VAPR 227301Recalls & reported problems
Depuy Mitek, LLC · VAPR Side Effect Electrode 3.5mm, Side Effect with Integrated Handpiece · Class 2
All models of VAPR (30)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10886705009671), received between Dec 10, 2018 and Oct 26, 2022 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Device Damaged Prior to Use” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (K082643); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Device Damaged Prior to Use2
- Electrical /Electronic Property Problem2
- Accessory Incompatible1
- Blocked Connection1
- Disconnection1
- Excessive Heating1
- Output Problem1
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- No Consequences Or Impact To Patient4
- Not Applicable2
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 26, 2022Electrical /Electronic Property Problem; Use of Device Problem; Electrical /Electronic Property Problem; Use of Device Problem
THIS IS REPORT 1 OF 2 FOR (B)(4). IT WAS REPORTED BY THE DISTRIBUTOR IN SLOVAKIA THAT DURING A KNEE ARTHROSCOPY PROCEDURE ON (B)(6) 2022, IT WAS OBSERVED THAT THE VAPR SIDE EFFECT ELECTRODE 3.5MM, SIDE EFFECT W/INTEGRATED HANDPIECE DEVICE DID NOT WORK. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE DISTAL TIP ON THE DEVICE SHOWED NO OBVIOUS SIGNS OF ACTIVATION, ELECTRICAL TESTING REVEALED THAT THE CAPACITANCE VALUE WAS SLIGHTLY ABOVE TOP SPECIFICATION VALUE, AND WHEN CONNECTED TO A GENERATOR THE DISPLAY VARIED BETWEEN CORRECT SETTINGS, INVALID INSTRUMENT AND INCORRECT DEFAULT SETTINGS. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15674842 ↗MalfunctionJan 12, 2021Output Problem; Output Problem
IT WAS REPORTED BY THE AFFILIATE IN (B)(6) THAT PREOPERATIVELY TO AN ARTHROSCOPY PROCEDURE ON (B)(6) 2020, IT WAS OBSERVED THAT AN ELECTRODE DEVICE PROMPTED AN UNSPECIFIED ALERT MESSAGE ON THE GENERATOR. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THERE WERE TISSUE DEBRIS BEING VISIBLE AROUND THE TIP ON THE ELECTRODE DEVICE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH A THREE MINUTE DELAY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11154320 ↗MalfunctionDec 3, 2020Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
IT WAS REPORTED BY THE SALES REP THAT DURING A ROTATOR CUFF REPAIR ON (B)(6) 2020, IT WAS OBSERVED THAT THE VAPR SIDE EFFECT ELECTRODE 3.5MM, SIDE EFFECT W/INTEGRATED HANDPIECE SHORTED. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. NO PATIENT CONSEQUENCES OR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10941548 ↗MalfunctionApr 29, 2020Device Damaged Prior to Use; Device Damaged Prior to Use
IT WAS REPORTED BY AFFILIATE VIA EMAIL THE VAPR3.5MM SIDEEFFECT 1-PCE ELECTRODE-EA, IMPLANT IS REQUESTED AND IT IS OPENED PROPERLY, THE ENVELOPE WHERE THE VAPR SIDE EFFECT ELECTRODE COMES FROM IS CAREFULLY OPENED SO THAT THE SURGICAL NURSE COULD TAKE THE AFOREMENTIONED MATERIAL, REFERENCE (B)(4), AFTER WHICH IT IS CONNECTED TO THE VAPR CONSOLE AND WE BECAME AWARE OF BOTH THE JOHNSON TECHNICIAN (B)(6) AND I, WHO REPORTED AN ERROR IN THE CONSOLE WHEN CONNECTING THE HANDPIECE, WE PROCEEDED TO CHANGE THE TRADITIONAL CONSOLE TO A VAPR VUE CONSOLE TO VERIFY IF THE PROBLEM WAS THE CONSOLE, BUT IT WAS NOT; SINCE THE ELECTRODE FAILED AGAIN, SO WE PROCEEDED TO OPEN ANOTHER VAPR TIP 227301 BY CONNECTING IT TO THE VAPR VUE CONSOLE WHERE IT WORKED WITHOUT ANY PROBLEM. NO PATIENT CONSEQUENCES. THERE WAS DELAY IN THE SURGERY OF 10 MINUTES. THE DEVICE IS AVAILABLE TO BE RETURNED FOR EVALUATION.
