DELTA XTEND 130738206Recalls & reported problems
Depuy (Ireland) · DELTA Xtend Standard Humeral PE Cup Dia38 /+6mm STD · Class 2 · Implant
All models of DELTA XTEND (194)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This DePuy Delta Xtend shoulder implant is a replacement part used in shoulder replacement surgery. Many of the 112 reports describe the shoulder dislocating or the implant loosening, often leading to another surgery with a more secure part. The reports include 108 injuries and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 116
- Deaths reported
- 0
- Injuries reported
- 112
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 116 adverse event reports that list this exact device (DI 10603295027379), received between Jun 23, 2017 and Sep 28, 2026 — classified as 112 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (52 reports).
10 other recalls cover devices cleared under the same FDA submission (K062250, K081620, K120174, K192855); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Device Dislodged or Dislocated52
- Adverse Event Without Identified Device or Use Problem48
- Naturally Worn11
- Insufficient Device Problem Information5
- Appropriate Device Problem Term/Code Not Available4
- Device-Device Incompatibility1
- Difficult to Insert1
Patient problems
- Joint Dislocation53
- No Code Available34
- Joint Laxity20
- Pain18
- Unspecified Infection17
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available5
- Unspecified Tissue Injury4
- Insufficient Information3
Recent adverse event reports
The 25 most recent of 116 reports, in the reporter's own words as published by the FDA.
InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem
Patient problems: Unspecified Infection
MAUDE report 50923063 ↗InjurySep 24, 2026Adverse Event Without Identified Device or Use Problem
It was reported that patient presented with a painful right shoulder. The surgeon revised the glenosphere and poly.Affected Side: Unknown Shoulder.
Patient problems: Pain
MAUDE report 50881944 ↗InjurySep 14, 2026Device Dislodged or Dislocated
It was reported that the patient revised due to dislocation.DOI:(b)(6) 2020.DOR: (b)(6) 2026.Affected side: Left shoulder.
Patient problems: Joint Dislocation
MAUDE report 50752337 ↗InjurySep 1, 2026Device Dislodged or Dislocated
It was reported that patient used arm too soon after last surgery and dislocated shoulder. The Surgeon chose to size up in implant sizing.DOI-(b)(6)2026DOR-(b)(6)2026Affected side-Right shoulder.
Patient problems: Joint Dislocation
MAUDE report 50550204 ↗InjuryApr 6, 2026Device Dislodged or Dislocated
IT WAS REPORTED THAT THE SHOULDER DISLOCATED FROM A PREVIOUS SURGERY. PUT IN A RETENTIVE POLY. DOI: UNKNOWN. DOR: (B)(6) 2026. AFFECTED SIDE: RIGHT SHOULDER.
Patient problems: Joint Dislocation
MAUDE report 24794127 ↗InjuryMar 3, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO JOINT INSTABILITY. THE POLYETHYLENE COMPONENT WAS EXCHANGED FOR A RETENTIVE LINER, AND THE REMOVED POLY WAS DISPOSED OF WITHOUT ISSUE. THERE WAS NO HARM TO THE PATIENT AND NO DELAY TO THE CASE. NO FURTHER INFORMATION. DOI: (B)(6) 2026, DOR: (B)(6) 2026, AFFECTED SIDE: LEFT.
Patient problems: Joint Laxity
MAUDE report 24496760 ↗InjuryJan 23, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO LOOSENING OF SHOULDER. AFFECTED SIDE: RIGHT SHOULDER.
Patient problems: Joint Laxity
MAUDE report 24161880 ↗InjuryNov 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT WAS RETURNED TO SURGERY ON (B)(6) 2025, TO REMOVE BONE SPURS. ORIGINAL SURGERY WAS ON (B)(6) 16. THERE WAS NO LOOSENING, NO ISSUES OR CONCERNS WITH IMPLANTS. THE POLY AND GLENOSPHERE WERE EXCHANGED TO GET ACCESS TO REMOVE BONE SPUR. DOI: (B)(6) 2016. DOR: (B)(6) 2025. AFFECTED SIDE: UNKNOWN.
Patient problems: Bone Notching
MAUDE report 23597264 ↗InjuryNov 10, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT NEEDED MORE STABILITY. DR. SWAPPED GEOSPHERES FOR 42+2 AND ADDED +9 SPACER WITH 42+6 RETENTIVE LINER. DOI: UNKNOWN, DOR: (B)(6) 2025, AFFECTED SIDE: RIGHT SHOULDER.
Patient problems: Joint Laxity
MAUDE report 23513387 ↗InjuryJul 7, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE PATIENT WAS DISLOCATING, SO THE SURGEON DECIDED TO PUT IN A LARGER GLENOSPHERE, HUMERAL SPACER, AND A DIFFERENT POLYETHYLENE CUP. DOI: (B)(6) 2023. DOR: (B)(6) 2025. AFFECTED SIDE: RIGHT SHOULDER.
