Grey Inter-Unit Interface Connectors ("IUI") on the… recallZ-1711-2008 · Cardinal Health 303 Inc DBA Alaris Products · initiated Mar 5, 2008
Terminated All recalls posted in June 2008
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Cardinal Health 303 Inc DBA Alaris Products or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
A channel error on an affected Alaris System module will stop an active infusion (or monitoring) with an audible and visual alarm. Channel errors with the Alaris System due to the Inter-Unit Interface (IUI) connectors on all Alaris System modules (manufactured on or before March 26, 2004). Grey Inter-Unit Interface Connectors ("IUI") on the following Alaris System modules built on or before March
FDA root cause category: Device Design
What the company is doing
An Urgent Medical Device Recall Notification letter was sent to consignees on April 10, 2008. Four customer letters were issued. The U19 IC issue affects only the Pump module and the U9 IC issue affects only the Syringe and PCA modules. Therefore there will be one letter for customers affected by all three issues Ul9 IC, U9 IC, and the IUI, one letter for affected Ul9 IC and U9 IC customers, one letter for U19 IC-only customers and one letter for U9 IC-only customers. All mailings were conducted through consignees (distributors as well as direct customers for the affected devices). Distributors will be sent multiple copies of the Recall Notification and Customer Response Cards. The Recall letters all includeRequired Actions sections which give their customers specific information about the issue and what the customers are to do and includes a web site link for each specific problem. Each recall letter also has chart of Cardinal Health contacts support.
Recall details
- Recall number
- Z-1711-2008
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Mar 5, 2008
- Posted by the FDA
- Jun 10, 2008
- Recalling firm
- Cardinal Health 303 Inc DBA Alaris Products
- Quantity
- 77,515 Units
- Product code
- FRN
- FDA submissions
- K010966, K012383, K022677, K023264, K043299, K950419
- Distribution
- Class II Recall - Worldwide Distribution --- USA including states of AR, AZ, CA, CT, CO, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, ME, MD MI, MN, MO, MS, NC, NE, NH, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI and WY, Guam and country of Canada. ****CDRH Comments: "Note for District Office to follow-up" The firm has not provided a grand total of devices affected by these problems. It is possible that devices suffering from U19 & U9 problems may also experience IUI problems. This recall has 3 problems with 6 modules. One problem affects only 2 modules, one problem affects 6 modules. For simplicity and expediency, CDRH agrees to keep the file as is, rather than separate into 6 products and have complicated explanations of the problems. 77,515 devices affected by IUI problem. 85,286 devices affected by U19 problem. 8,286 devices affected by U9 problem. ****
Product description
Grey Inter-Unit Interface Connectors ("IUI") on the following Alaris System modules built on or before March 26, 2004: (1) Alaris PC point-of-care unit ("PC Unit") (Model 8000)* (a.k.a. Medley PC Unit), (2) Alaris Pump Module (Model 8100) (a.k.a. Medley Pump Module), (3) Alaris Syringe Module (Model 8110) (a.k.a. Medley Syringe Module), (4) Alaris PCA Module (Model 8120) (a.k.a. Medley PCA Module), (5) Alaris SpO2 Module (Models 8210) (a.k.a. Medley SpO2 Module) with Nellcor technology, and (6) Alaris SpO2 Module (Models 8220) (a.k.a. Medley SpO2 Module) with Masimo technology, CardinalHealth, San Diego, CA 92130
Code information (models, lots, serial numbers)
Pending
Other devices under the same FDA submission
7 devices cleared under K010966, K012383, K022677, K023264, K043299, K950419 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Alaris 8100Carefusion 303, Inc. · Pump, infusion2 recalls2 deaths650,218 reports
- Alaris 8110Carefusion 303, Inc. · Pump, infusion2 recalls4 deaths21,956 reports
- Alaris 8110Carefusion 303, Inc. · Pump, infusion2 recalls2 deaths15,547 reports
- Alaris 8120Carefusion 303, Inc. · Pump, infusion, pca2 recalls4 deaths14,640 reports
- Alaris 8210Carefusion 303, Inc. · Oximeter168 reports
- Alaris 8220Carefusion 303, Inc. · Oximeter77 reports
- Alaris 8120Carefusion 303, Inc. · Pump, infusion, pca1 recall5 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 47335).
- Z-1710-2008 · TerminatedU19 Socketed Integrated Circuit ("U19 IC") on the logic board of Alaris Pump modules (aka. Medley Pump modules), Model 8100 built on or before September 8, 2005. CardinalHealth, S…
- Z-1712-2008 · TerminatedU9 Socket Integrated Circuit ("U9 IC") on the display board of the following Alaris System modules built on or before October 4, 2005: (1) Alaris Syringe module (Model 8110) (a.k.a…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.