VAPR 225303Recalls & reported problems
Depuy Mitek, LLC · VAPR End Effect Electrode 3.5 mm x 160 mm · Class 2
All models of VAPR (30)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10886705009275), received between Jul 13, 2021 and Mar 21, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Output Problem” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (K082643); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Output Problem2
- Use of Device Problem2
- Break1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 21, 2025Break; Use of Device Problem; Break; Use of Device Problem
IT WAS REPORTED BY THE SALES REP THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, THE VAPR 3.5MM END STR DEVICE SOUNDED AN OPERATING TONE BUT DID NOT BURN AT ALL. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE DEVICE RETURN WIRE CAME APART. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21666895 ↗MalfunctionOct 1, 2024Use of Device Problem; Output Problem; Use of Device Problem; Output Problem
IT WAS REPORTED BY THE SALES REP FROM JAPAN THAT DURING A SHOULDER JOINT SURGICAL PROCEDURE ON (B)(6) 2024 THE VAPR 3.5MM END STR -EA DEVICE SUTURE OF THE TENDON PLATE RUPTURED UNDER MICROSCOPE. THE DEVICE STOPPED BURNING AT SOME POINT. THE HANDPIECE WAS REPLACED BUT DID NOT BURN. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS A THIRTY MINUTE DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20347284 ↗MalfunctionJul 13, 2021Output Problem; Output Problem
IT WAS REPORTED BY THE AFFILIATE IN (B)(6) THAT DURING AN ARTHROSCOPIC ROTATOR CUFF REPAIR PROCEDURE ON (B)(6) 2021, IT WAS OBSERVED THAT THE VAPR 3.5MM END STR ELECTRODE DEVICE SUDDENLY BECAME OUT OF ORDER WHILE IN USE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITHOUT DELAY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED. THE DEVICE WAS BRAND NEW AND ITS FIRST USE WHEN THE ISSUE OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12160568 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K082643. One submission often covers many models, so check the product and lot details in each record.
- Z-0133-2013 · Mar 12, 2012 · TerminatedDePuy Mitek VAPR Hook Electrode 3.5mm Hook Electrode with Integrated Handpiece Catalog Number: 227305 The DePuy Mitek VAPR Electrodes for use with the VAPR Radiofrequency Syste
- Z-2301-2010 · Jun 7, 2010 · TerminatedMitek VAPR Side Effect Electrode (integrated) Product Code: 227301 The VAPR Electrodes (end effect, side effect, wedge, angled, hook, flex, side effect short and wedge short) and V
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page