessive heat being generated that poses a risk of patient burns, fire, and property damage.
- Z-0625-2007 · HoMedics, Inc. · posted Mar 21, 2007 · Terminated
Heating pads may have a loose electrical connection, which could result in excessive heat being generated that poses a risk of patient burns, fire, and property damage.
- Z-0626-2007 · HoMedics, Inc. · posted Mar 21, 2007 · Terminated
Heating pads may have a loose electrical connection, which could result in excessive heat being generated that poses a risk of patient burns, fire, and property damage.
- Z-0642-2007 · Boston Scientific · posted Mar 21, 2007 · Terminated
Boston Scientific is voluntarily recalling one lot/batch of Katzen Infusion Wires because they have identified that the label on the carton may indicate a different length device than what is actually in the carton. The affected batch was manufactured…
- Z-0497-2007 · Gebauer Company · posted Mar 20, 2007 · Terminated
Microbial Contamination-certain lots of product failed USP <61> Microbial Limits Testing for total aerobic count during 6 month stability testing.
- Z-0632-2007 · Baxter Healthcare Corp. · posted Mar 20, 2007 · Terminated
There is the potential for non-detection of, and no alarm for, upstream occlusion for Flo-Gard Volumetric Infusion Pumps, if an occlusion occurs above a flexible chamber such as a drip chamber, Buretrol or blood filter.
- Baxter Flo-Gard 6301 Dual Channel Volumetric Infusion Pump, catalog #2M8064, 2M8064R (refurbished) ↗Z-0633-2007 · Baxter Healthcare Corp. · posted Mar 20, 2007 · Terminated
There is the potential for non-detection of, and no alarm for, upstream occlusion for Flo-Gard Volumetric Infusion Pumps, if an occlusion occurs above a flexible chamber such as a drip chamber, Buretrol or blood filter.
- Z-0634-2007 · Baxter Healthcare Corp. · posted Mar 20, 2007 · Terminated
There is the potential for non-detection of, and no alarm for, upstream occlusion for Flo-Gard Volumetric Infusion Pumps, if an occlusion occurs above a flexible chamber such as a drip chamber, Buretrol or blood filter.
- Baxter Flo-Gard 6300 Dual Channel Volumetric Infusion Pump, catalog #2M8048, 2M8048R (refurbished) ↗Z-0635-2007 · Baxter Healthcare Corp. · posted Mar 20, 2007 · Terminated
There is the potential for non-detection of, and no alarm for, upstream occlusion for Flo-Gard Volumetric Infusion Pumps, if an occlusion occurs above a flexible chamber such as a drip chamber, Buretrol or blood filter.
- Z-0638-2007 · Viasys Respiratory Care, Inc.dba Bird Products · posted Mar 20, 2007 · Terminated
The graphical user interface may become slow in its response and in some cases non-responsive. Under some circumstances this situation may cause the ventilator to delay the annunciation of alarms.
- Z-0636-2007 · Philips Medical Systems (Cleveland) Inc · posted Mar 17, 2007 · Terminated
Two conditions have been identified with the Philips Medical System Brilliance Bore ,version 2.2.1 Tumor Localization, that may lead the user to misinterpret the displayed data.
- Z-0640-2007 · Abbott Laboratories · posted Mar 17, 2007 · Terminated
Incorrect Test Results-product may show a higher than expected platelet background count when used on the CELL-DYN 4000 System and report patient results that are unacceptable (out-of-range).
- Z-0611-2007 · Guidant Cardiac Surgery · posted Mar 16, 2007 · Terminated
Vacuum Tubing Failure- vacuum tubing fails before the labeled two year shelf life. This may result in partial or complete vacuum loss during surgery.
- Z-0612-2007 · Guidant Cardiac Surgery · posted Mar 16, 2007 · Terminated
Vacuum Tubing Failure- vacuum tubing fails before the labeled two year shelf life. This may result in partial or complete vacuum loss during surgery.
- Z-0637-2007 · Cook, Inc. · posted Mar 16, 2007 · Terminated
These catheters do not include the proximal sidehole as required by the specification.
- Z-0639-2007 · Abbott Laboratories · posted Mar 16, 2007 · Terminated
Incorrect Test Results-test results may show a higher than expected platelet background count when used on the CELL-DYN 4000 System and report patient results that are unacceptable (out-of-range).
- Z-0627-2007 · Merit Medical Systems, Inc · posted Mar 15, 2007 · Terminated
Sterility of some units may be compromised due to damaged packaging.
- Z-0628-2007 · Merit Medical Systems, Inc · posted Mar 15, 2007 · Terminated
Sterility of some units may be compromised due to damaged packaging.
- Z-0629-2007 · Merit Medical Systems, Inc · posted Mar 15, 2007 · Terminated
Sterility of some units may be compromised due to damaged packaging.
