RF Ablation System Foot Switch, used with the EPT-1000 RF… recallZ-0973-2011 · Boston Scientific Corp · initiated Jul 14, 2009
Terminated All recalls posted in January 2011
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Boston Scientific Corp or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Product malfunction causes inadvertent delivery of energy.
FDA root cause category: Nonconforming Material/Component
What the company is doing
The firm, Boston Scientific, sent an "URGENT: Medical Device Recall" letter dated January 14, 2009, with attached recall instructions to all customers. The letter described the product, problem, and the action to be taken by the customers. The customers were instructed to complete and return the Reply Verification Tracking Form and Obtain an RGA via fax to 408-957-6204; and return the recalled product using BSC Federal Express Number 1-809-4713-2 via second day delivery to Boston Scientific Corporation, Distribution Center, 500 Commander Shea Blvd., Quincy, MA 02171. Boston Scientific Corporation sales representatives will contact each customer. If you have any questions, please contact 408-935-4952 or email: [email protected].
Recall details
- Recall number
- Z-0973-2011
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Jul 14, 2009
- Posted by the FDA
- Jan 27, 2011
- Recalling firm
- Boston Scientific Corp
- Quantity
- 1044 units
- Product code
- LPB
- FDA submissions
- P920047
- Distribution
- Worldwide distribution
Product description
RF Ablation System Foot Switch, used with the EPT-1000 RF Generator, Foot Switch Model Number UPN M0048400, and in kits with UPNs M0044810 and M0044811. Intended use: The EPT-1000xp Cardiac Ablation Controllers and Foot Switch are indicated for use in cardiac ablation procedures with standard and high power BSC cardiac ablation Catheters.
Code information (models, lots, serial numbers)
Show the full code information (9k characters)
Serial numbers 49615 56781 57259 59544 61161 62244 62643 62694 62881 63271 49889 56794 57260 59545 61162 62245 62644 62695 62882 63272 49895 56800 57261 59549 61163 62246 62645 62696 62883 63273 50100 56801 57263 59550 61164 62247 62646 62698 62884 63274 50108 56805 57267 59552 61165 62248 62647 62699 62885 63275 50531 56806 57613 59555 61166 62249 62648 62844 62886 63276 50807 56810 58739 59560 61167 62250 62649 62845 62887 63277 51710 56812 58742 59564 61168 62251 62650 62846 62888 63278 52362 56814 58743 59565 61170 62253 62651 62847 62889 63279 52363 56815 58745 59567 61171 62255 62653 62848 62890 63280 52364 56816 58746 59568 61173 62256 62654 62849 62891 63281 52365 56817 59156 59569 61870 62257 62655 62850 62892 63282 52369 57190 59186 59819 61871 62258 62657 62851 62893 63283 52370 57193 59187 61116 61873 62259 62658 62852 62894 63284 52371 57194 59189 61117 61875 62260 62659 62853 62895 63285 52377 57199 59190 61118 61876 62262 62660 62854 62896 63286 52394 57200 59191 61119 61880 62265 62663 62855 62897 63287 52395 57201 59192 61120 61881 62266 62665 62856 62898 63288 52397 57203 59193 61122 61882 62267 62666 62857 62899 63289 53543 57209 59194 61123 61885 62268 62667 62858 62900 63290 54061 57210 59195 61124 61886 62269 62668 62859 62901 63291 54561 57211 59196 61125 61887 62270 62669 62860 62902 63292 54562 57212 59197 61126 61889 62271 62670 62861 62903 63293 54563 57213 59198 61128 61890 62272 62671 62862 62904 63295 54908 57214 59199 61129 61891 62273 62672 62863 62905 63296 54909 57218 59200 61131 61892 62274 62673 62864 62906 63297 54924 57222 59201 61132 61893 62275 62674 62865 62907 63298 55202 57224 59202 61133 61895 62276 62675 62866 62908 63299 55203 57227 59203 61136 61896 62277 62676 62867 62909 63300 55204 57229 59204 61139 61897 62279 62678 62868 62910 63301 55205 57230 59205 61140 61899 62501 62680 62869 62911 63302 55207 57231 59207 61143 61902 62627 62681 62870 62912 63303 55220 57234 59208 61144 61903 62630 62682 62871 62913 63304 56764 57237 59209 61145 61904 62631 62683 62872 62914 63305 56768 57238 59211 61147 61906 62632 62684 62873 62915 63306 56769 57239 59214 61148 62237 62633 62686 62874 62916 63307 56770 57241 59535 61154 62238 62634 62687 62875 62917 63308 56771 57242 59536 61155 62239 62635 62689 62876 62918 63309 56774 57243 59537 61156 62240 62636 62690 62877 63267 63310 