Medical device recalls: November 2007
The FDA posted 126 medical device recall records in November 2007, from 52 recall events by 43 companies. 126 have not been classified yet.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Ge Healthcare · 4 events, 4 records
- Abbott Laboratories, Inc · 2 events, 3 records
- Toshiba American Med Sys Inc · 2 events, 3 records
- Edwards Lifesciences, Llc · 2 events, 2 records
- Electric Mobility Corp · 2 events, 2 records
- Philips Medical Systems (Cleveland) Inc · 2 events, 2 records
- Stryker Howmedica Osteonics Corp. · 2 events, 2 records
- Stryker Medical Div. of Stryker Corporation · 1 event, 21 records
FDA root cause
- Software design · 33
- Component change control · 23
- Packaging change control · 18
- Device Design · 14
- Process control · 9
- Packaging process control · 6
- Labeling design · 5
- Process design · 4
Not yet classified (126)
The FDA has posted these records but has not published a class for them yet.
- Z-0266-2008 · Telefelx Medical · posted Nov 29, 2007 · Terminated
The specified insulation is not present on the shaft of the devices, leaving a risk of arcing and/or damage to surrounding tissue.
- Z-0267-2008 · Telefelx Medical · posted Nov 29, 2007 · Terminated
The specified insulation is not present on the shaft of the devices, leaving a risk of arcing and/or damage to surrounding tissue.
- Z-0268-2008 · Telefelx Medical · posted Nov 29, 2007 · Terminated
The specified insulation is not present on the shaft of the devices, leaving a risk of arcing and/or damage to surrounding tissue.
- Z-0269-2008 · Siemens Medical Solutions USA Inc · posted Nov 29, 2007 · Terminated
Incorrect Decay Correction Results - ECAT System's software versions 7.2.2 and 7.4 calculate the decay correction for emission data incorrectly, while performing whole-body studies with all transmissions followed by all emissions.
- Z-0270-2008 · Siemens Medical Solutions USA Inc · posted Nov 29, 2007 · Terminated
Incorrect Decay Correction Results - ECAT System's software versions 7.2.2 and 7.4 calculate the decay correction for emission data incorrectly, while performing whole-body studies with all transmissions followed by all emissions.
- Z-0271-2008 · Siemens Medical Solutions USA Inc · posted Nov 29, 2007 · Terminated
Incorrect Decay Correction Results - ECAT System's software versions 7.2.2 and 7.4 calculate the decay correction for emission data incorrectly, while performing whole-body studies with all transmissions followed by all emissions.
- Z-0272-2008 · Siemens Medical Solutions USA Inc · posted Nov 29, 2007 · Terminated
Incorrect Decay Correction Results - ECAT System's software versions 7.2.2 and 7.4 calculate the decay correction for emission data incorrectly, while performing whole-body studies with all transmissions followed by all emissions.
- Z-0273-2008 · Siemens Medical Solutions USA Inc · posted Nov 29, 2007 · Terminated
Incorrect Data Results - ECAT System's software versions 7.2.2 and 7.4 may calculate the decay correction for emission data incorrectly while performing whole-body studies with all transmissions followed by all emissions.
- Z-0275-2008 · Zimmer Inc. · posted Nov 29, 2007 · Terminated
Mislabeled as to size; 54mm package contains 46 mm device and vice versa.
- Z-0276-2008 · Zimmer Inc. · posted Nov 29, 2007 · Terminated
Mislabeled as to size; 54mm package contains 46 mm device and vice versa.
- Z-0171-2008 · Cyberonics, Inc · posted Nov 28, 2007 · Terminated
Dissolution/Fractures to the leads of the VNS Therapy System
- Z-0172-2008 · Cyberonics, Inc · posted Nov 28, 2007 · Terminated
Dissolution/Fractures to the leads of the VNS Therapy System
- Z-0173-2008 · Cyberonics, Inc · posted Nov 28, 2007 · Terminated
Dissolution/Fractures to the leads of the VNS Therapy System
- Z-0177-2008 · Oridion Capnography Inc · posted Nov 28, 2007 · Terminated
Handle Cracking - Cracks in the handle will lead to disconnection of the handle from the monitor, causing the monitor to fall.
