Advanced Bionics Precision Linear Leads, part of the… recallZ-0299-2007 · Advanced Bionics Corporation · initiated Sep 21, 2006
Terminated All recalls posted in January 2007
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Advanced Bionics Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
A small number of unimplanted Precision Linear Leads may have been assembled with incorrect electrode material. At high stimulation levels, the metal may corrode and dissolved metals may enter the patient.
FDA root cause category: Other
What the company is doing
In United States, the clinician recall notification letters were distributed by Fed-Ex or U.S. Postal Service certified mail on 09/21/2006. For the other regions, the in-country representatives started providing notification to the clinicians on 09/21/2006. The recall notification letters will instruct physicians to immediately return the unimplanted units.
Recall details
- Recall number
- Z-0299-2007
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Sep 21, 2006
- Posted by the FDA
- Jan 10, 2007
- Recalling firm
- Advanced Bionics Corporation
- Quantity
- 18
- Product code
- LGW
- FDA submissions
- P030017
- Distribution
- Worldwide.
Product description
Advanced Bionics Precision Linear Leads, part of the Precision Spinal Cord Stimulation System. Model Numbers: SC-2138-30, SC-2138-50, SC-2138-50T and SC-2138-70.
Code information (models, lots, serial numbers)
Serial Numbers: 111744, 112122, 112141, 112224, 112233, 114114, 116547, 116553, 117860, 117907, 110490, 100600, 110625, 110809, 113745, 113780, 112055 and 118943
Other devices under the same FDA submission
94 devices cleared under P030017 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Precision Spectra ™ SC-1132Boston Scientific Neuromodulation Corporation · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF53 deaths11,269 reports
- Spectra WaveWriter™ SC-1160Boston Scientific Neuromodulation Corporation · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF31 deaths7,221 reports
- Precision™ SC-1110-02Boston Scientific Neuromodulation Corporation · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF26 deaths6,519 reports
- WaveWriter Alpha™ SC-1232Boston Scientific Neuromodulation Corporation · Stimulator, spinal-cord, totally implanted for pain relief1 recall41 deaths4,347 reports
- Precision™ Montage™ MRI SC-1200Boston Scientific Neuromodulation Corporation · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF22 deaths4,316 reports
- Linear™ ST SC-2218-50Boston Scientific Neuromodulation Corporation · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2,729 reports
- WaveWriter Alpha™ 16 SC-1216Boston Scientific Neuromodulation Corporation · Stimulator, spinal-cord, totally implanted for pain relief1 recall29 deaths2,190 reports
- Infinion™ CX SC-2317-70Boston Scientific Neuromodulation Corporation · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 recall1 death1,268 reports
- Linear™ ST SC-2218-70Boston Scientific Neuromodulation Corporation · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1,175 reports
- Precision Novi™ SC-1140Boston Scientific Neuromodulation Corporation · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF10 deaths1,001 reports
- Infinion™ 16 SC-2316-50EBoston Scientific Neuromodulation Corporation · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF15 deaths978 reports
- WaveWriter Alpha™ Prime SC-1432Boston Scientific Neuromodulation Corporation · Stimulator, spinal-cord, totally implanted for relief of pain due to diabetic neuropathy7 deaths971 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.