56777 57246 59540 61157 62241 62637 62691 62878 63268 63311 56778 57255 59541 61158 62242 62641 62692 62879 63269 63312 56779 57256 59543 61159 62243 62642 62693 62880 63270 63313 63314 63357 64253 64295 64884 64926 90574 91230 91403 91510 63315 63358 64254 64296 64885 64927 90575 91231 91404 91511 63316 63359 64255 64297 64886 64928 90576 91232 91405 91512 63317 63360 64256 64298 64887 64929 90577 91233 91406 91513 63318 63361 64257 64299 64888 64930 90578 91234 91407 91514 63319 63362 64258 64300 64889 64931 90579 91235 91408 91515 63320 63363 64259 64301 64890 64932 90580 91236 91409 91516 63321 63364 64260 64302 64891 64933 90581 91237 91410 91517 63322 63365 64261 64303 64892 64934 90582 91238 91411 91518 63323 63366 64262 64304 64893 64935 90583 91239 91412 91519 63324 64220 64263 64305 64894 64936 90584 91240 91413 91520 63325 64221 64264 64306 64895 64937 90585 91241 91414 91521 63326 64222 64265 64307 64896 64938 90586 91242 91415 91522 63327 64223 64266 64308 64897 64939 90587 91243 91416 91523 63328 64224 64267 64309 64898 64940 90588 91244 91417 91524 63329 64225 64268 64310 64899 64941 90589 91245 91418 91525 63330 64226 64269 64311 64900 64942 90590 91246 91419 91898 63331 64227 64270 64312 64901 64943 91006 91247 91420 94402 63332 64228 64271 64313 64902 64945 91007 91248 91421 94403 63333 64229 64272 64314 64903 64946 91008 91249 91422 94404 63334 64230 64273 64315 64904 64947 91009 91250 91423 94405 63335 64231 64274 64316 64905 64948 91010 91251 91424 94406 63336 64232 64275 64317 64906 64949 91011 91252 91425 94407 63337 64233 64276 64318 64907 64950 91012 91253 91426 94408 63339 64234 64277 64319 64908 64951 91013 91254 91427 94409 63340 64235 64278 64867 64909 64952 91014 91386 91428 94410 63341 64236 64279 64868 64910 64953 91015 91387 91429 94411 63342 64237 64280 64869 64911 64954 91016 91388 91430 94412 63343 64238 64281 64870 64912 64955 91017 91389 91431 94413 63344 64239 64282 64871 64913 64956 91018 91390 91432 94414 63345 64241 64283 64872 64914 64957 91019 91391 91433 94415 63346 64242 64284 64873 64915 64958 91020 91392 91434 94416 63347 64243 64285 64874 64916 64959 91021 91393 91435 94417 63348 64244 64286 64875 64917 64960 91022 91394 91501 94418 63349 64245 64287 64876 64918 90566 91023 91395 91502 94419 63350 64246 64288 64877 64919 90567 91024 91396 91503 94420 63351 64247 64289 64878 64920 90568 91025 91397 91504 94421 63352 64248 64290 64879 64921 90569 91026 91398 91505 94422 63353 64249 64291 64880 64922 90570 91027 91399 91506 94423 63354 64250 64292 64881 64923 90571 91028 91400 91507 94424 63355 64251 64293 64882 64924 90572 91029 91401 91508 94425 63356 64252 64294 64883 64925 90573 91030 91402 91509 94426 94427 94469 98621 100995 101067 101109 94428 94470 98622 100996 101068 101110 94429 94471 98623 100997 101069 101111 94430 94472 98624 100998 101070 101112 94431 98583 98625 100999 101071 101113 94432 98584 98626 101000 101072 101114 94433 98585 98627 101001 101073 102001 94434 98586 98628 101002 101074 102002 94435 98587 98629 101003 101075 102003 94436 98588 98630 101004 101076 102004 94437 98589 98631 101005 101077 102005 94438 98590 98632 101006 101078 102006 94439 98591 98820 101007 101079 102007 94440 98592 98821 101008 101080 102008 94441 98593 98822 101009 101081 102009 94442 98594 98823 101010 101082 102010 94443 98595 98824 101011 101083 102011 94444 98596 98825 101012 101084 102012 94445 98597 98826 101013 101085 102013 94446 98598 98827 101014 101086 102014 94447 98599 98828 101015 101087 102015 94448 98600 98829 101016 101088 102016 94449 98601 98830 101017 101089 102017 94450 98602 98831 101018 101090 102018 94451 98603 98832 101019 101091 102019 94452 98604 98833 101020 101092 102020 94453 98605 98834 101021 101093 102021 94454 98606 98835 101022 101094 102022 94455 98607 98836 101023 101095 102023 94456 98608 98837 101024 101096 102024 94457 98609 98838 101025 101097 102025 94458 98610 98839 101026 101098 102026 94459 98611 100985 101027 101099 102027 94460 98612 100986 101028 101100 102028 94461 98613 100987 101029 101101 102029 94462 98614 100988 101030 101102 102030 94463 98615 100989 101031 101103 102031 94464 98616 100990 101032 101104 102032 94465 98617 100991 101033 101105 102033 94466 98618 100992 101034 101106 102034 94467 98619 100993 101065 