TRESTLE LUXE Anterior Cervical Plating System, Product code… recallZ-2273-2012 · Alphatec Spine, Inc. · initiated May 29, 2012
Class II Terminated All recalls posted in August 2012
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Alphatec Spine, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
On May 9th 2012 Alphatec Spine received a complaint which reported that the distal tip of the Trestle Luxe Slide Alignment Tool (Part Number 71721), does not properly fit into the hexalobe screw on the Trestle Luxe Plate.
FDA root cause category: Device Design
What the company is doing
Alphatec Spine sent an "URGENT-MEDICAL DEVICE RECALL NOTIFICATION" letter dated May 29, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. A form was attached for customers to complete and return via fax to 760-431-0289. Contact Alphatec Spine's Customer Service Department at [email protected] questions regarding this recall.)
Recall details
- Recall number
- Z-2273-2012
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- May 29, 2012
- Posted by the FDA
- Aug 27, 2012
- Recalling firm
- Alphatec Spine, Inc.
- Quantity
- 38
- Product code
- KWQ
- FDA submissions
- K102820
- Distribution
- Nationwide Distribution.
Product description
TRESTLE LUXE Anterior Cervical Plating System, Product code KWQ. 510(k) K102820. The TRESTLE LUXE Anterior Cervical Plating System is a temporary device used to stabilize the cervical spine during bone fusion development. Device implants include a range of plate sizes and bone screws to provide the versatility required for the specific indications noted. Fixation is achieved by means of a rigid plate that is surgically attached to the spine with bone screws. The TRESTLE LUXE Anterior Cervical Plating System is intended for use in the anterior cervical spine (C2-C7). Patients with trauma (including fractures), spondylolisthesis, and pseudoarthrosis, degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spinal stenosis, tumors, and failed previous fusion.
Code information (models, lots, serial numbers)
Product code KWQ. 510(k) K102820 Part Number 71721. Lot numbers: 6629701, 6594701, 6434301, 6434302, 6434302R1.
Other devices under the same FDA submission
23 devices cleared under K102820 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Trestle Luxe 71002-034Alphatec Spine, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY6 reports
- Trestle Luxe 71004-064Alphatec Spine, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY3 reports
- Trestle Luxe 71540-16Alphatec Spine, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY3 reports
- Trestle Luxe 71001-020Alphatec Spine, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY2 reports
- Trestle Luxe 71002-026Alphatec Spine, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY2 reports
- Trestle Luxe 71002-032Alphatec Spine, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY2 reports
- Trestle Luxe 71002-037Alphatec Spine, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY2 reports
- Trestle Luxe 71240-14Alphatec Spine, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY2 reports
- TRESTLE LUXE 711-0234Alphatec Spine, Inc. · Appliance, fixation, spinal intervertebral body1 report
- TRESTLE LUXE 722-14014Alphatec Spine, Inc. · Appliance, fixation, spinal intervertebral body1 report
- Trestle Luxe 71001-022Alphatec Spine, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 report
- Trestle Luxe 71002-028Alphatec Spine, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page