Roche LIPC Lipase Colorimetric Reagent recallZ-0688-2008 · Roche Diagnostics Corp. · initiated Dec 3, 2007

Terminated All recalls posted in February 2008

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Roche Diagnostics Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

False Results- The reagent of the LDL-Cholesterol assay shows a carryover effect on the Lipase assay. Results may be falsely elevated by up to 50 units per liter when used on COBAS INTEGRA 400 or COBAS INTEGRA 400+ analyzers.

FDA root cause category: Process design

What the company is doing

Consignees were sent and Urgent Medical Device Correction notice dated 12/03/07. The notice provided instructions for conducting an additional wash cycle whenever both the LDL-Cholesterol assays and Lipases assay are used on the COBAS INTEGRA 400 or COBAS INTEGRA 400+ analyzers. Contact Roche Technical Support at 1-800-428-2336 for questions regarding these instructions.

Recall details

Recall number
Z-0688-2008
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Dec 3, 2007
Posted by the FDA
Feb 8, 2008
Recalling firm
Roche Diagnostics Corp.
Quantity
9,817 kits.
Product code
JFJ
FDA submissions
K972250
Distribution
Nationwide Distribution

Product description

Roche LIPC Lipase Colorimetric Reagent; Catalog Number: 03029590322. Roche Diagnostics Corp., Indianapolis, IN 46250

Code information (models, lots, serial numbers)

Lots 68580101, exp. 11/30/2007; 68711401, exp. 1/31/2008; 68873201, exp. 3/31/2008; 69119401, exp. 5/31/2008 and 69279301, exp. 7/31/2008.