101107 102035 94468 98620 100994 101066 101108 102036 102037 102038 102039 102040 102041 102042 102043 102044 102045 102046 102047 102048 102049 102050 7E272 A1335 2287657 3049205 3915571 031299-006 032900-030 070501-005 2318282 3049207 3915572 031299-008 041601-001 071601-006 2318283 3049209 3931890 031299-010 041601-002 071601-009 2318284 3049210 010702-008 031299-012 041601-003 071601-010 2415609 3049213 010702-009 031299-014 041601-004 081701-001 2415610 3076854 010702-011 032900-004 041601-005 082301-031 2520150 3076857 011502-003 032900-005 050801-001 082301-032 2520151 3076858 011502-004 032900-006 050801-002 082301-033 2520152 3217687 011502-005 032900-007 050801-003 082301-034 2520153 3217688 020199-001 032900-008 050801-004 090399-010 2891036 3295978 020199-003 032900-010 050801-005 091801-002 2891037 3295982 020199-004 032900-011 061801-001 111599-001 2891038 3446758 020199-005 032900-012 061801-002 111599-009 2891189 3456156 020199-007 032900-013 061801-003 111599-010 2891190 3473195 020199-008 032900-014 061801-004 111599-011 2973043 3473196 020199-009 032900-015 061801-005 11599-013 2973044 3473198 020199-012 032900-016 062601-006 120700-001 2973045 3473201 020199-020 032900-017 062601-007 120700-002 2973046 3473206 020199-026 032900-018 062601-008 120700-003 2973047 3492930 020402-029 032900-019 062601-009 120700-004 3025916 3527773 020402-030 032900-020 062601-010 120700-005 3025917 3527774 020402-27 032900-021 070501-001 120700-006 3025918 3527775 021102-002 032900-027 070501-002 120700-007 3025920 3527776 031299-003 032900-028 070501-003 120700-008 3025921 3527777 031299-004 032900-029 070501-004 120700-009 120700-010 121198-002 121198-003 2787163 3188652 010702-006 050101-004 062700-014 092601-003 102502-017 2787165 3188653 010803-001 050101-005 062700-015 092701-002 102502-018 2787166 3295983 021302-001 060502-001 072501-001 092701-003 103001-001 2787167 3295988 021302-002 060502-002 072501-002 092701-004 103001-002 2787168 3411186 021302-003 060502-003 072501-003 092701-005 103001-003 2891036 3411389 021302-004 060502-004 072501-004 101601-001 103001-004 2891189 3446759 021302-005 060502-005 072501-005 101601-002 103001-005 3019925 3456146 032900-012 060502-006 080502-004 101601-003 110701-001 3019926 3456152 041001-001 060502-007 080502-005 101601-004 110701-002 3188645 3527778 041001-002 060502-008 080502-006 101601-005 110701-003 3188646 3527779 041601-001 060502-009 080502-007 102502-011 121901-001 3188647 3527780 041601-002 060502-010 080502-008 102502-012 121901-002 3188648 010603-008 041601-003 062700-005 090399-007 102502-013 121901-003 3188649 010603-009 050101-001 062700-011 090501-001 102502-014 121901-004 3188650 010603-011 050101-002 062700-012 090501-002 102502-015 121901-005 3188651 010603-012 050101-003 062700-013 090501-004 102502-016
Other devices under the same FDA submission
15 devices cleared under P920047 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Blazer® II HTD M0045031TH0Boston Scientific Corporation · Cardiac ablation percutaneous catheter14 reports
- IntellaNav™ ST M004R5031TH0Boston Scientific Corporation · Cardiac ablation percutaneous catheter10 reports
- Blazer® II HTD M0045031THK20Boston Scientific Corporation · Cardiac ablation percutaneous catheter7 reports
- IntellaNav™ ST M004R5031THK20Boston Scientific Corporation · Cardiac ablation percutaneous catheter5 reports
- Blazer® II M0045031T0Boston Scientific Corporation · Cardiac ablation percutaneous catheter4 reports
- Blazer® II M0045031TK20Boston Scientific Corporation · Cardiac ablation percutaneous catheter3 reports
- Blazer Prime® HTD M004P5031TH0Boston Scientific Corporation · Cardiac ablation percutaneous catheter3 reports
- Blazer® II M0045031TK10Boston Scientific Corporation · Cardiac ablation percutaneous catheter2 reports
- Blazer Prime® HTD M004P5031THN40Boston Scientific Corporation · Cardiac ablation percutaneous catheter2 reports
- Blazer® II HTD M0045031THN40Boston Scientific Corporation · Cardiac ablation percutaneous catheter1 report
- Blazer® II M0045031TL0Boston Scientific Corporation · Cardiac ablation percutaneous catheter1 report
- Blazer® II M0045770TN40Boston Scientific Corporation · Cardiac ablation percutaneous catheter1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page