8 Fr 30cc Ultra 8 IAB recallZ-1212-2009 · Arrow International, Inc. · initiated Feb 2, 2009
Terminated All recalls posted in April 2009
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Arrow International, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
A fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.
FDA root cause category: Process design
What the company is doing
An "Urgent Medical Device Recall" letter dated February 2, 2009 was issued to consignees via US Certified Mail to notify users of the problem. Users were instructed to check their inventory for the affected products, quarantine, and return the them to Arrow International, Inc. Customers were also instructed to follow the "Mitigation Directions" in circumstances where alternate Intra-Aortic Balloon (IAB) catheters/adaptors or other forms of therapy are not available (until receipt of available product). Canadian accounts were notified on February 9, 2009 by courier and International accounts notified on 6 February 2009 and 10 February 2009 via Registered Post/UPS. Direct questions to your local Sales Representative or the Arrow International Customer Service at 1-800-523-8446 (US) or 1-919-361-4062 (outside US). Press Release issued April 13, 2009 is posted to firm's webiste: http://www.arrowintl.com/iab_recall.asp.
Recall details
- Recall number
- Z-1212-2009
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Feb 2, 2009
- Posted by the FDA
- Apr 16, 2009
- Recalling firm
- Arrow International, Inc.
- Quantity
- 4,258 units
- Product code
- DSP
- FDA submissions
- K000729, K010330, K021462, K040801
- Distribution
- Worldwide Distribution: USA, Argentina, Australia, Taiwan, Singapore, New Zealand, India, Brazil, Belgium, Korea, Peru, Finland, and Denmark.
Product description
8 Fr 30cc Ultra 8 IAB, IAB-05830-U, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counter-pulsation therapy, whereby balloon inflation in the aorta during diastole and deflation during systole increase blood supply to the heart muscle and decrease work of the left ventricle.
Code information (models, lots, serial numbers)
Lot/Serial Numbers: MF6122058 MF7065640 MF7076257 MF7076378 MF7097392 MF7118003 MF7118004 MF7118219 MF7128387 MF7128471 MF7128504 MF7128505 MF8018907 MF8018997 MF8030122 MF8039664 MF8039878 MF8040324 MF8040325 MF8040638 MF8050911 MF8061125 MF8061176 MF8061334 MF8061500 MF8061615 MF8071782 MF8072074 MF8082267 MF8082584 MF8092783 MF8092784 MF8093000 MF8093001 MF8103155 MF8103262 MF8103263 MF8124064 MF8124178 MF8124064 MF8124307 and MF8124408.
Other devices under the same FDA submission
14 devices cleared under K000729, K010330, K021462, K040801 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- ARROW IPN000260Teleflex Incorporated · System, balloon, intra-aortic and control4 deaths698 reports
- ARROW IPN000257Teleflex Incorporated · System, balloon, intra-aortic and control3 deaths233 reports
- ARROW IPN915322Teleflex Incorporated · System, balloon, intra-aortic and control1 recall29 reports
- ARROW IPN039793Teleflex Incorporated · System, balloon, intra-aortic and control62 reports
- ARROW IPN915321Teleflex Incorporated · System, balloon, intra-aortic and control1 recall7 reports
- ARROW IPN915320Teleflex Incorporated · System, balloon, intra-aortic and control1 recall0 reports
- ARROW IPN000255Teleflex Incorporated · System, balloon, intra-aortic and control32 reports
- ARROW IPN039788Teleflex Incorporated · System, balloon, intra-aortic and control29 reports
- ARROW IPN039782Teleflex Incorporated · System, balloon, intra-aortic and control23 reports
- ARROW IPN039778Teleflex Incorporated · System, balloon, intra-aortic and control16 reports
- ARROW IPN038075Teleflex Incorporated · System, balloon, intra-aortic and control12 reports
- ARROW IPN038074Teleflex Incorporated · System, balloon, intra-aortic and control7 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 51080).
- Z-1210-2009 · Terminated8 Fr 30cc Narrow Flex IAB Catheter Kit with Generica Adapters Product Number: IAB-04830-U, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counte…
- Z-1211-2009 · Terminated8 Fr 40cc Narrow Flex IAB Catheter Kit with Generic Adapters Product Number: IAB-04840-U, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counter…
- Z-1213-2009 · Terminated8 Fr 40cc Ultra 8 IAB, IAB-05840-U, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counter-pulsation therapy, whereby balloon inflation in the a…
- Z-1214-2009 · Terminated7.5 Fr 30cc UltraFlex IAB, IAB-06830-U, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counter-pulsation therapy, whereby balloon inflation in …
- Z-1215-2009 · Terminated7.5 Fr 40cc UltraFlex IAB, IAB-06840-U, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counter-pulsation therapy, whereby balloon inflation in t…
- Z-1216-2009 · Terminated8 Fr 30cc FIBEROPTIX IAB, IAB-05830-LWS, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counter-pulsation therapy, whereby balloon inflation in …
- Z-1217-2009 · Terminated8 Fr 40cc FIBEROPTIX IAB, IAB-05840-LWS, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counter-pulsation therapy, whereby balloon inflation in …
- Z-1218-2009 · Terminated7 FR 30CC Rediguard IAB, IAB-S730C, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counter-pulsation therapy, whereby balloon inflation in the a…
- Z-1219-2009 · Terminated8 FR 40CC Rediguard IAB, IAB-S840C, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counter-pulsation therapy, whereby balloon inflation in the a…
- Z-1220-2009 · Terminated9 FR 50CC Rediguard IAB, IAB-R950-U, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counter-pulsation therapy, whereby balloon inflation in the …
- Z-1221-2009 · TerminatedARROW 40cc Drive Line Kits, REF IAK-02692, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counter-pulsation therapy, whereby balloon inflation …
- Z-1222-2009 · TerminatedARROW 50cc Drive Line Kits, REF IAK-02693, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counter-pulsation therapy, whereby balloon inflation …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page