Skip to content

Foundation Hip System Bipolar Modular, Size 42mm OD recallZ-0084-2008 · Encore Medical, Lp · initiated Aug 2, 2007

Terminated All recalls posted in November 2007

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Encore Medical, Lp or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Device mislabeled. Hip prosthesis components validated for 5 years were labeled as having a 6 year expiration date.

FDA root cause category: Packaging change control

What the company is doing

Firm notified consignees of recall via phone, e-mail and overnight letters beginning on 08/02/07. Consignees asked to return devices for a properly labeled replacement.

Recall details

Recall number
Z-0084-2008
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Aug 2, 2007
Posted by the FDA
Nov 7, 2007
Recalling firm
Encore Medical, Lp
Quantity
382 units.
Product code
KWY
FDA submissions
K953510
Distribution
Worldwide, including USA, Greece, Japan, and Saudi Arabia.

Product description

Foundation Hip System Bipolar Modular, Size 42mm OD; Cat. #412-01-042; Sterile; assembly required by physician prior to implant; manufactured by Encore Medical, L.P., Austin, TX 78758.

Code information (models, lots, serial numbers)

Lot #: 644021, 648331, 672341, 675901, 678461, 678471, 678481, 678491, 678501, 678511, 678521, 693251, 694501, 697771, 732831, 733871, 736411, 736741, 738071, 738081, 738411, 738421, 770041, 774781, 774791, 777391, 779571, 779581, 783901, 783911, 788351, 794371, 794381, 811151, 811161, 835501, 835511, 836161, 844591, 845951, 848661, 854461, 854471, 854481, 854491, 857681, 874941, 874951, 882561, 882571, 882581, 893551, 893561, 893571, 903121, 903131, 933751, 944241, 944251, 949391, 949401, 964171, 964181, 964191 and 964201.

Other devices under the same FDA submission

19 devices cleared under K953510 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 44788).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page