Electrode Belt, component of the LifeVest Wearable… recallZ-0936-2007 · Zoll Lifecor Corporation · initiated Apr 27, 2007

Terminated All recalls posted in June 2007

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Zoll Lifecor Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The electrodes do not properly release conductive gel

FDA root cause category: Other

What the company is doing

Patients currently using the product should continue to do so while the product removal is underway. Product replacements will begin with the belts currently in the possession of the sales reps. By replacing these first any new patients will be receiving new or refurbished belts with new Therapy Electrodes that have not been exposed to methanol. As patients complete their use of the LifeVest and return their device to the firm the belt's refurbishment process will include replacement of all 3 Therapy Electrodes. The firm expects this process to be completed in approximately 3 months based on the average use of time and current production capabilities. The process may take an additional month for devices distributed overseas. Physicians with active patients were notified via letter sent certified mail on 5/3/07. Long term use patients (patients using the device longer than 3 months) will be notified of the product removal via telephone and follow up letter. Each long term patient will be provided with a replacement belt. They will be instructed to return their existing belt in the same shipping box. Due to the low residual risk no notification is planned for short term use patients. Prescribing physicians have the option to notify their patients if they choose to do so.

Recall details

Recall number
Z-0936-2007
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Apr 27, 2007
Posted by the FDA
Jun 15, 2007
Recalling firm
Zoll Lifecor Corporation
Quantity
944 belts
Product code
MVK
FDA submissions
P010030
Distribution
Worldwide: The product was shipped to patients in AZ, CA, CT, FL, GA, IL, IN, KY, LA, MA, MI, MN, MO, MT, NC, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WI, and WV. The product was also shipped to Germany, United Kingdom, Italy, Spain, France, and Denmark.

Product description

Electrode Belt, component of the LifeVest Wearable Defibrillator, Product number 10A0889-A01. Specifically the 3 Therapy Electrode Pads attached to the Electrode Belt.

Code information (models, lots, serial numbers)

Show the full code information (10k characters)