- Z-0178-2008 · Oridion Capnography Inc · posted Nov 28, 2007 · Terminated
Handle Cracking - Cracks in the handle will lead to disconnection of the handle from the monitor, causing the monitor to fall.
- Z-0179-2008 · Edwards Lifesciences, Llc · posted Nov 28, 2007 · Terminated
Failure to deploy.
- Z-0265-2008 · GE Healthcare · posted Nov 28, 2007 · Terminated
Aortic Length over-estimated: incorrect measurements when using the Aorta analysis protocol or customized protocols. Until a new software version is installed the aorta analysis protocol or the customized protocols for an aneurysm should NOT BE USED.
- Z-0290-2008 · Wright Medical Technology Inc · posted Nov 27, 2007 · Terminated
Breakage: breaking of the connecting foot that is on the distal surface of the instrument
- Z-0291-2008 · Pioneer Surgical Technology · posted Nov 27, 2007 · Terminated
The pins may come out of the instrument and, if this happens, they could fall into/remain in the patient.
- Z-0293-2008 · Toshiba American Med Sys Inc · posted Nov 27, 2007 · Terminated
System Lock-up: When performing DAS One Shot, the system may lockup due to a software timing error.
- Z-0294-2008 · Medtronic Navigation, Inc · posted Nov 27, 2007 · Terminated
Reversed Instrument Position : a defect in the software version 3.0 causes an error in the position of surgical instruments and relative motion are reversed 180' on the StealthStation image. (If the StealthStation is not used, or if the patient…
- Z-0296-2008 · Xoran Technologies, Inc. · posted Nov 27, 2007 · Terminated
The hard drive may crash, resulting in loss of data from patient CT scans.
- Z-0047-2008 · Boston Scientific CRM Corp · posted Nov 24, 2007 · Terminated
Atypical Charge Time Behavior: 1)-End of Life (EOL) or ERI is displayed during mid-life (typically 24-48 months), even though battery capacity remains available. 2)- Extended ERI charge time limits: Charge times during mid-life may remain below a normal…
- Z-0048-2008 · Boston Scientific CRM Corp · posted Nov 24, 2007 · Terminated
Erroneous Display/Atypical Charge Time Behavior: 1)-End of Life (EOL) or ERI is displayed during mid-life (typically 24-48 months), even though battery capacity remains available. 2)- Extended ERI charge time limits: Charge times during mid-life may remain…
- Z-0049-2008 · Boston Scientific CRM Corp · posted Nov 24, 2007 · Terminated
Atypical Charge Time Behavior: 1)-End of Life (EOL) or ERI is displayed during mid-life (typically 24-48 months), even though battery capacity remains available. 2)- Extended ERI charge time limits: Charge times during mid-life may remain below a normal…
- Z-0050-2008 · Boston Scientific CRM Corp · posted Nov 24, 2007 · Terminated
Atypical Charge Time Behavior: 1)-End of Life (EOL) or ERI is displayed during mid-life (typically 24-48 months), even though battery capacity remains available. 2)- Extended ERI charge time limits: Charge times during mid-life may remain below a normal…
- Z-0051-2008 · Boston Scientific CRM Corp · posted Nov 24, 2007 · Terminated
Atypical Charge Time Behavior: 1)-End of Life (EOL) or ERI is displayed during mid-life (typically 24-48 months), even though battery capacity remains available. 2)- Extended ERI charge time limits: Charge times during mid-life may remain below a normal…
- Z-0052-2008 · Boston Scientific CRM Corp · posted Nov 24, 2007 · Terminated
Atypical Charge Time Behavior: 1)-End of Life (EOL) or ERI is displayed during mid-life (typically 24-48 months), even though battery capacity remains available. 2)- Extended ERI charge time limits: Charge times during mid-life may remain below a normal…
- Z-0053-2008 · Boston Scientific CRM Corp · posted Nov 24, 2007 · Terminated
Atypical Charge Time Behavior: 1)-End of Life (EOL) or ERI is displayed during mid-life (typically 24-48 months), even though battery capacity remains available. 2)- Extended ERI charge time limits: Charge times during mid-life may remain below a normal…
- Z-0255-2008 · Biomet, Inc. · posted Nov 24, 2007 · Terminated
Mislabeled as to size: the package actaully contained a 71 mm tray.
