NexGen Complete Knee Solution Stemmed Tibial recallZ-0120-2011 · Zimmer Inc. · initiated Jan 26, 2010
Terminated All recalls posted in October 2010
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Zimmer Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Zimmer had determined that two lots of these tibial implants have incorrectly positioned or missing flange plugs that were not seated in the device upon receipt to the customer. A missing flange plug could lead to a delay of surgery or possibility of wear leading to more surgery.
FDA root cause category: Employee error
What the company is doing
The firm sent an URGENT MARKET WITHDRAWAL to its sales force and distributors on 1/26/2010. The letter identifies the affected product and explains the reason for the Market Withdrawal. The letter also explains the clinical implications and instructs consignees to locate any unused devices and quarantine them immediately. A physical count should be carried out of all affected devices and recorded on the Inventory Return Certification Form that's included with the letter. A copy of the letter should be faxed to Zimmer, Inc. and the affected product should be returned along with the form to the firm.
Recall details
- Recall number
- Z-0120-2011
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Jan 26, 2010
- Posted by the FDA
- Oct 21, 2010
- Recalling firm
- Zimmer Inc.
- Quantity
- 61 units
- Product code
- JWH
- FDA submissions
- K933785
- Distribution
- Worldwide Distribution -- US, Canada, Singapore Australia, and India.
Product description
NexGen Complete Knee Solution Stemmed Tibial Component Precoat Size 3, REF 00-5980-037-01, Sterile, Zimmer, Warsaw, IN. Intended for mating with a UHMWPE articular surface and Zimaloy Cobalt-Chromium-Molybdenum alloy femoral component and is indicated for cemented use only.
Code information (models, lots, serial numbers)
Lot Number: 61313261.
Other devices under the same FDA submission
89 devices cleared under K933785 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- NexGen® 5980-47-01Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall13 reports
- NexGen® 5988-20-14Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall3 reports
- NexGen® 5988-20-17Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall0 reports
- NexGen® 5986-37-01Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER32 reports
- NexGen® 5986-37-02Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER32 reports
- NexGen® 5980-37-02Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER23 reports
- NexGen® 5986-47-01Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER21 reports
- NexGen® 5972-65-32Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER20 reports
- NexGen® 5980-37-01Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER18 reports
- NexGen® 5980-47-02Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER17 reports
- NexGen® 5988-12-15Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER17 reports
- NexGen® 5986-47-02Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER16 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 56701).
- Z-0119-2011 · TerminatedNexGen Complete Knee Solution Stemmed Tibial Component Precoat Size 6, REF 00-5980-047-02, Sterile, Zimmer, Warsaw, IN. Intended for mating with a UHMWPE articular surface and Zi…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page