- Z-0630-2007 · Boston Scientific Corp · posted Mar 15, 2007 · Terminated
Sterile barrier may have been compromised. Firm received one complaint which demonstrated that the inner and outer seals of the port tray packaging had been compromised.
- Z-0631-2007 · Boston Scientific Corp · posted Mar 15, 2007 · Terminated
Sterile barrier may have been compromised. Firm received one complaint which demonstrated that the inner and outer seals of the port tray packaging had been compromised.
- Z-0617-2007 · Terumo Cardiovascular Systems Corp · posted Mar 14, 2007 · Terminated
Internal friction may cause the centrifugal manual drive to be inoperable when a disposable pump is attached.
- Z-0618-2007 · Baxter Healthcare Corp. · posted Mar 14, 2007 · Terminated
Baxter has received reports of disconnections/leaks during use of vented Paclitaxel administration sets.
- Z-0619-2007 · Baxter Healthcare Corp. · posted Mar 14, 2007 · Terminated
Baxter has received reports of disconnections/leaks during use of vented Paclitaxel administration sets.
- Z-0620-2007 · Baxter Healthcare Corp. · posted Mar 14, 2007 · Terminated
Baxter has received reports of disconnections/leaks during use of vented Paclitaxel administration sets.
- Z-0621-2007 · Remel, Inc. · posted Mar 14, 2007 · Terminated
False negative results for sorbitol fermenting bacteria such as E. Coli 0157.
- Z-0584-2007 · Conmed Corporation · posted Mar 13, 2007 · Terminated
Firm was made aware of instances where the sterile barrier of the instruments had been compromised.
- Z-0585-2007 · Conmed Corporation · posted Mar 13, 2007 · Terminated
Firm was made aware of instances where the sterile barrier of the instruments had been compromised.
- Z-0586-2007 · Conmed Corporation · posted Mar 13, 2007 · Terminated
Firm was made aware of instances where the sterile barrier of the instruments had been compromised.
- Z-0587-2007 · Conmed Corporation · posted Mar 13, 2007 · Terminated
Firm was made aware of instances where the sterile barrier of the instruments had been compromised.
- Z-0588-2007 · Conmed Corporation · posted Mar 13, 2007 · Terminated
Firm was made aware of instances where the sterile barrier of the instruments had been compromised.
- Z-0589-2007 · Conmed Corporation · posted Mar 13, 2007 · Terminated
Firm was made aware of instances where the sterile barrier of the instruments had been compromised.
- Z-0590-2007 · Conmed Corporation · posted Mar 13, 2007 · Terminated
Firm was made aware of instances where the sterile barrier of the instruments had been compromised.
- Z-0591-2007 · Conmed Corporation · posted Mar 13, 2007 · Terminated
Firm was made aware of instances where the sterile barrier of the instruments had been compromised.
- Z-0592-2007 · Conmed Corporation · posted Mar 13, 2007 · Terminated
Firm was made aware of instances where the sterile barrier of the instruments had been compromised.
- Z-0593-2007 · Conmed Corporation · posted Mar 13, 2007 · Terminated
Firm was made aware of instances where the sterile barrier of the instruments had been compromised.
- Z-0594-2007 · Inverness Medical Professional Diagnostics · posted Mar 13, 2007 · Terminated
Lot PX018A is exhibiting a decrease in sensitivity affecting the qualitative result around the cut off.
- Z-0595-2007 · Abbott Molecular · posted Mar 13, 2007 · Terminated
The LSI D13S319 SO/13q34 SA/CEP 12 SG probe of the two mixture DNA-Probe was contaminated with LSI ATM SG/p53 SO probe. This contamination would result in 10 FISH signals in each normal cell rather than the expected 6 FISH signals in each normal cell.
- Z-0604-2007 · Boston Scientific Corporation · posted Mar 13, 2007 · Terminated
Guidewire restriction during use (excessive adhesive in the guidewire lumen)
- Z-0605-2007 · Boston Scientific Corporation · posted Mar 13, 2007 · Terminated
Guidewire restriction during use (excessive adhesive in the guidewire lumen)
- Z-0606-2007 · Boston Scientific Corporation · posted Mar 13, 2007 · Terminated
Guidewire restriction during use (excessive adhesive in the guidewire lumen)
- Z-0607-2007 · Boston Scientific Corporation · posted Mar 13, 2007 · Terminated
Guidewire restriction during use (excessive adhesive in the guidewire lumen)
- Z-0608-2007 · Boston Scientific Corporation · posted Mar 13, 2007 · Terminated
Guidewire restriction during use (excessive adhesive in the guidewire lumen)
- Z-0609-2007 · Boston Scientific Corporation · posted Mar 13, 2007 · Terminated
Guidewire restriction during use (excessive adhesive in the guidewire lumen)
- Z-0610-2007 · Alcon Research, Ltd · posted Mar 13, 2007 · Terminated
Product is misbranded; 19 gauge Schlera Plugs distributed in containers labeled as 19 gauge Trochar Canula Plugs.