Patient problems: Not Applicable
MAUDE report 10012076 ↗MalfunctionApr 29, 2020Device Damaged Prior to Use; Device Damaged Prior to Use
IT WAS REPORTED BY AFFILIATE VIA EMAIL THE VAPR 3.5 MM SIDE EFFECT 1-PCE ELECTRODE-EA, OPENED PROPERLY, IT CONNECTED NORMALLY TO THE VAPR VUE CONSOLE, WITH A WIRELESS PEDAL, WHEN THE DOCTOR USED IT, DID NOT WORK, IT MARKED AN ERROR, SO IT WAS PRECEDED TO CHANGE THE CONSOLE TO VAP3, THINKING THAT IT WAS THE CONSOLE THAT HAD THE PROBLEM, BUT IN THE NEW CONSOLE IT ALSO MARKED ERROR. THE DR. REQUESTED THAT THE PIECE NOT BE CHARGED, FOR THIS REASON THE HOSPITAL ASKED US FOR THE REPLACEMENT WITHOUT CHARGE, AS WELL AS PUNCTUAL FOLLOW-UP TO THE DIAGNOSIS AND RETURN OF THE DEFECTIVE MATERIAL. THE SURGERY CONTINUED WITHOUT PROBLEMS. THE DEVICE IS AVAILABLE TO BE RETURNED FOR EVALUATION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10011746 ↗MalfunctionFeb 26, 2020Blocked Connection; Blocked Connection
IT WAS REPORTED BY THE AFFILIATE VIA EMAIL THAT DURING AN UNKNOWN PROCEDURE WITHOUT SWITCHING ON THE VAPR3.5MM SIDEEFFECT 1-PCE ELECTRODE-EA THE PORTAL GAVE THE ERROR MESSAGE "ELECTRODE NOT RECOGNIZED". THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. NO PATIENT CONSEQUENCE AND NO SURGICAL DELAY WAS REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 9759442 ↗MalfunctionJan 27, 2020Accessory Incompatible; Accessory Incompatible
IT WAS REPORTED BY SALES CONSULTANT VIA COMPLAINT SUBMISSION TOOL, THAT PRE-OPERATIVELY TO A SHOULDER REPAIR A VAPR SIDE EFFECT ELECTRODE 3.5MM, SIDE EFFECT W/INTEGRATED HANDPIECE, PRESENTED A MALFUNCTION. NO SURGICAL DELAY OR PATIENT CONSEQUENCE REPORTED. ADDITIONAL INFORMATION REPORTED A TWO MINUTE SURGICAL DELAY AND THE CASE WAS COMPLETED WITH ANOTHER READILY AVAILABLE DEVICE. THE AFFILIATE REPORTED SURGICAL INTERVENTION WAS NOT REQUIRED AND THE FAILURE OCCURRED WHEN THE MACHINE DID NOT RECOGNIZE THE ELECTRODE AFTER CONNECTION. IT WAS REPORTED THE DEVICE WILL NOT BE RETURNED FOR EVALUATION BECAUSE IT HAS BEEN DISCARDED.
Patient problems: Not Applicable
MAUDE report 9634801 ↗MalfunctionDec 26, 2019Excessive Heating; Excessive Heating
IT WAS REPORTED BY THE AFFILIATE VIA COMPLAINT SUBMISSION TOOL THAT DURING AN UNKNOWN PROCEDURE THE VAPR 3.5MM SIDEEFFECT 1-PCE ELECTRODE-EA TURNED RED COLOR WHEN IT WAS BEEN USED, LIKE IT WAS OVERHEATING. THERE WAS NO PATIENT CONSEQUENCE AND NO SURGICAL DELAY WAS REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9520212 ↗MalfunctionDec 10, 2018Disconnection; Disconnection
IT WAS REPORTED BY THE SALES REP THAT DURING ROTATOR CUFF REPAIR SURGICAL PROCEDURE, IT WAS OBSERVED THAT WHEN THE CUSTOMER'S VAPR 3.5MM SIDE EFFECT ELECTRODE WAS PLUGGED INTO THE GENERATOR, THE DEVICE WAS RECOGNIZED BUT WOULD NOT ABLATE OR COAGULATE. THE PROCEDURE WAS COMPLETED WITH ANOTHER LIKE DEVICE USING THE SAME GENERATOR WITH NO HARM TO THE PATIENT BUT THERE WAS A ONE MINUTE DELAY TO GRAB ANOTHER ELECTRODE. THE ELECTRODE IS NOT A REPROCESSED DEVICE. THE SALES REP WAS NOT PRESENT FOR THE CASE THEREFORE COULD NOT PROVIDE ANY FURTHER INFORMATION. THE DEVICE WILL BE RETURNING FOR EVALUATION. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8146333 ↗
FDA clearance and classification
Cleared through Special 510(k) K082643 on Dec 19, 2008 (decision: Substantially Equivalent). Submitted by Depuy Mitek, A Johnson & Johnson Company as “3.5 Mm Side Effect Electrode With Integrated Handpiece, Model 227301, 3.5 Mm Angled Side Effect Electrode With Integrate”. The FDA review took 99 days. FDA record ↗
The FDA regulates product code GEI (electrosurgical, cutting & coagulation & accessories) under 21 CFR 878.4400 as a Class II device. 21 CFR 878.4400 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K082643. One submission often covers many models, so check the product and lot details in each record.
- Z-0133-2013 · Mar 12, 2012 · TerminatedDePuy Mitek VAPR Hook Electrode 3.5mm Hook Electrode with Integrated Handpiece Catalog Number: 227305 The DePuy Mitek VAPR Electrodes for use with the VAPR Radiofrequency Sy
- Z-2301-2010 · Jun 7, 2010 · TerminatedMitek VAPR Side Effect Electrode (integrated) Product Code: 227301 The VAPR Electrodes (end effect, side effect, wedge, angled, hook, flex, side effect short and wedge short) and V
Common questions about the VAPR 227301
Has the VAPR 227301 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10886705009671), based on the FDA recall file dated Nov 27, 2026. However, 2 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the VAPR 227301?
The FDA's MAUDE database lists 9 adverse event reports for this device, received between Dec 10, 2018 and Oct 26, 2022. In those reports, the most common specific device problems are “Device Damaged Prior to Use” (2), “Electrical /Electronic Property Problem” (2) and “Accessory Incompatible” (1); the most common patient problem is “Not Applicable” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the VAPR 227301?
None of the reports are classified as deaths, 0 as injuries and 9 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the VAPR 227301 FDA approved?
It was cleared through Special 510(k) K082643 on Dec 19, 2008. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the VAPR 227301, and is it still sold?
The FDA's device database lists Depuy Mitek, LLC as the company responsible for it (brand VAPR, model 227301). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,137 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,038 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,891 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,046 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall957 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories929 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES874 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories831 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page