Skip to content

U9 Socket Integrated Circuit ("U9 IC") on the display board… recallZ-1712-2008 · Cardinal Health 303 Inc DBA Alaris Products · initiated Mar 5, 2008

Terminated All recalls posted in June 2008

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Cardinal Health 303 Inc DBA Alaris Products or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A channel error on an affected Alaris System module will stop an active infusion (or monitoring) with an audible and visual alarm. Channel errors due to failure of the U9 Socketed Integrated Circuits (U9 IC) on the display board of the Alaris Syringe Module ("Syringe module") and Alaris PCA module ("PCA module") for modules manufactured on or before October 4, 2005. The user will be required to re

FDA root cause category: Device Design

What the company is doing

An Urgent Medical Device Recall Notification letter was sent to consignees on April 10, 2008. Four customer letters were issued. The U19 IC issue affects only the Pump module and the U9 IC issue affects only the Syringe and PCA modules. Therefore there will be one letter for customers affected by all three issues Ul9 IC, U9 IC, and the IUI, one letter for affected Ul9 IC and U9 IC customers, one letter for U19 IC-only customers and one letter for U9 IC-only customers. All mailings were conducted through consignees (distributors as well as direct customers for the affected devices). Distributors will be sent multiple copies of the Recall Notification and Customer Response Cards. The Recall letters all includeRequired Actions sections which give their customers specific information about the issue and what the customers are to do and includes a web site link for each specific problem. Each recall letter also has chart of Cardinal Health contacts support.

Recall details

Recall number
Z-1712-2008
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Mar 5, 2008
Posted by the FDA
Jun 10, 2008
Recalling firm
Cardinal Health 303 Inc DBA Alaris Products
Quantity
8,286 units
Product code
FRN
FDA submissions
K010966, K012383, K022677, K023264, K043299, K950419
Distribution
Class II Recall - Worldwide Distribution --- USA including states of AR, AZ, CA, CT, CO, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, ME, MD MI, MN, MO, MS, NC, NE, NH, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI and WY, Guam and country of Canada. ****CDRH Comments: "Note for District Office to follow-up" The firm has not provided a grand total of devices affected by these problems. It is possible that devices suffering from U19 & U9 problems may also experience IUI problems. This recall has 3 problems with 6 modules. One problem affects only 2 modules, one problem affects 6 modules. For simplicity and expediency, CDRH agrees to keep the file as is, rather than separate into 6 products and have complicated explanations of the problems. 77,515 devices affected by IUI problem. 85,286 devices affected by U19 problem. 8,286 devices affected by U9 problem. ****

Product description

U9 Socket Integrated Circuit ("U9 IC") on the display board of the following Alaris System modules built on or before October 4, 2005: (1) Alaris Syringe module (Model 8110) (a.k.a. Medley Syringe module) (2) Alaris PCA module (Model 8120) (a.k.a. Medley PCA module), CardinalHealth, San Diego, CA 92130

Code information (models, lots, serial numbers)

Pending

Other devices under the same FDA submission

7 devices cleared under K010966, K012383, K022677, K023264, K043299, K950419 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 47335).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page