Patient problems: Joint Dislocation
MAUDE report 22412404 ↗InjuryMar 24, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT FELL AND NEEDED MORE LATERALIZED IMPLANTS FOR STABILITY. DOI: UNKNOWN. DOR: UNKNOWN. AFFECTED SIDE: UNKNOWN SHOULDER SIDE.
Patient problems: Fall; Joint Laxity
MAUDE report 21678324 ↗InjuryMar 24, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT DISLOCATION RIGHT SHOULDER FOLLOWING IMPLANTATION OF REVERSE TOTAL SHOULDER. DOI: (B)(6) 2025. DOR: (B)(6) 2025. AFFECTED SIDE: RIGHT SHOULDER.
Patient problems: Joint Dislocation
MAUDE report 21678003 ↗InjuryFeb 6, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO INCREASE TENSION IN SHOULDER. DOI: UNKNOWN. AFFECTED SIDE: UNKNOWN.
Patient problems: Joint Laxity; Unspecified Tissue Injury
MAUDE report 21322635 ↗InjuryDec 3, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT PATIENT WAS BROUGHT BACK FOR DISLOCATION. DOI- (B)(6) 2024, DOR- (B)(6) 2024, AFFECTED SIDE- LEFT SHOULDER.
Patient problems: Joint Dislocation
MAUDE report 20832385 ↗InjuryNov 21, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT PATIENT HAD REVISED FOR INSTABILITY. DOI: (B)(6) 2024. DOR: (B)(6) 2024. AFFECTED SIDE- RIGHT SHOULDER.
Patient problems: Joint Laxity
MAUDE report 20747968 ↗InjuryOct 9, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
ADDITIONAL INFORMATION RECEIVED THAT THERE WAS NO SURGICAL DELAY.
Patient problems: Joint Dislocation
MAUDE report 20406164 ↗InjuryJul 11, 2024Naturally Worn; Naturally Worn
IT WAS REPORTED THAT PATIENT WAS REVISED DUE TO GLENOID WAS GROSSLY LOOSE. IMPLANT WAS PLACED HIGH ON THE GLENOID FACE. CEMENTED STEM. ENTIRE GLENOID CONSTRUCT WAS PULLED OUT WITH AND OSTEOTOME AND KOCHER. METAGLENE TWO LOCKING SCREWS AND 38 GLENOSPHERE. THE SUPERIOR LOCKING SCREW WAS BROKEN AND MOST OF IT WAS LEFT IN THE PATIENTS SUPERIOR GLENOID. POLY WAS REMOVED AS WELL AND HAD WEAR ON THE MEDIAL HALF OF THE RIM OF THE POLY. THEY DID NOT HAVE ANY IMPLANT RECORDS AND THE REF NUMBERS WERE NOT VISIBLE TO IDENTIFY SIZES AND LENGTHS. DOI: UNKNOWN. DOR: (B)(6) 2024. AFFECTED SIDE: LEFT SHOULDER.
Patient problems: Insufficient Information
MAUDE report 19722248 ↗InjuryJun 3, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT HAD INITIAL SURGERY DONE (B)(6) 2016 AT THE HOSPITAL BY THE SURGEON. THE PATIENT'S SHOULDER WAS INFECTED, AND THE SURGEON FELT THE PATIENT WOULD BENEFIT FROM SINGLE STAGE WASHOUT. EPI, POLY, AND GLENOSPHERE WERE REMOVED. THE SHOULDER WAS WASHED OUT AND NEW COMPONENTS WERE IMPLANTED. STABILITY WAS CONFIRMED AND NO DELAYS OR ADVERSE EVENTS TOOK PLACE. DOI: (B)(6) 2016, DOR: (B)(6) 2024 , AFFECTED SIDE: RIGHT SHOULDER.
Patient problems: Unspecified Infection
MAUDE report 19449371 ↗InjuryJun 3, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT PATIENT WAS REVISED DUE TO INFECTED REVERSE TOTAL SHOULDER. ALL IMPLANTS WERE REMOVED AND A SPACER WAS IMPLANTED. THERE WAS NO LOOSENING. DOI: (B)(6) 2023. DOR: (B)(6) 2024. AFFECTED SIDE: LEFT SHOULDER.
Patient problems: Unspecified Infection
MAUDE report 19446616 ↗InjuryMay 1, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
THE PATIENT HAD REVISED REVERSE FOR DISLOCATION. PATIENT HAD A REVERSE FOR A FRACTURE AWHILE BACK. UNKNOWN SURGICAL DELAY. DOI: UNKNOWN DOR: (B)(6) 2024 AFFECTED SIDE: RIGHT SHOULDER.
Patient problems: Joint Dislocation
MAUDE report 19222635 ↗InjuryFeb 15, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
THIS CASE LOGGED TO CAPTURE THE PREVIOUS REVISION MENTIONED IN (B)(4). SURGEON DID ORIGINAL OPERATION (B)(6) 2020 AND FIRST REVISION 2 MONTHS LATER FOR DISLOCATION.