- Z-0257-2008 · Abbott Laboratories, Inc · posted Nov 24, 2007 · Terminated
Sample problems; Five issues identified related to software utilized by the ACCELERATOR APS System Input/Output Module, to include: Under three specific conditions, the post-aspiration radio-frequency identification (RF ID) read of the tube carrier is not…
- Z-0259-2008 · Boston Scientific Target · posted Nov 24, 2007 · Terminated
Mislabeling: Catheters were found to be mislabeled (90 cm labeled devices actually contains 100 cm length catheter)
- Z-0260-2008 · Boston Scientific Target · posted Nov 24, 2007 · Terminated
Mislabeling: Catheters were found to be mislabeled (100 cm labeled devices actually contains 90cm length catheter)
- Z-0262-2008 · Edwards Lifesciences, Llc · posted Nov 24, 2007 · Terminated
MAP offset errors: If users are "slaving" mean arterial pressure (MAP) from the bedside monitor into the Vigilance II Monitor for the purpose of the continuous calculation of systemic vascular resistance (SVR), at times, on some bedside monitors, the…
- Z-0263-2008 · Baxter Bioscience · posted Nov 24, 2007 · Terminated
Sterility (package integrity) compromised: This recall is being conducted due to incomplete seal of the product package at the corner of the outer tray and tyvek lid.
- Z-0264-2008 · B. Braun Interventional Systems · posted Nov 24, 2007 · Terminated
Sterility (package integrity) compromised: package sealing problem leaving packaging being partially open on one of the four sealed Tyvek pouch edges.
- Z-0174-2008 · Electric Mobility Corp · posted Nov 21, 2007 · Terminated
Seat and/or lift actuator problems: 1) Models equipped with a Seat Lift, the mast screw and washer holding the Seat Lift Actuator in place will loosen break off or fall out; 2) Models assembled with Seat Lift Actuator or Seat Post Mounting Bolts, the seat…
- Z-0175-2008 · Electric Mobility Corp · posted Nov 21, 2007 · Terminated
Seat and/or lift actuator problems: 1) Models equipped with a Seat Lift, the mast screw and washer holding the Seat Lift Actuator in place will loosen break off or fall out; 2) Models assembled with Seat Lift Actuator or Seat Post Mounting Bolts, the seat…
- Z-0180-2008 · Philips Medical Systems (Cleveland) Inc · posted Nov 21, 2007 · Terminated
Incorrect simulation and/or beam placement: An anomaly has been identified with the Philips VoxelQ workstation utlizing AcQPlan software version 4.2, 4.2.1, 4.9, 4.9.1, 4.9.2, 4.9.3, 5.0, 5.0.1, 5.0.2 or 5.0.3 when computing absolute patient marking…
- Z-0212-2008 · Thoratec Corp · posted Nov 21, 2007 · Terminated
Pneumatic Leaks: air leaks developing in the pneumatic driveline when subjected to excessive stress at the junction of the percutaneous line to the pump housing due to sharp bending when implanted (paracorporeal-external-position).
- Z-0244-2008 · Applied Biotech Inc · posted Nov 21, 2007 · Terminated
Sensitivity: Kits contain test devices which may not meet the label claims for sensitivity, therefore use of such devices may produce false negative results with patient samples at or near the limit of detection of the product.
- Z-0245-2008 · Applied Biotech Inc · posted Nov 21, 2007 · Terminated
Sensitivity: Kits contain test devices which may not meet the label claims for sensitivity, therefore use of such devices may produce false negative results with patient samples at or near the limit of detection of the product.
- Z-0176-2008 · Depuy Orthopaedics, Inc. · posted Nov 17, 2007 · Terminated
Package Integrity: The outer peelable pouch seals may be compromised due to creasing, although the inner pouch seals are intact.
- Z-0253-2008 · Toshiba American Med Sys Inc · posted Nov 17, 2007 · Terminated
Multiple problems including : (1) The surface temperature of the transducer tip can rise and exceed the limit set by the international standard. (2) Incorrect display of the TI indices on the display monitor. The displayed value is greater or smaller than…
- Z-0254-2008 · Toshiba American Med Sys Inc · posted Nov 17, 2007 · Terminated
Multiple problems including : (1) The surface temperature of the transducer tip can rise and exceed the limit set by the international standard. (2) Incorrect display of the TI indices on the display monitor. The displayed value is greater or smaller than…
- Z-0182-2008 · Pacific Bioscience Lab, Inc. · posted Nov 16, 2007 · Terminated
Overheating: The handle or charger/cradle can overheat, melting the plastic, which potentially can lead to smoke and fire and has caused mild user burns.
