Skip to content

BD PrepStain System, Model Number: 799-14000-00 recallZ-1090-2009 · Bd Diagnostic Systems Tripath · initiated Sep 26, 2008

Terminated All recalls posted in March 2009

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Bd Diagnostic Systems Tripath or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Software Error in the non-gyn program of some devices may cause non-gyn samples to be incorrectly transferred from centrifuge tubes to slides potentially result in mismatched patient results.

FDA root cause category: Software change control

What the company is doing

BD Diagnostics began contacting known non-gyn customers by phone on 09/26/2008. Customers were notified by Urgent - Important Product Information letter on 09/29/2008. They were advised to stop using the system until they complete a list of actions provided in the letter. A telephone contact number for Technical Support was provided in the letter. A permanent software based solution will be provided to customers when completed. They anticipate the software will be available in October 2008.

Recall details

Recall number
Z-1090-2009
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Sep 26, 2008
Posted by the FDA
Mar 19, 2009
Recalling firm
Bd Diagnostic Systems Tripath
Quantity
662 units
Product code
MKQ
FDA submissions
P970018
Distribution
Worldwide Distribution

Product description

BD PrepStain System, Model Number: 799-14000-00; Catalog Number: 05CR000021, 05CR00021R, 799-13000-00, 799-13000-00R, 799-14000-00 and 799-14000-00R; Catalog Number: 05CR000021, 05CR000021R, 799-13000-00, 799-13000-00R, 799-14000-00 and 7914000-00R. PrepStain is a liquid-based thin layer cell preparation device for preparation and staining of Prep smear slides from cervical specimens in a liquid based format (gyn use) and of cytology examination slides from non-cervical specimens such as urine and lung lavage.

Code information (models, lots, serial numbers)

GYN software version 1.2, Non-GYN version 2.80; GYN software version 1.3.0.1, Non-GYN software version 2.80.

Other devices under the same FDA submission

13 devices cleared under P970018 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.