Patient problems: Joint Dislocation
MAUDE report 18712458 ↗InjuryJan 2, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT WAS REVISED DUE TO INSTABILITY. DOI: UNKNOWN, DOR: (B)(3) 2023, AFFECTED SIDE: UNKNOWN.
Patient problems: Joint Laxity
MAUDE report 18424882 ↗InjuryDec 1, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE LINER EXCHANGE WAS DONE. PATIENT EXPERIENCED DISLOCATION, SO THE PHYSICIAN ADDED A BIGGER POLY. THERE WAS NO SURGICAL DELAY. DOI: (B)(6) 2023. DOR: (B)(6) 2023. AFFECTED SIDE: RIGHT SHOULDER.
Patient problems: Joint Dislocation
MAUDE report 18246298 ↗InjuryNov 30, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT PATIENT HAD BEEN UNSTABLE AND SEVERE KYPHOSIS EVIDENT WHICH COULD POTENTIALLY BE A CAUSE OF THE INSTABILITY. DOI: (B)(6) 2018. DOR: NOV 21, 2023. AFFECTED SIDE: RIGHT SHOULDER.
Patient problems: Joint Laxity
MAUDE report 18240446 ↗InjuryNov 15, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT PATIENT HAD A RSA FOR A PROXIMAL HUMERAL FRACTURE ON (B)(6) 2023. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY. AFTER SURGERY, INFECTION WAS CONFIRMED. A REMOVAL SURGERY WILL BE PERFORMED AS SOON AS A PREPARATION IS READY. DOI: (B)(6) 2023 RIGHT SHOULDER.
Patient problems: Unspecified Infection
MAUDE report 18145786 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K062250, K081620, K120174, K192855. One submission often covers many models, so check the product and lot details in each record.
- Z-0834-2018 · Aug 31, 2017 · TerminatedDELTA XTEND" Reverse Shoulder System Modular Centered and Eccentric Epiphysis Implants intended for use in total shoulder or hemi-shoulder replacement procedures in patients with n
- Z-0833-2018 · Aug 31, 2017 · TerminatedDELTA XTEND" Reverse Shoulder System Modular Centered and Eccentric Epiphysis Implants are intended for use in total shoulder or hemi-shoulder replacement procedures in patients wi
- Z-0533-2009 · Jul 10, 2008 · TerminatedDepuy DELTA Xtend, Standard Humerus PE Cup, Dia 38/+6 STD, sterile, Depuy France; REF 130738206. The device is used as a shoulder implant.
- Z-0538-2009 · Jul 10, 2008 · TerminatedDepuy DELTA Xtend, Standard Humeral PE Cup, Dia 42/+9 STD, sterile, Depuy France; REF 130742209. The device is used as a shoulder implant.
- Z-0536-2009 · Jul 10, 2008 · TerminatedDepuy DELTA Xtend, Standard Humeral PE Cup, Dia 42/+3 STD, sterile, Depuy France; REF 130742203. The device is used as a shoulder implant.
- Z-0539-2009 · Jul 10, 2008 · TerminatedDepuy DELTA Xtend, Standard Humeral PE Cup, Dia 42+6R RET, sterile, Depuy France; REF 130742106. The device is used as a shoulder implant.
- Z-0537-2009 · Jul 10, 2008 · TerminatedDepuy DELTA Xtend, Standard Humeral PE Cup, Dia 42/+6 STD, sterile, Depuy France; REF 130742206. The device is used as a shoulder implant.
- Z-0534-2009 · Jul 10, 2008 · TerminatedDepuy DELTA Xtend, Standard Humeral PE Cup, Dia 38/+9 STD, sterile, Depuy France; REF 130738209. The device is used as a shoulder implant.
- Z-0535-2009 · Jul 10, 2008 · TerminatedDepuy DELTA Xtend, Standard Humeral PE Cup, Dia 38/+6R RET, sterile, Depuy France; REF 130738106. The device is used as a shoulder implant.
- Z-0532-2009 · Jul 10, 2008 · TerminatedDepuy DELTA Xtend, Standard humeral PE Cup Dia 38/+3 STD, sterile, Depuy France; REF 130738203. The device is used as a shoulder implant.
Other PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED devices
- DJO SURGICAL 509-02-032Encore Medical, L.P. · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED364 reports
- DELTA XTEND 130790036Depuy (Ireland) · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED171 reports
- DELTA XTEND 130770018Depuy (Ireland) · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED159 reports
- TRIGEN 71118200Smith & Nephew, Inc. · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED123 reports
- DJO SURGICAL 509-02-432Encore Medical, L.P. · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED119 reports
- COMPREHENSIVE SHOULDER SYSTEM 113633Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED2 recalls4 reports
- DELTA XTEND 130742206Depuy (Ireland) · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED99 reports
- GLOBAL UNITE 1100-12-100Depuy (Ireland) · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED1 recall29 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page