- Z-0214-2008 · Stryker Howmedica Osteonics Corp. · posted Nov 16, 2007 · Terminated
Mislabeled: One lot of screws (05x70mm) is labeled as containing 70mm, but may actually contain 75mm length screws.
- Matrix MAX3 Actuator, Catalog Number : MAX30 A300545A252L0V000, SKF USA, Inc. Norristown, PA 19403 ↗Z-0215-2008 · Skf USA Inc · posted Nov 16, 2007 · Terminated
Sudden failure of the actuator during lifting operation
- Z-0216-2008 · Skf USA Inc · posted Nov 16, 2007 · Terminated
Sudden failure of the actuator during lifting operation
- Z-0217-2008 · Skf USA Inc · posted Nov 16, 2007 · Terminated
Sudden failure of the actuator during lifting operation
- Z-0059-2008 · Philips Medical Systems North America Co. Phillips · posted Nov 10, 2007 · Terminated
System Shut Down -- Unexpected Shut Down during use of the Optional Remote Control.
- Z-0060-2008 · Philips Medical Systems North America Co. Phillips · posted Nov 10, 2007 · Terminated
System Shut Down -- Unexpected Shut Down during use of the Optional Remote Control.
- Z-0181-2008 · GE OEC Medical Systems, Inc · posted Nov 10, 2007 · Terminated
Tracking Accuracy - Image intensifier supplier's change on the C-Arm could lead to patient injury, due to incorrect anatomical location of the instrument during a procedure.
- Z-0219-2008 · Stryker Medical Div. of Stryker Corporation · posted Nov 10, 2007 · Terminated
The brake cams may crack or develop excessive wear, resulting in the brakes not locking the wheels, or the stretcher not being locked firmly in position.
- Z-0220-2008 · Stryker Medical Div. of Stryker Corporation · posted Nov 10, 2007 · Terminated
The brake cams may crack or develop excessive wear, resulting in the brakes not locking the wheels, or the stretcher not being locked firmly in position.
- Z-0221-2008 · Stryker Medical Div. of Stryker Corporation · posted Nov 10, 2007 · Terminated
The brake cams may crack or develop excessive wear, resulting in the brakes not locking the wheels, or the stretcher not being locked firmly in position.
- Z-0222-2008 · Stryker Medical Div. of Stryker Corporation · posted Nov 10, 2007 · Terminated
The brake cams may crack or develop excessive wear, resulting in the brakes not locking the wheels, or the stretcher not being locked firmly in position.
- Z-0223-2008 · Stryker Medical Div. of Stryker Corporation · posted Nov 10, 2007 · Terminated
The brake cams may crack or develop excessive wear, resulting in the brakes not locking the wheels, or the stretcher not being locked firmly in position.
- Z-0224-2008 · Stryker Medical Div. of Stryker Corporation · posted Nov 10, 2007 · Terminated
The brake cams may crack or develop excessive wear, resulting in the brakes not locking the wheels, or the stretcher not being locked firmly in position.
- Z-0225-2008 · Stryker Medical Div. of Stryker Corporation · posted Nov 10, 2007 · Terminated
The brake cams may crack or develop excessive wear, resulting in the brakes not locking the wheels, or the stretcher not being locked firmly in position.
- Z-0226-2008 · Stryker Medical Div. of Stryker Corporation · posted Nov 10, 2007 · Terminated
The brake cams may crack or develop excessive wear, resulting in the brakes not locking the wheels, or the stretcher not being locked firmly in position.
- Z-0227-2008 · Stryker Medical Div. of Stryker Corporation · posted Nov 10, 2007 · Terminated
The brake cams may crack or develop excessive wear, resulting in the brakes not locking the wheels, or the stretcher not being locked firmly in position.
- Z-0228-2008 · Stryker Medical Div. of Stryker Corporation · posted Nov 10, 2007 · Terminated
The brake cams may crack or develop excessive wear, resulting in the brakes not locking the wheels, or the stretcher not being locked firmly in position.