- Z-0613-2007 · Gyrus ACMI Corporation · posted Mar 13, 2007 · Terminated
Product sterility may be compromised due to failure of the tray seal packaging
- Z-0614-2007 · Gyrus ACMI Corporation · posted Mar 13, 2007 · Terminated
Product sterility may be compromised due to failure of the tray seal packaging
- Z-0615-2007 · Gyrus ACMI Corporation · posted Mar 13, 2007 · Terminated
Product sterility may be compromised due to failure of the tray seal packaging
- Z-0616-2007 · Stentor Inc A Phillips Medical Systems Co. · posted Mar 13, 2007 · Terminated
A defect may cause patient image orientation markers to be incorrectly labeled on Multi-Planar Reformation (MPR) images. Left and Right Anatomical orientation markers may be reversed.
- Z-0413-2007 · Abbott Laboratories MPG · posted Mar 8, 2007 · Terminated
The ARCHITECT AUSAB and ARCHITECT Anti-HCV assay parameter default interpretation screens when using ARCHITECT Assay CD-ROM US Version 06E58-21 does not align with the result interpretation options in the ARCHITECT AUSAB Reagent Package Insert (PI)…
- Z-0575-2007 · Abbott Vascular, Inc. · posted Mar 8, 2007 · Terminated
Premature release of the vessel locator wings, which stabilize the device prior to clip deployment, will result in no hemostasis.
- Z-0577-2007 · Ekos Corp · posted Mar 8, 2007 · Terminated
Ekos Corporation received three reports that distal radiopaque marker bands on the EndoWave Drug Delivery Catheter became detached from the catheter during use and remained in the patient.
- Z-0578-2007 · Siemens Medical Solutions USA, Inc · posted Mar 8, 2007 · Terminated
C-arm gantry could rapidly descend without command during use.
- Z-0579-2007 · Siemens Medical Solutions USA, Inc · posted Mar 8, 2007 · Terminated
C-arm gantry could rapidly descend without command during use
- Z-0580-2007 · Defibtech, LLC · posted Mar 8, 2007 · Terminated
Device Malfunction-The self-test software may allow a self-test to clear a previously detected low battery condition. If this occurs, the operator may be unaware of the low battery and the device may be "unable" to deliver a defibrillation shock, which…
- Z-0581-2007 · Defibtech, LLC · posted Mar 8, 2007 · Terminated
Device Malfunction-The self-test software may allow a self-test to clear a previously detected low battery condition. If this occurs, the operator may be unaware of the low battery and the device may be "unable" to deliver a defibrillation shock, which…
- Z-0583-2007 · Medtronic Perfusion Systems · posted Mar 8, 2007 · Terminated
One lot of Custom Perfusion Sets, Catalog 2T10R1 was incorrectly assembled. The one way vacuum relief valve on the 102 inch line is attached backwards, thus preventing flow.
- Z-0596-2007 · J. Jamner Surgical Instruments, Inc. · posted Mar 8, 2007 · Terminated
The JariTrak Table Clamp may not securely fasten to the bed rail of the OR table during set up of the retractor system. In the event that the clamp is not securely fastened to the bed rail, the system may move during surgery. The firm received several…
- Z-0597-2007 · Kensey Nash Corp · posted Mar 8, 2007 · Terminated
Face seal may wear excessively and cause particulate matter.
- Z-0567-2007 · Wright Medical Technology Inc · posted Mar 7, 2007 · Terminated
Product specification failure which could result in improper locking of the cup onto the femoral head.
- Z-0568-2007 · Wright Medical Technology Inc · posted Mar 7, 2007 · Terminated
Product specification failure which could result in improper locking of the cup onto the femoral head.
- Z-0569-2007 · Wright Medical Technology Inc · posted Mar 7, 2007 · Terminated
Product specification failure which could result in improper locking of the cup onto the femoral head.
- Z-0570-2007 · Wright Medical Technology Inc · posted Mar 7, 2007 · Terminated
Product specification failure which could result in improper locking of the cup onto the femoral head.
- Z-0571-2007 · Wright Medical Technology Inc · posted Mar 7, 2007 · Terminated
Product specification failure which could result in improper locking of the cup onto the femoral head.
- Z-0572-2007 · Wright Medical Technology Inc · posted Mar 7, 2007 · Terminated
Product specification failure which could result in improper locking of the cup onto the femoral head.
- Z-0573-2007 · Wright Medical Technology Inc · posted Mar 7, 2007 · Terminated
Product specification failure which could result in improper locking of the cup onto the femoral head.
