Foundation Hip System Bipolar Modular, Size 46mm OD recallZ-0088-2008 · Encore Medical, Lp · initiated Aug 2, 2007
Terminated All recalls posted in November 2007
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Encore Medical, Lp or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Device mislabeled; hip prosthesis components validated for 5 years were labeled as having a 6 year expiration date.
FDA root cause category: Packaging change control
What the company is doing
Firm notified consignees of recall via phone, e-mail and overnight letters beginning on 08/02/07. Consignees asked to return devices for a properly labeled replacement.
Recall details
- Recall number
- Z-0088-2008
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Aug 2, 2007
- Posted by the FDA
- Nov 7, 2007
- Recalling firm
- Encore Medical, Lp
- Quantity
- 605 units.
- Product code
- KWY
- FDA submissions
- K953510
- Distribution
- Worldwide, including USA, Greece, Japan, and Saudi Arabia.
Product description
Foundation Hip System Bipolar Modular, Size 46mm OD; Cat. #412-01-046; Sterile; assembly required by physician prior to implant; manufactured by Encore Medical, L.P., Austin, TX 78758.
Code information (models, lots, serial numbers)
Lot #: 644031, 672331, 677061, 677981, 677991, 678691, 678701, 678711, 678721, 679651, 679661, 697671, 691281, 691291, 693261, 699171, 730241, 735791, 736491, 736501, 736511, 736771, 737811, 738441, 738451, 738461, 770101, 770111, 770221, 773281, 777421, 778431, 782551, 788441, 788451, 810911, 811211, 813501, 830751, 830761, 830911, 830921, 835531, 835541, 842361, 842371, 847411, 848681, 848691, 853541, 857731, 857741, 857751, 875571, 875581, 878371, 878381, 878391, 885381, 885401, 892331, 894241, 894251, 905611, 907121, 907131, 907141, 907211, 925801, 925811, 927241, 927251, 927261, 932761, 949471, 949491, 949851, 951101, 951111, 951121, 958941, 966131, 966141 and 982721.
Other devices under the same FDA submission
19 devices cleared under K953510 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- DJO SURGICAL 412-02-046Encore Medical, L.P. · Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented14 reports
- DJO SURGICAL 412-02-048Encore Medical, L.P. · Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented14 reports
- DJO SURGICAL 412-02-045Encore Medical, L.P. · Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented12 reports
- DJO SURGICAL 412-02-052Encore Medical, L.P. · Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented11 reports
- DJO SURGICAL 412-02-044Encore Medical, L.P. · Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented8 reports
- DJO SURGICAL 412-02-047Encore Medical, L.P. · Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented8 reports
- DJO SURGICAL 412-02-050Encore Medical, L.P. · Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented6 reports
- DJO SURGICAL 412-02-042Encore Medical, L.P. · Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented5 reports
- DJO SURGICAL 412-02-054Encore Medical, L.P. · Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented5 reports
- DJO SURGICAL 412-02-049Encore Medical, L.P. · Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented4 reports
- DJO SURGICAL 412-02-051Encore Medical, L.P. · Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented4 reports
- DJO SURGICAL 412-02-056Encore Medical, L.P. · Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented4 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 44788).
- Z-0082-2008 · TerminatedFoundation Hip System Bipolar Modular, Size 40mm OD; Cat. #412-01-040; Sterile; assembly required by physician prior to implant; manufactured by Encore Medical, L.P., Austin, TX 7…
- Z-0083-2008 · TerminatedFoundation Hip System Bipolar Modular, Size 41mm OD; Cat. #412-01-041; Sterile; assembly required by physician prior to implant; manufactured by Encore Medical, L.P., Austin, TX 7…
- Z-0084-2008 · TerminatedFoundation Hip System Bipolar Modular, Size 42mm OD; Cat. #412-01-042; Sterile; assembly required by physician prior to implant; manufactured by Encore Medical, L.P., Austin, TX 7…
- Z-0085-2008 · TerminatedFoundation Hip System Bipolar Modular, Size 43mm OD; Cat. #412-01-043; Sterile; assembly required by physician prior to implant; manufactured by Encore Medical, L.P., Austin, TX 7…
- Z-0086-2008 · TerminatedFoundation Hip System Bipolar Modular, Size 44mm OD; Cat. #412-01-044; Sterile; assembly required by physician prior to implant; manufactured by Encore Medical, L.P., Austin, TX 7…
- Z-0087-2008 · TerminatedFoundation Hip System Bipolar Modular, Size 45mm OD; Cat. #412-01-045; Sterile; assembly required by physician prior to implant; manufactured by Encore Medical, L.P., Austin, TX 7…
- Z-0089-2008 · TerminatedFoundation Hip System Bipolar Modular, Size 47mm OD; Cat. #412-01-047; Sterile; assembly required by physician prior to implant; manufactured by Encore Medical, L.P., Austin, TX 7…
- Z-0090-2008 · TerminatedFoundation Hip System Bipolar Modular, Size 48mm OD; Cat. #412-01-048; Sterile; assembly required by physician prior to implant; manufactured by Encore Medical, L.P., Austin, TX 7…
- Z-0091-2008 · TerminatedFoundation Hip System Bipolar Modular, Size 49mm OD; Cat. #412-01-049; Sterile; assembly required by physician prior to implant; manufactured by Encore Medical, L.P., Austin, TX 7…
- Z-0092-2008 · TerminatedFoundation Hip System Bipolar Modular, Size 50mm OD; Cat. #412-01-050; Sterile; assembly required by physician prior to implant; manufactured by Encore Medical, L.P., Austin, TX 7…
- Z-0093-2008 · TerminatedFoundation Hip System Bipolar Modular, Size 51mm OD; Cat. #412-01-051; Sterile; assembly required by physician prior to implant; manufactured by Encore Medical, L.P., Austin, TX 7…
- Z-0094-2008 · TerminatedFoundation Hip System Bipolar Modular, Size 52mm OD; Cat. #412-01-052; Sterile; assembly required by physician prior to implant; manufactured by Encore Medical, L.P., Austin, TX 7…
- Z-0095-2008 · TerminatedFoundation Hip System Bipolar Modular, Size 53mm OD; Cat. #412-01-053; Sterile; assembly required by physician prior to implant; manufactured by Encore Medical, L.P., Austin, TX 7…
- Z-0096-2008 · TerminatedFoundation Hip System Bipolar Modular, Size 54mm OD; Cat. #412-01-054; Sterile; assembly required by physician prior to implant; manufactured by Encore Medical, L.P., Austin, TX 7…
- Z-0097-2008 · TerminatedFoundation Hip System Bipolar Modular, Size 56mm OD; Cat. #412-01-056; Sterile; assembly required by physician prior to implant; manufactured by Encore Medical, L.P., Austin, TX 7…
- Z-0098-2008 · TerminatedFoundation Hip System Bipolar Modular, Size 58mm OD; Cat. #412-01-058; Sterile; assembly required by physician prior to implant; manufactured by Encore Medical, L.P., Austin, TX 7…
- Z-0099-2008 · TerminatedFoundation Hip System Bipolar Modular, Size 60mm OD; Cat. #412-01-060; Sterile; assembly required by physician prior to implant; manufactured by Encore Medical, L.P., Austin, TX 7…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page