- Z-0229-2008 · Stryker Medical Div. of Stryker Corporation · posted Nov 10, 2007 · Terminated
The brake cams may crack or develop excessive wear, resulting in the brakes not locking the wheels, or the stretcher not being locked firmly in position.
- Z-0230-2008 · Stryker Medical Div. of Stryker Corporation · posted Nov 10, 2007 · Terminated
The brake cams may crack or develop excessive wear, resulting in the brakes not locking the wheels, or the stretcher not being locked firmly in position.
- Z-0231-2008 · Stryker Medical Div. of Stryker Corporation · posted Nov 10, 2007 · Terminated
The brake cams may crack or develop excessive wear, resulting in the brakes not locking the wheels, or the stretcher not being locked firmly in position.
- Z-0232-2008 · Stryker Medical Div. of Stryker Corporation · posted Nov 10, 2007 · Terminated
The brake cams may crack or develop excessive wear, resulting in the brakes not locking the wheels, or the stretcher not being locked firmly in position.
- Z-0233-2008 · Stryker Medical Div. of Stryker Corporation · posted Nov 10, 2007 · Terminated
The brake cams may crack or develop excessive wear, resulting in the brakes not locking the wheels, or the stretcher not being locked firmly in position.
- Z-0234-2008 · Stryker Medical Div. of Stryker Corporation · posted Nov 10, 2007 · Terminated
The brake cams may crack or develop excessive wear, resulting in the brakes not locking the wheels, or the stretcher not being locked firmly in position.
- Z-0235-2008 · Stryker Medical Div. of Stryker Corporation · posted Nov 10, 2007 · Terminated
The brake cams may crack or develop excessive wear, resulting in the brakes not locking the wheels, or the stretcher not being locked firmly in position.
- Z-0236-2008 · Stryker Medical Div. of Stryker Corporation · posted Nov 10, 2007 · Terminated
The brake cams may crack or develop excessive wear, resulting in the brakes not locking the wheels, or the stretcher not being locked firmly in position.
- Z-0237-2008 · Stryker Medical Div. of Stryker Corporation · posted Nov 10, 2007 · Terminated
The brake cams may crack or develop excessive wear, resulting in the brakes not locking the wheels, or the stretcher not being locked firmly in position.
- Z-0238-2008 · Stryker Medical Div. of Stryker Corporation · posted Nov 10, 2007 · Terminated
The brake cams may crack or develop excessive wear, resulting in the brakes not locking the wheels, or the stretcher not being locked firmly in position.
- Z-0239-2008 · Stryker Medical Div. of Stryker Corporation · posted Nov 10, 2007 · Terminated
The brake cams may crack or develop excessive wear, resulting in the brakes not locking the wheels, or the stretcher not being locked firmly in position.
- Z-0119-2008 · Opti Medical Systems Inc · posted Nov 9, 2007 · Terminated
Incorrect results/stability failure; Actual device stability studies show sodium and chloride sensors age differently than anticipated when stored at the high end of labeled storage, therefore subsequent measurements give results outside the allowed error…
- Z-0155-2008 · Siemens Medical Solutions USA, Inc · posted Nov 9, 2007 · Terminated
Mistreatment of radiation therapy due to inadvertent parameter changes or miss-calibration in Service Mode. The miss-calibration can affect the beam profile and Virtual Wedge angle. This leads to patient mistreatment and alteration of future treatment…
- Z-0156-2008 · Siemens Medical Solutions USA, Inc · posted Nov 9, 2007 · Terminated
Mistreatment of radiation therapy due to inadvertent parameter changes or miss-calibration in Service Mode. The miss-calibration can affect the beam profile and Virtual Wedge angle. This leads to patient mistreatment and alteration of future treatment…
- Z-0157-2008 · Siemens Medical Solutions USA, Inc · posted Nov 9, 2007 · Terminated
Mistreatment of radiation therapy due to inadvertent parameter changes or miss-calibration in Service Mode. The miss-calibration can affect the beam profile and Virtual Wedge angle. This leads to patient mistreatment and alteration of future treatment…
- Z-0158-2008 · Siemens Medical Solutions USA, Inc · posted Nov 9, 2007 · Terminated
Mistreatment of radiation therapy due to inadvertent parameter changes or miss-calibration in Service Mode. The miss-calibration can affect the beam profile and Virtual Wedge angle. This leads to patient mistreatment and alteration of future treatment…
- Z-0159-2008 · Siemens Medical Solutions USA, Inc · posted Nov 9, 2007 · Terminated
Mistreatment of radiation therapy due to inadvertent parameter changes or miss-calibration in Service Mode. The miss-calibration can affect the beam profile and Virtual Wedge angle. This leads to patient mistreatment and alteration of future treatment…