Product number 10A0889-A01. Serial numbers with active patients: 79000004, 79000055, 79000064, 79000073, 79000106, 79000157, 79000181, 79000205, 79000208, 79000217, 79000226, 79000238, 79000244, 79000259, 79000274, 79000277, 79000286, 79000289, 79000298, 79000328, 79000346, 79000358, 79000364, 79000367, 79000373, 79000376, 79000379, 79000388, 79000403, 79000409, 79000445, 79000454, 79000457, 79000466, 79000487, 79000490, 79000493, 79000508, 79000511, 79000529, 79000532, 79000547, 79000556, 79000562, 79000574, 79000577, 79000595, 79000598, 79000601, 79000604, 79000616, 79000631, 79000634, 79000637, 79000640, 79000652, 79000661, 79000670, 79000686, 79000685, 79000694, 79000697, 79000715, 79000727, 79000730, 79000739, 79000754, 79000760, 79000769, 79000790, 79000796, 79000799, 79000802, 79000811, 79000823, 79000829, 79000832, 79000835, 79000838, 79000841, 79000844, 79000847, 79000853, 79000856, 79000886, 79000889, 79000901, 79000904, 79000925, 79000931, 79000934, 79000937, 79000943, 79000958, 79000973, 79000979, 79000982, 79000994, 79001003, 79001018, 79001021, 79001024, 79001027, 79001030, 79001033, 79001036, 79001042, 79001048, 79001054, 79001060, 79001066, 79001078, 79001081, 79001093, 79001099, 79001102, 79001111, 79001117, 79001147, 79001150, 79001153, 79001156, 79001162, 79001198, 79001201, 79001207, 79001216, 79001219, 79001228, 79001237, 79001243, 79001255, 79001273, 79001276, 79001288, 79001300, 79001318, 79001339, 79001354, 79001357, 79001363, 79001369, 79001372, 79001375, 79001378,79001384, 79001387, 79001399, 79001402, 79001405, 79001408, 79001417, 79001423, 79001429, 79001441, 79001444, 79001456, 79001459, 79001462, 79001468, 79001486, 79001492, 79001498, 79001501, 79001504, 79001522, 79001525, 79001528, 79001534, 79001537, 79001540, 79001558, 79001564, 79001570, 79001573, 79001582, 79001585, 79001588, 79001594, 79001597, 79001603, 79001630 79001639, 79001642, 79001648, 79001654, 79001657, 79001660, 79001666, 79001672, 79001675, 79001681, 79001690, 79001693, 79001699, 79001714, 79001717, 79001732, 79001738, 79001741, 79001744, 79001750, 79001753, 79001756, 79001759, 79001762, 79001765, 79001771, 79001777, 79001780, 79001786, 79001795, 79001798, 79001801, 79001804, 79001810, 79001822, 79001834, 79001837, 79001840, 79001846, 79001849, 79001852, 79001867, 79001879, 79001882, 79001906, 79001912, 79001918, 79001924, 79001939, 79001942, 79001945, 79001954, 79001957, 79001963, 79001969, 79001984, 79001996, 79002002, 79002005, 79002011, 79002017, 79002026, 79002041, 79002044, 79002047, 79002053, 79002056, 79002059, 79002068, 79002071, 79002074, 79002089, 79002110, 79002116, 79002119, 79002122, 79002128, 79002140, 79002152, 79002155, 79002164, 79002170, 79002176, 79002179, 79002182, 79002185, 79002191, 79002194, 79002200, 79002206, 79002215, 79002218, 79002221, 79002224, 79002227, 79002233, 79002236, 79002248, 79002251, 79002263, 79002275, 79002278, 79002287, 79002290, 79002293, 79002305, 79002308, 79002320, 79002329, 79002332, 79002338, 79002347, 79002359, 79002368, 79002377, 79002392, 79002404, 79002413, 79002416, 79002422, 79002425, 79002428, 79002440, 79002446, 79002449, 79002452, 79002455, 79002458, 79002479, 79002482, 79002494, 79002497, 79002509, 79002512, 79002521, 79002524, 79002530, 79002533, 79002539, 79002554, 79002560, 79002563, 79002566, 79002578, 79002587, 79002593, 79002602, 79002605, 79002608, 79002617, 79002620, 79002623, 79002629, 79002635, 79002644, 79002647, 79002650, 79002659, 79002665, 79002671, 79002674, 79002677, 79002698, 79002701, 79002710, 79002728, 79002740, 79002749, 79002761, 79002764, 79002767, 79002770, 79002776, 79002779, 79002782, 79002785, 79002788, 79002794, 79002800, 79002806, 79002818, 79002827, 79002833, 79002842, 79002848, 79002860, 79002863, 79002866, 79002869, 79002875, 79002878, 79002881, 79002884, 79002896, 79002899, 79002902, 79002911, 79002917, 79002932, 79002938, 79002941, 79002950, 79002962, 79002965, 79002968, 79002971, 79002974, 79002977, 79002980, 79002983, 79002989, 79002992, 79003007, 79003010, 79003016, 79003019, 79003028, 79003031, 79003034, 79003037, 79003049, 79003052, 79003055, 79003058, 79003067, 79003070, 79003079, 79003082, 79003088, 79003097, 79003103, 79003118, 79003121, 79003130, 79003133, 79003136, 79003142, 79003151, 79003157, 79003166, 79003172, 79003175, 79003178, 79003190, 79003196, 79003199, 79003205, 79003211, 79003223, 79003229, 79003235, 79003244, 79003262, 79003265, 79003274, 79003277, 79003295, 79003316, 79003325, 79003328, 79003331, 79003346, 79003349, 79003373, 79003382, 79003400, 79003415, 79003445, 79003454, 79003457, 79003460, 79003463, 79003469, 79003472, 79003478, 79003493, 79003508, 79003511, 79003520, 79003526, 79003529, 79003538, 79003544, 79003550, 79003553, 79003556, 79003568, 79003583, 79003589, 79003592, 79003598, 79003601, 79003604, 79003607, 79003613, 79003616, 79003628, 79003631, 