Lifeline AED DDU-100 semiautomatic external defibrillator… recallZ-1781-2010 · Defibtech LLC · initiated May 21, 2010
Terminated All recalls posted in June 2010
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Defibtech LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The AED when used with an affected battery pack, may falsely detect an error condition, cancel charge and not provide therapy
FDA root cause category: Device Design
What the company is doing
Defibtech, LLC issued an "Urgent Medical Device Safety Information and Correction" notification dated May 20, 2010 by email and hard copy via US Mail. The notification instructed Consignees on proper identification of affected product and steps to take for product replacement. For further information, contact your distributor or Defibtech Technical Support at [email protected] or call 1-877-453-4507 or 1-203-453-4507, 8:30 AM to 5:00 PM Monday - Friday, EST.
Recall details
- Recall number
- Z-1781-2010
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- May 21, 2010
- Posted by the FDA
- Jun 10, 2010
- Recalling firm
- Defibtech LLC
- Product code
- MKJ
- FDA submissions
- K013896
- Distribution
- Worldwide Distribution -- United States, Canada, Australia, Jordan, Spain, Israel, UK, Poland, Italy, Netherlands, Malaysia and Ireland.
Product description
Lifeline AED DDU-100 semiautomatic external defibrillator used with the DBP-2800 Battery Pack, Manufactured by Defibtech, LLC. Seymour, CT 06483. The DBP-2800 battery pack is an accessory to the AED and does not have its own intended use. The AED is intended for use on victims of sudden cardiac arrest when the patient is unconscious, unresponsive and not breathing.
Code information (models, lots, serial numbers)
Serial Numbers range from 202001005 through 202005916 and 206001001 through 206009871.
Other devices under the same FDA submission
10 devices cleared under K013896 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Lifeline Configuration DDU-100ADefibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)1 death1,285 reports
- Lifeline DDU-100ADefibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)1 recall3 deaths640 reports
- Lifeline Configuration DDU-100EDefibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)1 recall469 reports
- Lifeline DDU-100EDefibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)1 recall177 reports
- ReviveR Configuration DDU-100BDefibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)185 reports
- ReviveR DDU-100BDefibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)71 reports
- DDU-1XX Series Adult Rescue Pads DDP-100Defibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)1 recall1 report
- DDU-1XX Series Battery Pack DBP-1400Defibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)3 reports
- Lifeforce Configuration DDU-100CDefibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)2 reports
- DDU-1XX series Battery Pack Configuration DBP-1400Defibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)2 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 55497).
- Z-1782-2010 · TerminatedReviveR AED semiautomatic external defibrillator used with the DBP-2800 Battery Pack. Manufactured by Defibtech, LLC. Seymour, CT 06483. The DBP-2800 battery pack is an accesso…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page