- Z-0160-2008 · Siemens Medical Solutions USA, Inc · posted Nov 9, 2007 · Terminated
Mistreatment of radiation therapy due to inadvertent parameter changes or miss-calibration in Service Mode. The miss-calibration can affect the beam profile and Virtual Wedge angle. This leads to patient mistreatment and alteration of future treatment…
- Z-0161-2008 · Siemens Medical Solutions USA, Inc · posted Nov 9, 2007 · Terminated
Mistreatment of radiation therapy due to inadvertent parameter changes or miss-calibration in Service Mode. The miss-calibration can affect the beam profile and Virtual Wedge angle. This leads to patient mistreatment and alteration of future treatment…
- Z-0162-2008 · Siemens Medical Solutions USA, Inc · posted Nov 9, 2007 · Terminated
Mistreatment of radiation therapy due to inadvertent parameter changes or miss-calibration in Service Mode. The miss-calibration can affect the beam profile and Virtual Wedge angle. This leads to patient mistreatment and alteration of future treatment…
- Z-0163-2008 · Siemens Medical Solutions USA, Inc · posted Nov 9, 2007 · Terminated
Mistreatment of radiation therapy due to inadvertent parameter changes or miss-calibration in Service Mode. The miss-calibration can affect the beam profile and Virtual Wedge angle. This leads to patient mistreatment and alteration of future treatment…
- Z-0164-2008 · Siemens Medical Solutions USA, Inc · posted Nov 9, 2007 · Terminated
Mistreatment of radiation therapy due to inadvertent parameter changes or miss-calibration in Service Mode. The miss-calibration can affect the beam profile and Virtual Wedge angle. This leads to patient mistreatment and alteration of future treatment…
- Z-0165-2008 · Siemens Medical Solutions USA, Inc · posted Nov 9, 2007 · Terminated
Mistreatment of radiation therapy due to inadvertent parameter changes or miss-calibration in Service Mode. The miss-calibration can affect the beam profile and Virtual Wedge angle. This leads to patient mistreatment and alteration of future treatment…
- Z-0166-2008 · Siemens Medical Solutions USA, Inc · posted Nov 9, 2007 · Terminated
Mistreatment of radiation therapy due to inadvertent parameter changes or miss-calibration in Service Mode. The miss-calibration can affect the beam profile and Virtual Wedge angle. This leads to patient mistreatment and alteration of future treatment…
- Z-0167-2008 · Siemens Medical Solutions USA, Inc · posted Nov 9, 2007 · Terminated
Mistreatment of radiation therapy due to inadvertent parameter changes or miss-calibration in Service Mode. The miss-calibration can affect the beam profile and Virtual Wedge angle. This leads to patient mistreatment and alteration of future treatment…
- Z-0168-2008 · Siemens Medical Solutions USA, Inc · posted Nov 9, 2007 · Terminated
Mistreatment of radiation therapy due to inadvertent parameter changes or miss-calibration in Service Mode. The miss-calibration can affect the beam profile and Virtual Wedge angle. This leads to patient mistreatment and alteration of future treatment…
- Z-0169-2008 · Siemens Medical Solutions USA, Inc · posted Nov 9, 2007 · Terminated
Mistreatment of radiation therapy due to inadvertent parameter changes or miss-calibration in Service Mode. The miss-calibration can affect the beam profile and Virtual Wedge angle. This leads to patient mistreatment and alteration of future treatment…
- Z-0256-2008 · Bio-Rad Labroatories Inc. · posted Nov 9, 2007 · Terminated
Mislabeling: Reporting terminology in the Syphilis IgG APF CD is not consistent with the distributed Instructions for Use. (Non-Reactive and Reactive rather than Negative and Positive)
- Z-0075-2008 · GE Healthcare · posted Nov 8, 2007 · Terminated
Design Problems Including: 1) Hot Brake (Longitudinal Overheating) potential risk of burns 2) Brake Locks and Sudden Releases 3) Emergency Stop Circuit Button; open instead of closed 4) Longitudinal Brakes Release does not engage properly 5) Tilting…
- Z-0213-2008 · Varian Medical Systems Inc · posted Nov 8, 2007 · Terminated
Unexpected loss of Treatement Field Add-Ons (Wedges, Blocks, etc) during Ximatron simulation. If an unapproved treatment field is created without an MLC, but with an add-on, the add on will be deleted without warning. If this is not detected and…
- Z-0218-2008 · Cardinal Health 303 Inc DBA Alaris Products · posted Nov 8, 2007 · Terminated
Leaks: there is a risk of leakage at the female end of the Texium closed male luer (CML) when attached to a spin luer. (NOTE: users who connect the female portion of the Texium closed male luer to fixed (one-piece) male luers on tubing or on syringes are…
- Z-0082-2008 · Encore Medical, Lp · posted Nov 7, 2007 · Terminated
Device mislabeled. Hip prosthesis components validated for 5 years were labeled as having a 6 year expiration date.
