GE LightSpeed VCT Scanner System, models 5212920-300… recallZ-0368-2011 · GE Healthcare, LLC · initiated Sep 3, 2010

Terminated All recalls posted in November 2010

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from GE Healthcare, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

GE has become aware of a potential set of circumstances that could cause x-ray continuation during an unexpected table stop on certain LightSpeed VCT and BrightSpeed systems.

FDA root cause category: Software design

What the company is doing

An Urgent Medical Device Correction letter dated September 2, 2010 was sent to consignees beginning September 3, 2010. This letter addressed the LightSpeed VCT systems. Another Urgent Medical Device Correction letter, dated September 24, 2010 was sent September 24, 2010 for the BrightSpeed Systems. The letters described the issue and provided safety instructions for the health care provider to be aware if the table should stop unexpectedly and the x-ray remains on to manually stop the scan until the devices software is upgraded. If you have any questions, please contact the US (800) 437-1171, Japan 0120-055-919, Korea 1544-6119, Australia/New Zealand 800 659 465, China 800-810-8188; for other countries, please contact your local GE Healthcare Services Representative.

Recall details

Recall number
Z-0368-2011
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Sep 3, 2010
Posted by the FDA
Nov 15, 2010
Recalling firm
GE Healthcare, LLC
Quantity
68
Product code
JAK
FDA submissions
K052855, K082761, K082816
Distribution
Worldwide distribution: USA including states of: MI, NM, OH, TX, and Puerto Rico and countries including: Algeria, Argentina, Australia, Austria, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Guatemala, Honduras, Hungary, India, Indonesia, Ireland, Islamic Republic of Iran, Israel, Italy, Jamaica, Japan, Kenya, Lebanon, Malaysia, Mauritania, Mexico, Morocco, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Pakistan, Paraguay, Philippines, Poland, Portugal, Republic of Korea, Romania, Russian Federation, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Togo, Tunisia, Turkey, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Viet Nam.

Product description

GE LightSpeed VCT Scanner System, models 5212920-300, 5212920-310, 5311595-10. Composed of a gantry, patient table, operator console, computer, and PDU and includes image acquisition hardware, image acquisition and reconstruction software, associated accessories and connections/interfaces to accessories. GE Healthcare, 3000 North Grandview, Waukesha, WI 53188 USA. Is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes, including Axial, Cine, Helical (Volumetric), Cardiac, and Gated acquisitions.

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

Model 5212920-300, serial numbers: 00000021775YC2. 00000408285CN5. 00000409136CN9. 00000412029CN1. 00000412391CN5. 00000414803CN7. 00000415599CN0. 00000416813CN4. 00000417885CN1.   Model 5212920-310, serial numbers: 00000416740CN9. 00000417311CN8. 00000417519CN6. 00000417884CN4. 00000417886CN9. 00000417916CN4. 00000417917CN2. 00000417918CN0. 00000417919CN8. 00000417925CN5. 00000417926CN3. 00000417989CN1. 00000417990CN9. 00000418000CN6. 00000418055CN0. 00000418056CN8. 00000418057CN6. 00000418058CN4. 00000418059CN2. 00000418060CN0. 00000418062CN6. 00000418063CN4. 00000418064CN2. 00000418065CN9. 00000418126CN9. 00000418141CN8. 00000418142CN6. 00000418179CN8. 00000418218CN4. 00000418219CN2. 00000418220CN0. 00000418221CN8. 00000418222CN6. 00000418246CN5. 00000418260CN6. 00000418261CN4. 00000418266CN3. 00000418272CN1. 00000418293CN7. 00000418296CN0. 00000418337CN2. 00000418355CN4. 00000418365CN3. 00000418400CN8. 00000418404CN0. 00000418467CN7. 00000418541CN9. 00000418542CN7. 00000418581CN5. 00000418586CN4. 00000418604CN5. 00000418706CN8. 00000418717CN5. 00000418789CN4. 00000418790CN2. 00000418793CN6.   Model 5311595-10, serial numbers: 00000031546CO5. 00000031835CO2. 00000031838CO6.

Other devices under the same FDA submission

3 devices cleared under K052855, K082761, K082816 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 56814).