Lifeline Configuration DDU-100ERecalls & reported problems

Defibtech, L.L.C. · Class 3

All models of Lifeline Configuration (2)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Defibtech Lifeline AED is a portable external defibrillator used in sudden cardiac arrest. Many of its 469 reports describe the unit failing to power on or its battery draining early. The reports include no injuries and no deaths, and reports in the last 12 months fell to 64 from 242 before. 1 open recall names this device, started in March 2025, over missing Italian-language labeling and instructions required for Switzerland.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
469
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 469 adverse event reports that list this exact device (DI 10815098020031), received between Mar 21, 2023 and Aug 10, 2026 — classified as 469 malfunctions.

The most frequently reported specific device problem is “Failure to Power Up” (180 reports).

1 FDA recall record name this device's identifier, 1 still open.

5 other recalls cover devices cleared under the same FDA submission (K013896); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 23 reportsJan 2024: 14 reportsFeb 2024: 9 reportsMar 2024: 11 reportsApr 2024: 13 reportsMay 2024: 14 reportsJun 2024: 7 reportsJul 2024: 18 reportsAug 2024: 8 reportsSep 2024: 18 reportsOct 2024: 15 reportsNov 2024: 24 reportsDec 2024: 7 reportsJan 2025: 18 reportsFeb 2025: 6 reportsMar 2025: 23 reportsApr 2025: 5 reportsMay 2025: 19 reportsJun 2025: 33 reportsJul 2025: 32 reportsAug 2025: 29 reportsSep 2025: 13 reportsOct 2025: 33 reportsNov 2025: 14 reportsDec 2025: 37 reportsJan 2026: 12 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202637
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 18, 2025Class IIOpen, Classified

    It was noticed there was no Italian translation included for the AED. Article 16 of the Swiss Medical Devices Ordinance (MedDO) states that product information (labelling and instructions for use) should be provided in all three official languages of Switzerland (German, French, and Italian).

    Z-1690-2025 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Failure to Power Up180
  • Premature Discharge of Battery150
  • Failure to Deliver Shock/Stimulation42
  • Battery Problem39
  • No Audible Prompt/Feedback28
  • Inappropriate/Inadequate Shock/Stimulation14
  • Connection Problem9
  • Failure to Sense4

Patient problems

  • No Clinical Signs, Symptoms or Conditions461
  • Insufficient Information6
  • Cardiac Arrest2

Recent adverse event reports

The 25 most recent of 469 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 10, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50192581 ↗
  • MalfunctionJan 19, 2026Failure to Power Up

    AN INTERNATIONAL DISTRIBUTOR REPORTED THAT THEIR CUSTOMER'S AED WOULD NOT POWER ON; HOWEVER, WHEN TESTED WITH A SPARE BATTERY PACK, IT DID POWER ON AND PROMPTED A SERVICE CODE. THE CUSTOMER DID NOT REPORT THIS OCCURRING DURING PATIENT USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24109567 ↗
  • MalfunctionJan 19, 2026Failure to Power Up

    AN INTERNATIONAL DISTRIBUTOR REPORTED THAT THEIR CUSTOMER'S AED WOULD NOT POWER ON; HOWEVER, WHEN TESTED WITH A SPARE BATTERY PACK, IT DID POWER ON AND PROMPTED A SERVICE CODE. THE CUSTOMER DID NOT REPORT THIS OCCURRING DURING PATIENT USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24109566 ↗
  • MalfunctionJan 19, 2026Battery Problem

    AN INTERNATIONAL DISTRIBUTOR REPORTED A CUSTOMER'S BATTERY PACK WILL NOT STAY LATCHED IN THEIR AED. THE CUSTOMER DID NOT REPORT THIS OCCURRING DURING PATIENT USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24109560 ↗
  • MalfunctionJan 19, 2026Battery Problem

    AN INTERNATIONAL DISTRIBUTOR REPORTED A CUSTOMER'S BATTERY PACK WILL NOT STAY LATCHED IN THEIR AED. THE CUSTOMER DID NOT REPORT THIS OCCURRING DURING PATIENT USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24109558 ↗
  • MalfunctionJan 13, 2026No Audible Prompt/Feedback

    AN INTERNATIONAL DISTRIBUTOR REPORTED A CUSTOMER'S AED WILL NOT SPEAK. THE CUSTOMER DID NOT REPORT THIS OCCURRING DURING PATIENT USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24052897 ↗
  • MalfunctionJan 13, 2026No Audible Prompt/Feedback

    AN INTERNATIONAL DISTRIBUTOR REPORTED A CUSTOMER'S AED WILL NOT SPEAK. THE CUSTOMER DID NOT REPORT THIS OCCURRING DURING PATIENT USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24052896 ↗
  • MalfunctionJan 13, 2026Premature Discharge of Battery