- Z-0083-2008 · Encore Medical, Lp · posted Nov 7, 2007 · Terminated
Device mislabeled. Hip prosthesis components validated for 5 years were labeled as having a 6 year expiration date.
- Z-0084-2008 · Encore Medical, Lp · posted Nov 7, 2007 · Terminated
Device mislabeled. Hip prosthesis components validated for 5 years were labeled as having a 6 year expiration date.
- Z-0085-2008 · Encore Medical, Lp · posted Nov 7, 2007 · Terminated
Device mislabeled. Hip prosthesis components validated for 5 years were labeled as having a 6 year expiration date.
- Z-0086-2008 · Encore Medical, Lp · posted Nov 7, 2007 · Terminated
Device mislabeled. Hip prosthesis components validated for 5 years were labeled as having a 6 year expiration date.
- Z-0087-2008 · Encore Medical, Lp · posted Nov 7, 2007 · Terminated
Device mislabeled; hip prosthesis components validated for 5 years were labeled as having a 6 year expiration date.
- Z-0088-2008 · Encore Medical, Lp · posted Nov 7, 2007 · Terminated
Device mislabeled; hip prosthesis components validated for 5 years were labeled as having a 6 year expiration date.
- Z-0089-2008 · Encore Medical, Lp · posted Nov 7, 2007 · Terminated
Device mislabeled; hip prosthesis components validated for 5 years were labeled as having a 6 year expiration date.
- Z-0090-2008 · Encore Medical, Lp · posted Nov 7, 2007 · Terminated
Device mislabeled; hip prosthesis components validated for 5 years were labeled as having a 6 year expiration date.
- Z-0091-2008 · Encore Medical, Lp · posted Nov 7, 2007 · Terminated
Device mislabeled; hip prosthesis components validated for 5 years were labeled as having a 6 year expiration date.
- Z-0092-2008 · Encore Medical, Lp · posted Nov 7, 2007 · Terminated
Device mislabeled; hip prosthesis components validated for 5 years were labeled as having a 6 year expiration date.
- Z-0093-2008 · Encore Medical, Lp · posted Nov 7, 2007 · Terminated
Device mislabeled; hip prosthesis components validated for 5 years were labeled as having a 6 year expiration date.
- Z-0094-2008 · Encore Medical, Lp · posted Nov 7, 2007 · Terminated
Device mislabeled; hip prosthesis components validated for 5 years were labeled as having a 6 year expiration date.
- Z-0095-2008 · Encore Medical, Lp · posted Nov 7, 2007 · Terminated
Device mislabeled; hip prosthesis components validated for 5 years were labeled as having a 6 year expiration date.
- Z-0096-2008 · Encore Medical, Lp · posted Nov 7, 2007 · Terminated
Device mislabeled; hip prosthesis components validated for 5 years were labeled as having a 6 year expiration date.
- Z-0097-2008 · Encore Medical, Lp · posted Nov 7, 2007 · Terminated
Device mislabeled; hip prosthesis components validated for 5 years were labeled as having a 6 year expiration date.
