Lifeline DDU-100ERecalls & reported problems
Defibtech, L.L.C. · Class 3
All models of Lifeline (2)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This DefibTech Lifeline defibrillator has 174 adverse event reports, with no deaths and 1 injury. The reports mostly describe the unit failing to power on, batteries discharging early, and service codes appearing, and many say the problem happened outside patient use. One recall is open, started in March 2025, over labels and instructions that lack an Italian translation. Reports rose to 172 in the latest 12 months from 1 in the prior 12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 177
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 177 adverse event reports that list this exact device (DI 00815098020034), received between Sep 29, 2022 and Sep 30, 2026 — classified as 1 injury, 176 malfunctions.
The most frequently reported specific device problem is “Failure to Power Up” (73 reports).
1 FDA recall record name this device's identifier, 1 still open.
5 other recalls cover devices cleared under the same FDA submission (K013896); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Mar 18, 2025Open, Classified
It was noticed there was no Italian translation included for the AED. Article 16 of the Swiss Medical Devices Ordinance (MedDO) states that product information (labelling and instructions for use) should be provided in all three official languages of Switzerland (German, French, and Italian).
Z-1690-2025 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Failure to Power Up73
- Premature Discharge of Battery65
- Battery Problem18
- No Audible Prompt/Feedback10
- Failure to Deliver Shock/Stimulation8
- Audible Prompt/Feedback Problem2
- Insufficient Device Problem Information1
Patient problems
- No Clinical Signs, Symptoms or Conditions176
- Cardiac Arrest1
Recent adverse event reports
The 25 most recent of 177 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Failure to Power Up
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50953764 ↗MalfunctionSep 29, 2026Premature Discharge of Battery
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50927997 ↗MalfunctionSep 29, 2026Failure to Power Up
An international distributor reported that their customer's AED would not power on; however, when tested with a spare battery pack, it did power on and prompted a service code. The customer did not report this occurring during patient use.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50927982 ↗MalfunctionSep 29, 2026Failure to Power Up
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50927975 ↗MalfunctionSep 23, 2026Failure to Power Up
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50873432 ↗MalfunctionSep 22, 2026Premature Discharge of Battery
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50859373 ↗MalfunctionSep 22, 2026Premature Discharge of Battery
An international distributor reported their customer's AED battery was depleted. The customer did not report this occurring during patient use.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50853440 ↗MalfunctionSep 22, 2026Failure to Power Up
An international distributor reported their customer's AED would not power on. The customer did not report this occurring during patient use.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50853438 ↗MalfunctionSep 14, 2026Failure to Power Up
A customer reported that their AED's ASI is flashing red, and the AED will not power on. The customer did not report this occurring during patient use.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50755247 ↗MalfunctionSep 14, 2026Premature Discharge of Battery
An international distributor reported their customer's AED is flashing red and prompting a "replace battery pack now warning." The customer did not report this occurring during patient use.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50753888 ↗MalfunctionSep 14, 2026Failure to Power Up
An international distributor reported that their customer's AED would not power on; however, when tested with a spare battery pack, it powered on. The customer did not report this occurring during patient use.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50753881 ↗MalfunctionSep 11, 2026Insufficient Device Problem Information
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50741147 ↗MalfunctionSep 10, 2026No Audible Prompt/Feedback
An international distributor reported a customer's AED will not speak. The customer did not report this occurring during patient use.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50714653 ↗MalfunctionSep 10, 2026Premature Discharge of Battery
An international distributor reported their customer's AED is flashing red and prompting a "replace battery pack now warning". The customer did not report this occurring during patient use.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50714650 ↗MalfunctionSep 8, 2026Premature Discharge of Battery
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50676678 ↗MalfunctionAug 31, 2026Failure to Power Up
An international distributor reported their customer's AED ASI is flashing red, and the device is reporting a service code. The customer did not report this occurring during patient use.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50524498 ↗MalfunctionAug 31, 2026Failure to Power Up
An international distributor reported that their customer's AED would not power on; however, when tested with a spare battery pack, the AED powered on. The customer did not report this occurring during patient use.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50524489 ↗MalfunctionAug 31, 2026Failure to Power Up
An international distributor reported that their customer's AED would not power on; however, when tested with a spare battery pack, it did power on and prompted a service code. The customer did not report this occurring during patient use.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50524483 ↗MalfunctionAug 28, 2026Premature Discharge of Battery
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50498136 ↗MalfunctionAug 28, 2026Premature Discharge of Battery
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50498132 ↗MalfunctionAug 28, 2026No Audible Prompt/Feedback
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50498129 ↗MalfunctionAug 28, 2026No Audible Prompt/Feedback
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50498123 ↗MalfunctionAug 26, 2026Failure to Power Up
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50463612 ↗MalfunctionAug 24, 2026Failure to Power Up
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50419050 ↗MalfunctionAug 13, 2026Failure to Power Up
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50230434 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K013896. One submission often covers many models, so check the product and lot details in each record.
- Z-2064-2011 · Apr 1, 2011 · TerminatedDefibtech Lifeline AED and ReviveR AED, Model: DDU-100 Series AEDs Software version 2.004 or earlier The product under correction is the DDU-100 AED. The AED is indicated for
- Z-0077-2011 · Aug 27, 2010 · TerminatedDDC-6-AA Data Card, an accessory to the Sentry DDU-1 00 semiautomatic external defibrillator (AED). The DDC-6-AA data card is an accessory to the AED and does not have its own inte
- Z-1781-2010 · May 21, 2010 · TerminatedLifeline AED DDU-100 semiautomatic external defibrillator used with the DBP-2800 Battery Pack, Manufactured by Defibtech, LLC. Seymour, CT 06483. The DBP-2800 battery pack is an
- Z-1782-2010 · May 21, 2010 · TerminatedReviveR AED semiautomatic external defibrillator used with the DBP-2800 Battery Pack. Manufactured by Defibtech, LLC. Seymour, CT 06483. The DBP-2800 battery pack is an accesso
- Z-0580-2007 · Feb 17, 2007 · TerminatedDefibtech Sentry Semi-Automatic External Defibrilator (AED) sold as 'Lifeline AED' brand -Model DDU-100A (version 2.002 and earlier)
Other Automated External Defibrillators (Non-Wearable) devices
- NA 861290Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)12 deaths5,167 reports
- NA 861288Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)21 deaths5,015 reports
- NA 861289Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)7 deaths3,385 reports
- R SERIES 30001005201310012Zoll Medical Corporation · Automated external defibrillators (non-wearable)10 deaths1,720 reports
- PHILIPS HeartStart FRxPhilips Electronics North America Corporation · Automated external defibrillators (non-wearable)26 deaths1,586 reports
- LIFEPAK CR2 99512Physio-Control, Inc. · Automated external defibrillators (non-wearable)8 deaths1,502 reports
- Lifeline Configuration DDU-100ADefibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)1 death1,285 reports
- PHILIPS HeartStart FR3Philips Electronics North America Corporation · Automated External Defibrillators (Non-Wearable)10 deaths1,188 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.