- Z-0574-2007 · Abiomed, Inc. · posted Mar 7, 2007 · Terminated
Unit may alarm ''Low Flow'' and ''Low Pressure'' due to manufacturing material in the pressure source.
- Z-0576-2007 · Medtronic Xomed, Inc. · posted Mar 7, 2007 · Terminated
A labeling discrepancy was internally identified for the surface electrode component of REF 8263210 AccuGuide EMG Injection Needle, 30G X 25MM. The outside box label correctly indicates 'Content: 5 sterile Injection Needles and 10 non-sterile Surface…
- Coulter LH700 Series Hematology Analyzers, Part Number 6605632 with 2B Software (2B3, 2B4 and 2B5) ↗Z-0566-2007 · Beckman Coulter Inc · posted Mar 6, 2007 · Terminated
There is a potential for erroneous yet credible results for Body Fluids on the LH700 series hematology analyzers when a cassette label fails to read and the subsequent sample is cycled in Body Fluid mode.
- Z-0547-2007 · Safety & Supplies Co. Inc. · posted Mar 3, 2007 · Terminated
Counterfeit product (manufacturer unknown)
- Z-0582-2007 · Becton Dickinson & Company · posted Mar 3, 2007 · Terminated
The needle becomes detached from the hub/syringe
- Z-0344-2007 · Advanced Medical Optics, Inc. · posted Mar 1, 2007 · Terminated
Due to a potential sterility issue with the product container. Three lots sold in Japan were found to have bacterial contamination, which compromised sterility. Because of this production-line issue at the firm's plant in China, AMO is recalling lots that…
- Z-0345-2007 · Advanced Medical Optics, Inc. · posted Mar 1, 2007 · Terminated
Due to a potential sterility issue with the product container. Three lots sold in Japan were found to have bacterial contamination, which compromised sterility. Because of this production-line issue at the firm's plant in China, AMO is recalling lots that…
- Z-0346-2007 · Advanced Medical Optics, Inc. · posted Mar 1, 2007 · Terminated
Due to a potential sterility issue with the product container. Three lots sold in Japan were found to have bacterial contamination, which compromised sterility. Because of this production-line issue at the firm''s plant in China, AMO is recalling lots that…
- Z-0347-2007 · Advanced Medical Optics, Inc. · posted Mar 1, 2007 · Terminated
Due to a potential sterility issue with the product container. Three lots sold in Japan were found to have bacterial contamination, which compromised sterility. Because of this production-line issue at the firm''s plant in China, AMO is recalling lots that…
- Z-0348-2007 · Advanced Medical Optics, Inc. · posted Mar 1, 2007 · Terminated
Due to a potential sterility issue with the product container. Three lots sold in Japan were found to have bacterial contamination, which compromised sterility. Because of this production-line issue at the firm''s plant in China, AMO is recalling lots that…
- Z-0535-2007 · St. Jude Medical / Daig Division · posted Mar 1, 2007 · Terminated
Limited number of FastCath Transseptal Catheter Introducers (REF #406850) are susceptible to the irrigation extension tube on the side port becoming disconnected when a pull force is applied.
- Z-0545-2007 · Siemens Medical Solutions Diagnostics · posted Mar 1, 2007 · Terminated
Under limited circumstances, an incorrect patient result could be printed on the optional chartable patient report.
- Z-0560-2007 · Alcon Laboratories, Inc · posted Mar 1, 2007 · Terminated
Individually packaged surgical blades, labeled as sterile, found with incomplete seals; blades purchased for use in custom surgical packs.
- Z-0561-2007 · Alcon Laboratories, Inc · posted Mar 1, 2007 · Terminated
Individually packaged surgical blades, labeled as sterile, found with incomplete seals; blades purchased for use in custom surgical packs.
- Z-0562-2007 · Alcon Laboratories, Inc · posted Mar 1, 2007 · Terminated
Individually packaged surgical blades, labeled as sterile, found with incomplete seals; blades purchased for use in custom surgical packs.
- Z-0563-2007 · Alcon Laboratories, Inc · posted Mar 1, 2007 · Terminated
Individually packaged surgical blades, labeled as sterile, found with incomplete seals; blades purchased for use in custom surgical packs.
- Z-0564-2007 · Alcon Laboratories, Inc · posted Mar 1, 2007 · Terminated
Individually packaged surgical blades, labeled as sterile, found with incomplete seals; blades purchased for use in custom surgical packs.
- Z-0565-2007 · Abbott Laboratories · posted Mar 1, 2007 · Terminated
Results for tests of RBC, MCV, RDW, MPV and PLT may be falsely elevated or falsely decreased when the Cell Dyn Diluent/Sheath reagent is used with Cell-Dyn 4000, Ruby and Sapphire systems.