- Z-0098-2008 · Encore Medical, Lp · posted Nov 7, 2007 · Terminated
Device mislabeled; hip prosthesis components validated for 5 years were labeled as having a 6 year expiration date.
- Z-0099-2008 · Encore Medical, Lp · posted Nov 7, 2007 · Terminated
Device mislabeled; hip prosthesis components validated for 5 years were labeled as having a 6 year expiration date.
- Z-0100-2008 · Smiths Medical ASD, Inc. · posted Nov 7, 2007 · Terminated
Mislabeled. The product is labeled as a size 9.0 mm tracheostomy tube, but the package contains a 7.0 mm tracheostomy tube.
- Z-0113-2008 · Ge Healthcare · posted Nov 7, 2007 · Terminated
When a spectrum of a Pulsed Wave Doppler shows an angle correction unequal to 0 is reloaded from the archive or the image is sent to a DICOM station, this angle correction is mistakenly disregarded resulting in an underestimation of the flow velocities.
- Z-0114-2008 · Warsaw Orthopedic, Inc. dba Medtronic SOFAMOR DANEK · posted Nov 7, 2007 · Terminated
Packaging weakness may compromise sterility and instructions for use are inadequate.
- Z-0121-2008 · Philips Medical Systems (Cleveland) Inc · posted Nov 7, 2007 · Terminated
The maximum intensity projection (MIP) images produced from respiratory gated images by the Tumor Localization (Tumor LOC) software application on Brilliance Big Bore CT scanners will be labeled incorrectly for certain patient orientations and view…
- Z-0144-2008 · Karl Storz Endoscopy America Inc · posted Nov 7, 2007 · Terminated
Outflow Difference: The accuracy of the Equimat displayed value could be influenced by external factors of the cart/stand and may result in a fluctuating or inaccurate reading.
- Z-0146-2008 · Abbott Laboratories, Inc · posted Nov 7, 2007 · Terminated
Incorrect Sample/Test Identification: Due to a software timing defect with certain versions of ARCHITECT cSystem Software, test results may be incorrectly associated to a Sample Identification during specific sample-processing conditions of the Robotic…
- Z-0147-2008 · Abbott Laboratories, Inc · posted Nov 7, 2007 · Terminated
Incorrect Sample/Test Identification: Due to a software timing defect with certain versions of ARCHITECT cSystem Software, test results may be incorrectly associated to a Sample Identification during specific sample-processing conditions of the Robotic…
- Z-0170-2008 · GE Healthcare · posted Nov 7, 2007 · Terminated
Incorrect patient positioning; when moving the patient into the bore may lead to wrist and hand injury.
- Z-0081-2008 · Medtronic Inc. Cardiac Rhythm Managment · posted Nov 6, 2007 · Terminated
An issue with the Paceart System - 2006 First Edition (Get Connected Edition) exists. Under certain circumstances, Paceart System Generic ICD reports may not accurately reflect patient VT/VF, SVT/NST and Mode Switch/AT/AF episode detection data. When…
- Z-0129-2008 · Welch Allyn Protocol, Inc · posted Nov 6, 2007 · Terminated
Failure to Deliver Shock; a defective capacitor may cause the delay or non-delivery of the defibrillating shock which may result in failure to resuscitate the patient
- Z-0001-2008 · Bio-Logic Systems Corp · posted Nov 3, 2007 · Terminated
Bio-logic System Corp Camera Pole may weaken and detach at the base of the telescoping camera pole during transport causing pole to break at the welded base and fall.
- Z-0064-2008 · Abbott Diagnostic International, Ltd. · posted Nov 3, 2007 · Terminated
Incorrect Results -- Error codes 1062, 1063, 1113, and 1118 observed when running patient samples on the AxSYM Digoxin III assay, rather than a concentration, is reported for the patient sample.
- Z-0133-2008 · EMD Chemicals Inc · posted Nov 3, 2007 · Terminated
Out of Spec -- The Batch (Lot) was incorrectly released after being rejected for out of specification of high azure and eosin absorbance at the end of the run and particulate matter at the end of the run.
- Z-0080-2008 · Stryker Howmedica Osteonics Corp. · posted Nov 1, 2007 · Terminated
The plasma spray may delaminate from the proximal portion of the humeral stem.
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.