Mitek VAPR Side Effect Electrode (integrated) recallZ-2301-2010 · DePuy Mitek, Inc., a Johnson & Johnson Co. · initiated Jun 7, 2010
Terminated All recalls posted in August 2010
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from DePuy Mitek, Inc., a Johnson & Johnson Co. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Product does not meet the required packaging specifications for pouch seal integrity which could lead to a breach in sterility
FDA root cause category: Packaging process control
What the company is doing
The firm, Depuy Mitek, sent an "Urgent Voluntary Product Recall" letter dated June 7, 2010, to all customers. The letter described the product, problem and action to be taken by the customers. The customers were instructed to not use any of the identified recalled products; to isolate all inventory of the affected product; complete the Business Reply Form, and return the affected products along with the form to Depuy Mitek; ATTN: Recall Coordinator, 50 Scotland Boulevard, Bridgewater, MA 02324 or fax completed form to 1-508-828-3750. Note the IMPORTANT PRODUCT INFORMATION: Any product that contains a round green sticker has been 100% inspected, meets specifications and is acceptable for use. Please Do Not Return any product that is labeled with the green dots because this product is not affected by the recall. If you have any questions or concerns with regard to this program, please contact your local Depuy Mitek Representative or DePuy Mitek Customer Service directly at 1-800-382-4682.
Recall details
- Recall number
- Z-2301-2010
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Jun 7, 2010
- Posted by the FDA
- Aug 27, 2010
- Recalling firm
- DePuy Mitek, Inc., a Johnson & Johnson Co.
- Quantity
- 112936 units
- Product code
- GEI
- FDA submissions
- K082643
- Distribution
- Worldwide distribution: USA and countries including Canada, Argentina, Australia, Bahrain, Belgium, Brazil, Chile, China, Columbia, Costa Rica, Czech Republic, Ecuador, Egypt, France, Germany, Greece, Guatemala, Hong Kong, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Jordan, Kazakhastan, Kuwait, Lebanon, Maylasia, Mexico, Netherlands, New Zealand, Oman, Pakistan, Palestine, Panama, Peru, Poland, Qatar, Russia, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Portugal, Sweden, Switzerland, Syria, Thailand ,Turkey, UK, Ukraine, UAE, Venezuela, Vietnam, and Yemen.
Product description
Mitek VAPR Side Effect Electrode (integrated) Product Code: 227301 The VAPR Electrodes (end effect, side effect, wedge, angled, hook, flex, side effect short and wedge short) and VAPR Suction Electrodes (S9 and S ) with Integrated Handpiece with Integrated Handpiece, when used with the DePuy Mitek VAPR Electrosurgical System, are intended for resection, ablation, excision of soft tissue, hemostasis of blood vessels and coagulation of soft tissues in patients requiring arthroscopic surgery of the knee, shoulder, ankle, elbow and wrist.
Code information (models, lots, serial numbers)
All lots of unexpired product
Other devices under the same FDA submission
12 devices cleared under K082643 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- VAPR 225301Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES31 reports
- VAPR 227305Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES27 reports
- VAPR 227312Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES10 reports
- VAPR 227301Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES9 reports
- VAPR 227203Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES7 reports
- VAPR 227211Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES6 reports
- VAPR 225305Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES4 reports
- VAPR 227213Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES4 reports
- VAPR 225303Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES3 reports
- VAPR 227252Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES3 reports
- VAPR 227302Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES2 reports
- VAPR 227202Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page