    AN INTERNATIONAL DISTRIBUTOR REPORTED THEIR CUSTOMER'S AED WOULD NOT POWER ON, BUT IT POWERED ON WITH A SPARE BATTERY. THE CUSTOMER DID NOT REPORT THIS OCCURRING DURING PATIENT USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24052862 ↗
  • MalfunctionJan 6, 2026Inaudible or Unclear Audible Prompt/Feedback

    AN INTERNATIONAL DISTRIBUTOR REPORTED A CUSTOMER'S AED'S AUDIO PROMPTS ARE NOT CLEAR. THE CUSTOMER DID NOT REPORT THIS OCCURRING DURING PATIENT USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23979779 ↗
  • MalfunctionJan 6, 2026Failure to Power Up

    A CUSTOMER REPORTED THAT THEIR AED WOULD NOT POWER ON; HOWEVER, WHEN TESTED WITH A SPARE BATTERY PACK, IT DID POWER ON AND PROMPTED A SERVICE CODE. THE CUSTOMER DID NOT REPORT THIS OCCURRING DURING PATIENT USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23974071 ↗
  • MalfunctionJan 6, 2026Premature Discharge of Battery

    A CUSTOMER REPORTED THAT THEIR AED'S ASI IS FLASHING RED, AND THE AED WILL NOT POWER ON. THE CUSTOMER DID NOT REPORT THIS OCCURRING DURING PATIENT USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23974052 ↗
  • MalfunctionJan 5, 2026No Audible Prompt/Feedback

    AN INTERNATIONAL DISTRIBUTOR REPORTED A CUSTOMER'S AED WILL NOT SPEAK. THE CUSTOMER DID NOT REPORT THIS OCCURRING DURING PATIENT USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23963862 ↗
  • MalfunctionJan 5, 2026Failure to Power Up

    A CUSTOMER REPORTED THAT THEIR AED'S ASI IS BLANK, AND THE AED WILL NOT POWER ON. THE CUSTOMER DID NOT REPORT THIS OCCURRING DURING PATIENT USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23959152 ↗
  • MalfunctionDec 20, 2025Failure to Power Up; Failure to Power Up

    A CUSTOMER REPORTED THAT THEIR AED'S ASI IS FLASHING RED, AND THE AED WILL NOT POWER ON. THE CUSTOMER DID NOT REPORT THIS OCCURRING DURING PATIENT USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23858854 ↗
  • MalfunctionDec 18, 2025Failure to Power Up; Failure to Power Up

    A CUSTOMER REPORTED THAT THEIR AED'S ASI IS BLANK, AND THE AED WILL NOT POWER ON. THE CUSTOMER DID NOT REPORT THIS OCCURRING DURING PATIENT USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23839866 ↗
  • MalfunctionDec 18, 2025Premature Discharge of Battery; Premature Discharge of Battery

    AN INTERNATIONAL DISTRIBUTOR REPORTED THAT THEIR CUSTOMER'S AED WOULD NOT POWER ON; HOWEVER, WHEN TESTED WITH A SPARE BATTERY PACK, IT DID POWER ON AND PROMPTED A SERVICE CODE. THE CUSTOMER DID NOT REPORT THIS OCCURRING DURING PATIENT USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23839863 ↗
  • MalfunctionDec 18, 2025No Audible Prompt/Feedback; No Audible Prompt/Feedback

    AN INTERNATIONAL DISTRIBUTOR REPORTED A CUSTOMER'S AED WILL NOT SPEAK. THE CUSTOMER DID NOT REPORT THIS OCCURRING DURING PATIENT USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23839855 ↗
  • MalfunctionDec 17, 2025Inaudible or Unclear Audible Prompt/Feedback; Inaudible or Unclear Audible Prompt/Feedback

    AN INTERNATIONAL DISTRIBUTOR REPORTED A CUSTOMER'S AED WILL NOT SPEAK. THE CUSTOMER DID NOT REPORT THIS OCCURRING DURING PATIENT USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23835459 ↗
  • MalfunctionDec 17, 2025Inaudible or Unclear Audible Prompt/Feedback; Inaudible or Unclear Audible Prompt/Feedback

    AN INTERNATIONAL DISTRIBUTOR REPORTED A CUSTOMER'S AED WILL NOT SPEAK. THE CUSTOMER DID NOT REPORT THIS OCCURRING DURING PATIENT USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23835458 ↗
  • MalfunctionDec 17, 2025Failure to Power Up; Failure to Power Up

    A CUSTOMER REPORTED THAT THEIR AED'S ASI IS FLASHING RED, AND REPORTING A SERVICE CODE. THE CUSTOMER DID NOT REPORT THIS OCCURRING DURING PATIENT USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23832255 ↗
  • MalfunctionDec 17, 2025Premature Discharge of Battery; Premature Discharge of Battery

    A CUSTOMER REPORTED THEIR AED IS FLASHING RED AND PROMPTING A "REPLACE BATTERY PACK NOW" WARNING. THEY REPORTED THIS DID NOT OCCUR DURING PATIENT USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23829587 ↗
  • MalfunctionDec 15, 2025Premature Discharge of Battery; Premature Discharge of Battery

