HeartMate 12 volt sealed lead acid Battery Clips that are… recallZ-0422-2010 · Thoratec Corp · initiated Oct 14, 2009

Terminated All recalls posted in November 2009

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Thoratec Corp or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Patient reports alerted Thoratec that during the process of disconnecting power leads from the battery clips, the threaded connector of the battery clip housing was observed to be loose, and in some cases dislodged from the battery clip housing. The problem affects HeartMate 12V SLA Battery Clips including those used for the HeartMate II LVAS and HeartMate XVE systems. If not detected and handled

FDA root cause category: Other

What the company is doing

Firm has issued a notification letter to consignees, dated October 14, 2009 asking them to return proof of receipt, and to remove all affected unused devices from stock, and asking ongoing patients to return to hospitals for inspection, replacement if needed, and retraining on the process of switching power sources and safely switching to a back-up system controller. Contact Thoratec Customer Services at 1-800-456-1477 for any questions or concerns about this issue.

Recall details

Recall number
Z-0422-2010
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Oct 14, 2009
Posted by the FDA
Nov 24, 2009
Recalling firm
Thoratec Corp
Quantity
8,194 battery clips
Product code
DSQ
FDA submissions
P060040, P920014
Distribution
Worldwide - USA- AL, AR, AZ, CA, CO, CT, FL, GA, IA, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, NC, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI and the District Columbia. Internationally to Austria, Belgium, Bulgaria, Canada, Czech Republic, Denmark, France, Germany, Greece, Ireland, Israel, Italy, Lebanon, Lithuania, Luxembourg, Monaco, Netherlands, Singapore, South Africa, Sweden, Switzerland, Taiwan and the UK.

Product description

HeartMate 12 volt sealed lead acid Battery Clips that are components of the HeartMate II LVAS (PMA P060040) and the HeartMate XVE LVAS (PMA P920014) Systems. The HeartMate II and the HeartMate XVE LVAS are intended for use as a bridge to transplantation in cardiac transplant candidates at risk of imminent death from non-reversible left ventricular failure. The HeartMate XVE LVAS is also indicated for use in patients with New York Heart Association (NYHA) Class IIIB or IV end-stage left ventricular failure who have received optimal medical therapy for at least 45 of the last 60 days and who are not candidates for cardiac transplantation. The HeartMate II and XVE LVAS are intended for use both inside and outside the hospital, for transportation of Ventricular Assist Device (VAD) patients via ground ambulance, fixed-wing aircraft or helicopter. The devices consist of an implantable blood pump connected to an external System Controller by a percutaneous lead. The external controller is powered by AC electrical power, via the Power Module or Power Base Unit, or by battery power with a pair of rechargeable HeartMate batteries. Two batteries are used simultaneously for power. The LVAS is also designed to operate on a single battery, facilitating battery changes without interruption of VAD support.

Code information (models, lots, serial numbers)

Lot numbers are visible on the clips: 458560206, 581080207, 581080507, 34331720107, 34404690108, 34532990109, 343317201407, 34331720207,34331720507, 34404690208, 34404690308, 34404690408, 34432810108, 34450070108, 34477720108, 34570570109, 34577730109, 34331720307.

Other devices under the same FDA submission

7 devices cleared under P060040, P920014 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.