Depuy DELTA Xtend, Standard Humerus PE Cup, Dia 38/+6 STD… recallZ-0533-2009 · Depuy Orthopaedics, Inc. · initiated Jul 10, 2008
Terminated All recalls posted in January 2009
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Depuy Orthopaedics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Compressed air may become trapped between the cup and the stem, preventing the tapered walls from engaging and causing the cup to feel unstable.
FDA root cause category: Device Design
What the company is doing
A letter, "Urgent Device Correction" dated July 8, 2008 was sent via certified mail to implanting surgeons advising them of the issues and guidance on how to mitigate the problem. A communication was also sent to all sales representatives and territory offices on July 15, 2008. For questions, please contact Depuy Orthopaedics, Inc. Customer Service at 1-800-473-3789.
Recall details
- Recall number
- Z-0533-2009
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Jul 10, 2008
- Posted by the FDA
- Jan 15, 2009
- Recalling firm
- Depuy Orthopaedics, Inc.
- Quantity
- 1,521 of all devices.
- Product code
- KWS
- FDA submissions
- K062250, K081620
- Distribution
- Nationwide Distribution.
Product description
Depuy DELTA Xtend, Standard Humerus PE Cup, Dia 38/+6 STD, sterile, Depuy France; REF 130738206. The device is used as a shoulder implant.
Code information (models, lots, serial numbers)
All lots.
Other devices under the same FDA submission
38 devices cleared under K062250, K081620 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- DELTA XTEND 130760000Depuy (Ireland) · shoulder prosthesis, reverse configuration223 reports
- DELTA XTEND 130738203Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED175 reports
- DELTA XTEND 130790036Depuy (Ireland) · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED171 reports
- DELTA XTEND 130770018Depuy (Ireland) · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED159 reports
- DELTA XTEND 130760138Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED146 reports
- DELTA XTEND 130790030Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED132 reports
- DELTA XTEND 130790024Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED122 reports
- DELTA XTEND 130738206Depuy (Ireland) · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED116 reports
- DELTA XTEND 130760142Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED116 reports
- DELTA XTEND 130742203Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED104 reports
- DELTA XTEND 130742206Depuy (Ireland) · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED99 reports
- DELTA XTEND 130738209Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED87 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 49112).
- Z-0532-2009 · TerminatedDepuy DELTA Xtend, Standard humeral PE Cup Dia 38/+3 STD, sterile, Depuy France; REF 130738203. The device is used as a shoulder implant.
- Z-0534-2009 · TerminatedDepuy DELTA Xtend, Standard Humeral PE Cup, Dia 38/+9 STD, sterile, Depuy France; REF 130738209. The device is used as a shoulder implant.
- Z-0535-2009 · TerminatedDepuy DELTA Xtend, Standard Humeral PE Cup, Dia 38/+6R RET, sterile, Depuy France; REF 130738106. The device is used as a shoulder implant.
- Z-0536-2009 · TerminatedDepuy DELTA Xtend, Standard Humeral PE Cup, Dia 42/+3 STD, sterile, Depuy France; REF 130742203. The device is used as a shoulder implant.
- Z-0537-2009 · TerminatedDepuy DELTA Xtend, Standard Humeral PE Cup, Dia 42/+6 STD, sterile, Depuy France; REF 130742206. The device is used as a shoulder implant.
- Z-0538-2009 · TerminatedDepuy DELTA Xtend, Standard Humeral PE Cup, Dia 42/+9 STD, sterile, Depuy France; REF 130742209. The device is used as a shoulder implant.
- Z-0539-2009 · TerminatedDepuy DELTA Xtend, Standard Humeral PE Cup, Dia 42+6R RET, sterile, Depuy France; REF 130742106. The device is used as a shoulder implant.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page