    AN INTERNATIONAL DISTRIBUTOR REPORTED THEIR CUSTOMER'S AED BATTERY WAS DEPLETED. THE CUSTOMER DID NOT REPORT THIS OCCURRING DURING PATIENT USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23809755 ↗
  • MalfunctionDec 15, 2025Premature Discharge of Battery; Premature Discharge of Battery

    AN INTERNATIONAL DISTRIBUTOR REPORTED THEIR CUSTOMER'S AED BATTERY WAS DEPLETED. THE CUSTOMER DID NOT REPORT THIS OCCURRING DURING PATIENT USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23809750 ↗
  • MalfunctionDec 15, 2025Premature Discharge of Battery; Premature Discharge of Battery

    AN INTERNATIONAL DISTRIBUTOR REPORTED THEIR CUSTOMER'S AED WOULD NOT POWER ON, BUT IT POWERED ON WITH A SPARE BATTERY. THE CUSTOMER DID NOT REPORT THIS OCCURRING DURING PATIENT USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23809742 ↗
  • MalfunctionDec 15, 2025Premature Discharge of Battery; Premature Discharge of Battery

    AN INTERNATIONAL DISTRIBUTOR REPORTED THEIR CUSTOMER'S AED WOULD NOT POWER ON, BUT IT POWERED ON WITH A SPARE BATTERY. THE CUSTOMER DID NOT REPORT THIS OCCURRING DURING PATIENT USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23809736 ↗

FDA clearance and classification

Cleared through 510(k) K013896 on Jun 19, 2002 (decision: Substantially Equivalent - Subject to Tracking Reg.). Submitted by Defibtech, LLC as “Sentry Semiautomatic External Defibrillator; External Defibrillation Pads (Electrodes) Ddp-100; Battery Pack 1200mah; Ba”. The FDA review took 208 days. FDA record ↗

The FDA regulates product code MKJ (automated external defibrillators (non-wearable)) under 21 CFR 870.5310 as a Class III device. 21 CFR 870.5310 ↗

FDA definition of this device type: This device is a non-wearable prescription use only aed. These are devices that include automated external defibrillation. Automated external defibrillators use external pad-type electrodes to sense, detect, classify and treat (with an electrical shock) ventricular fibrillation. These devices are intended to be used on suspected victims of sudden cardiac arrest. A person in cardiac arrest is unresponsive and is not breathing normally. The device can be sold with prescription only.

Recalls under the same FDA submission

These recalls involve devices cleared under K013896. One submission often covers many models, so check the product and lot details in each record.

  • Z-2064-2011 · Apr 1, 2011 · Terminated
    Defibtech Lifeline AED and ReviveR AED, Model: DDU-100 Series AEDs Software version 2.004 or earlier The product under correction is the DDU-100 AED. The AED is indicated for
  • Z-0077-2011 · Aug 27, 2010 · Terminated
    DDC-6-AA Data Card, an accessory to the Sentry DDU-1 00 semiautomatic external defibrillator (AED). The DDC-6-AA data card is an accessory to the AED and does not have its own inte
  • Z-1781-2010 · May 21, 2010 · Terminated
    Lifeline AED DDU-100 semiautomatic external defibrillator used with the DBP-2800 Battery Pack, Manufactured by Defibtech, LLC. Seymour, CT 06483. The DBP-2800 battery pack is an
  • Z-1782-2010 · May 21, 2010 · Terminated
    ReviveR AED semiautomatic external defibrillator used with the DBP-2800 Battery Pack. Manufactured by Defibtech, LLC. Seymour, CT 06483. The DBP-2800 battery pack is an accesso
  • Z-0580-2007 · Feb 17, 2007 · Terminated
    Defibtech Sentry Semi-Automatic External Defibrilator (AED) sold as 'Lifeline AED' brand -Model DDU-100A (version 2.002 and earlier)

Common questions about the Lifeline Configuration DDU-100E

Has the Lifeline Configuration DDU-100E been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1690-2025 (Class II), began on Mar 18, 2025: It was noticed there was no Italian translation included for the AED. Article 16 of the Swiss Medical Devices Ordinance (MedDO) states that product information (labelling and instructions for use) should be provided… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Lifeline Configuration DDU-100E?

The FDA's MAUDE database lists 469 adverse event reports for this device, received between Mar 21, 2023 and Aug 10, 2026. In those reports, the most common specific device problems are “Failure to Power Up” (180), “Premature Discharge of Battery” (150) and “Failure to Deliver Shock/Stimulation” (42); the most common patient problem is “Cardiac Arrest” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Lifeline Configuration DDU-100E?

None of the reports are classified as deaths, 0 as injuries and 469 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Lifeline Configuration DDU-100E FDA approved?

It was cleared through 510(k) K013896 on Jun 19, 2002. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the Lifeline Configuration DDU-100E, and is it still sold?

The FDA's device database lists Defibtech, L.L.C. as the company responsible for it (brand Lifeline Configuration, model DDU-100E). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.