Suros ATEC 0912-20 Breast Biopsy and Excision System… recallZ-2285-2010 · Hologic, Inc. · initiated Aug 11, 2010
Terminated All recalls posted in August 2010
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Hologic, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The firm received a customer complaint that the a device labeled as ATEC 0912-12 was actually ATEC 0912-20.
FDA root cause category: Packaging
What the company is doing
The firm, Hologic, Inc., sent out "Urgent Medical Device Recall" letters by FedEx on August 11, 2010, to all customers. The letter describes the product, problem and action to be taken by the customers. The customers were instructed to check their inventories and to quarantine any of the recalled product; if the product was distributed further, identify the users and notify them at once of this product recall; segregate all products from this lot and quarantine them in preparation for return to Hologic, and complete and return the enclosed RECALL RESPONSE FORM by fax at 877-574-3255 or mail. Note: Replacement product cannot be sent until Hologic receives your response form. If you have any questions, please contact Hologic directly at 317-344-7630.
Recall details
- Recall number
- Z-2285-2010
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Aug 11, 2010
- Posted by the FDA
- Aug 26, 2010
- Recalling firm
- Hologic, Inc.
- Quantity
- 990
- Product code
- KNW
- FDA submissions
- K010400, K042290
- Distribution
- Worldwide distribution: USA and countries including Canada, Belgium, Algeria, Germany, London, Lebanon and Italy.
Product description
Suros ATEC 0912-20 Breast Biopsy and Excision System Handpiece, REF ATEC0912-12, sterile, Hologic, Inc. 6100 Technology Center Drive Indianapolis, IN. The handpiece is inserted into the breast to the site of interest. When the proper location/position is acheived, the handpiece cuts and removes biopsy tissue.
Code information (models, lots, serial numbers)
00316.
Other devices under the same FDA submission
15 devices cleared under K010400, K042290 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- ATEC 0912-20Hologic, Inc. · INSTRUMENT, BIOPSY32 reports
- ATEC SAPPHIRE 100Hologic, Inc. · INSTRUMENT, BIOPSY25 reports
- ATEC SAPPHIREHologic, Inc. · INSTRUMENT, BIOPSY18 reports
- ATEC 0914-20MRHologic, Inc. · INSTRUMENT, BIOPSY17 reports
- ATEC Introducer ILS 0914-20Hologic, Inc. · INSTRUMENT, BIOPSY13 reports
- ATEC PEARLHologic, Inc. · INSTRUMENT, BIOPSY8 reports
- ATEC 1212-20Hologic, Inc. · INSTRUMENT, BIOPSY7 reports
- ATEC SAPPHIRE25Hologic, Inc. · INSTRUMENT, BIOPSY5 reports
- ATEC Introducer ILS 0914-12Hologic, Inc. · INSTRUMENT, BIOPSY4 reports
- ATEC Sapphire 200Hologic, Inc. · INSTRUMENT, BIOPSY2 reports
- ATEC 0909-20Hologic, Inc. · INSTRUMENT, BIOPSY2 reports
- ATEC 0912-12Hologic, Inc. · INSTRUMENT, BIOPSY2 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page