Biosense Webster NaviStar RMT Diagnostic/Ablation Steerable… recallZ-0369-2010 · BioSense Webster, Inc. · initiated Oct 2, 2008
Terminated All recalls posted in November 2009
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from BioSense Webster, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Biosense Webster initiated the recall because of the potential concern with one of the Printed Circuit Board (PCB) components used in the NaviStar RMT Catheters, which may cause the ablation catheter icon to shift on the map viewer when RF energy is being applied. Therefore the user, attempting a linear ablation in the location displayed on the CARTO RMT screen, may ultimately apply RF energy to u
FDA root cause category: Process control
What the company is doing
The firm sent an "Urgent Medical Device Correction" letter to consignees on October 6, 2008 informing consignees. Users were advised to pay "close attention to the location of the ablation catheter icon on the map viewer when applying RF energy." If consignees suspect that they have a defective NAVISTAR RMT Catheter, they were instructed to notify Biosense Webster and to return the defective catheter for evaluation and refund/replacement. Customer were instructed to contact their local Biosense Webster representative for any questions or further information. No return reply was requested.
Recall details
- Recall number
- Z-0369-2010
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Oct 2, 2008
- Posted by the FDA
- Nov 19, 2009
- Recalling firm
- BioSense Webster, Inc.
- Quantity
- 2180 Units
- Product code
- LPB
- FDA submissions
- P990025
- Distribution
- CA, CT, FL, GA, IA, IL, IN, KS, KY, MA, MI, MN, MO, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, and WI
Product description
Biosense Webster NaviStar RMT Diagnostic/Ablation Steerable Tip 8mm DS Catheters; Part Numbers: D-1259-01 & D-1259-02
Code information (models, lots, serial numbers)
All lot numbers
Other devices under the same FDA submission
13 devices cleared under P990025 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- EZ Steer NAV BN7TCDF4LBiosense Webster Inc · Cardiac ablation percutaneous catheter2 deaths105 reports
- nGEN Generator D138402Biosense Webster Inc · Cardiac ablation percutaneous catheter2 deaths46 reports
- NaviStar NS7TCDL174HSBiosense Webster Inc · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING1 death44 reports
- NaviStar NS7TCFL174HSBiosense Webster Inc · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING1 death27 reports
- EZ Steer NAV BN7TCFJ4LBiosense Webster Inc · Cardiac ablation percutaneous catheter24 reports
- NaviStar NS7TCBL174HSBiosense Webster Inc · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING5 reports
- NaviStar NS7TCJL174HSBiosense Webster Inc · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING4 reports
- EZ Steer NAV BN7TCFF4LBiosense Webster Inc · Cardiac ablation percutaneous catheter3 reports
- NaviStar RMT NR7TCS4YUBiosense Webster Inc · Cardiac ablation percutaneous catheter1 death3 reports
- EZ Steer NAV BN7TCJJ4LBiosense Webster Inc · Cardiac ablation percutaneous catheter2 reports
- NaviStar RMT NR7TCS4YBiosense Webster Inc · Cardiac ablation percutaneous catheter2 reports
- EZ Steer NAV BN7TCDD4LBiosense Webster Inc · Cardiac ablation percutaneous catheter1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page