79003634, 79003637, 79003640, 79003646, 79003649, 79003652, 79003658, 79003661, 79003664, 79003670, 79003673, 79003697, 79003700, 79003703, 79003706, 79003715, 79003718, 79003721, 79003727, 79003730, 79003748, 79003757, 79003760, 79003763, 79003766, 79003769, 79003778, 79003784, 79003793, 79003796, 79003808, 79003823, 79003835, 79003838, 79003841, 79003844, 79003847, 79003850, 79003856, 79003862, 79003868, 79003871, 79003877, 79003889, 79003916, 79003919, 79003922, 79003949, 79003958, 79003961, 79003967, 79003976, 79003985, 79003988, 79004000, 79004003, 79004012, 79004021, 79004027, 79004042, 79004066, 79004078, 79004087, 79004096, 79004108, 79004117, 79004120, 79004126, 79004129, 79004135, 79004141, 79004144, 79004153, 79004162, 79004165, 79004210, and 79004216. Foreign serial numbers with active patients: 79000118, 79000724, 79000736, 79000343, 79000664, 79001075, 79001213, 79000310, 79000394, 79000139, 79000187, 79001231, 79001600, 79000256, 79000370, 79000463, 79000514, 79000592, 79000703, 79000745, 79000766, 79000808, 79001090, 79001171, 79001177, 79001315, 79001336, 79001345, 79001351, 79001390, 79001615, 79001825, 79001870, 79001900, 79001960, 79001966, 79002008, 79002062, 79002149, 79002203, 79002230, 79002272, 79002398, 79002464, 79002488, 79002491, 79002500, 79002545, 79002551, 79002584, 79002590, 79002611, 79002614, 79002680, 79002812, 79002953, 79003013, 79003022, 79003025, 79003094, 79003100, 79003124, 79003127, 79003139, 79003232, 79003247, 79003319, and 79003355. Serial numbers still under the firm''s control: 79000058, 79000190, 79000196, 79000211, 79000214, 79000241, 79000247, 79000253, 79000283, 79000325, 79000334, 79000340, 79000352, 79000391, 79000427, 79000439, 79000448, 79000475, 79000541, 79000544, 79000589, 79000643, 79000667, 79000712, 79000721, 79000772, 79000781, 79000784, 79000793, 79000814, 79000826, 79000850, 79000859, 79000862, 79000865, 79000874, 79000880, 79000895, 79000910, 79000913, 79000952, 79000961, 79000964, 79000985, 79000988, 79001012, 79001063, 79001072, 79001105, 79001120, 79001123, 79001138, 79001180, 79001225, 79001240, 79001252, 79001282, 79001291, 79001297, 79001303, 79001309, 79001312, 79001321, 79001324, 79001333, 79001360, 79001396, 79001420, 79001471, 79001489, 79001507, 79001546, 79001579, 79001591, 79001612, 79001621, 79001702, 79001723, 79001729, 79001792, 79001819, 79001828, 79001855, 79001861, 79001888, 79001903, 79001909, 79001927, 79001933, 79001951, 79001975, 79001981, 79001990, 79002032, 79002035, 79002038, 79002050, 79002065, 79002092, 79002095, 79002101, 79002125, 79002131, 79002134, 79002143, 79002146, 79002158, 79002173, 79002209, 79002245, 79002257, 79002269, 79002281, 79002302, 79002311, 79002314, 79002326, 79002350, 79002356, 79002395, 79002401, 79002410, 79002419, 79002434, 79002461, 79002470, 79002476, 79002569, 79002572, 79002575, 79002581, 79002596, 79002599, 79002632, 79002662, 79002683, 79002686, 79002692, 79002704, 79002713, 79002734, 79002743, 79002746, 79002773, 79002791, 79002803, 79002845, 79002851, 79002857, 79002893, 79002908, 79002923, 79002929, 79002935, 79002956, 79003019, 79003046, 79003085, 79003112, 79003148, 79003154, 79003160, 79003163, 79003169, 79003184, 79003187, 79003193, 79003214, 79003217, 79003226, 79003238, 79003241, 79003250, 79003253, 79003298, 79003301, 79003313, 79003322, 79003334, 79003340, 79003352, 79003379, 79003475, 79003481, 79003490, 79003496, 79003514, 79003541, 79003547, 79003571, 79003574, 79003577, 79003586, 79003643, 79003709, 79003724, 79003742, 79003751, 79003787, 79003790, 79003805, 79003814, 79003817, 79003832, 79003859, 79003865, 79003874, 79003892, 79003895, 79003904, 79003913, 79003925, 79003937, 79003940, 79003943, 79003952, 79003994, 79004006, 79004009, 79004030, 79004033, 79004036, 79004039, 79004045, 79004048, 79004063, 79004069, 79004072, 79004090, 79004093, 79004099, 79004102, 79004105, 79004111, 79004114, 79004132, 79004138, 79004147, 79004156, 79004159, 79004183, 79004186, 79004201, 79004204, 79004207, and 79004240. Foreign serial numbers still under the firm''s control: 79000361, 79000817, 79001132, 79001915, 79000499, 79000868, 79000898, 79001009, 79001414, 79001432, 79002284, 79003004, 79001258, 79001264, 79001279, 79001234, 79001249, 79000187, 79000559, 79001381, 79000265, 79000292, 79000349, 79000406, 79000424, 79000469, 79000520, 79000565, 79000625, 79000649, 79000673, 79000691, 79000877, 79001039, 79001087, 79001435, 79001474, 79001495, 79001513, 79001567, 79001606, 79001609, 79001774, 79001885, 79001930, 79001948, 79002020, 79002299, 79002365, 79002443, 79002503, 79002506, 79002668, 79002689, 79002755, 79002986, 32687900, 79003292, 79003310, 79003337, 79003358, 79003364, 79003403, 79003406, 79003409, and 79003412.

Other devices under the same FDA submission

12 devices cleared under P010030 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.