GE BrightSpeed Excel/Edge/Elite Select / Elite CT Scanner… recallZ-0369-2011 · GE Healthcare, LLC · initiated Sep 3, 2010
Terminated All recalls posted in November 2010
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from GE Healthcare, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
GE has become aware of a potential set of circumstances that could cause x-ray continuation during an unexpected table stop on certain LightSpeed VCT and BrightSpeed systems.
FDA root cause category: Software design
What the company is doing
An Urgent Medical Device Correction letter dated September 2, 2010 was sent to consignees beginning September 3, 2010. This letter addressed the LightSpeed VCT systems. Another Urgent Medical Device Correction letter, dated September 24, 2010 was sent September 24, 2010 for the BrightSpeed Systems. The letters described the issue and provided safety instructions for the health care provider to be aware if the table should stop unexpectedly and the x-ray remains on to manually stop the scan until the devices software is upgraded. If you have any questions, please contact the US (800) 437-1171, Japan 0120-055-919, Korea 1544-6119, Australia/New Zealand 800 659 465, China 800-810-8188; for other countries, please contact your local GE Healthcare Services Representative.
Recall details
- Recall number
- Z-0369-2011
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Sep 3, 2010
- Posted by the FDA
- Nov 15, 2010
- Recalling firm
- GE Healthcare, LLC
- Quantity
- 946
- Product code
- JAK
- FDA submissions
- K052855, K082761, K082816
- Distribution
- Worldwide distribution: USA including states of: MI, NM, OH, TX, and Puerto Rico and countries including: Algeria, Argentina, Australia, Austria, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Guatemala, Honduras, Hungary, India, Indonesia, Ireland, Islamic Republic of Iran, Israel, Italy, Jamaica, Japan, Kenya, Lebanon, Malaysia, Mauritania, Mexico, Morocco, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Pakistan, Paraguay, Philippines, Poland, Portugal, Republic of Korea, Romania, Russian Federation, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Togo, Tunisia, Turkey, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Viet Nam.
Product description
GE BrightSpeed Excel/Edge/Elite Select / Elite CT Scanner System, models 2377708-560, 5377708-540, 5377708-560. Composed of a gantry, patient table, operator console, computer, and PDU and includes image acquisition hardware, image acquisition and reconstruction software, associated accessories and connections/interfaces to accessories. Is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes, including Axial, Cine, Helical (Volumetric), Cardiac, and Gated acquisitions.
Code information (models, lots, serial numbers)
Show the full code information (15k characters)
Model 2377708-560, serial numbers: 00000018952YC2. 00000018953YC0. 00000018954YC8. 00000019505YC7. 00000019506YC5. 00000019507YC3. 00000019508YC1. 00000019509YC9. 00000019663YC4. 00000019664YC2. 00000019665YC9. 00000019772YC3. 00000019773YC1. 00000019898YC6. 00000019899YC4. 00000020095YC6. 00000020147YC5. 00000020277YC0. 00000020278YC8. 00000020708YC4. 00000020709YC2. 00000020710YC0. 00000020835YC5. 00000020888YC4. 00000020936YC1. 00000020937YC9. 00000021104YC5. 00000021105YC2. 00000021106YC0. 00000021286YC0. 00000021287YC8. 00000021288YC6. 00000021289YC4. 00000021485YC8. 00000021486YC6. 00000021487YC4. 00000021488YC2. 00000021638YC2. 00000021639YC0. 00000021640YC8. 00000021850YC3. 00000021851YC1. 00000021966YC7. 00000021967YC5. 00000021975YC8. 00000021976YC6. 00000022244YC8. 00000022245YC5. 00000023063YC1. 00000023602YC6. 00000023981YC4. 00000024234YC7. 00000024374YC1. 00000025318YC7. 00000208246HM9. 00000210066HM7. 00000210067HM5. 00000210068HM3. 00000210069HM1. 00000210070HM9. 00000210348HM9. 00000210349HM7. 00000210350HM5. 00000211354HM6. 00000211450HM2. 00000211451HM0. 00000212790HM0. 00000212791HM8. 00000212792HM6. 00000212793HM4. 00000214155HM4. 00000214156HM2. 00000214408HM7. 00000214769HM2. 00000214770HM0. 00000214771HM8. 00000214772HM6. 00000214903HM7. 00000215052HM2. 00000215053HM0. 00000215054HM8. 00000215055HM5. 00000215056HM3. 00000215057HM1. 00000215349HM2. 00000215476HM3. 00000215477HM1. 00000215478HM9. 00000215479HM7. 00000216172HM7. 00000216173HM5. 00000216174HM3. 00000216175HM0. 00000216536HM3. 00000216537HM1. 00000217051HM2. 00000217052HM0. 00000217053HM8. 00000217391HM2. 00000217392HM0. 00000217591HM7. 00000217592HM5. 00000217593HM3. 00000217796HM2. 00000217797HM0. 00000218234HM3. 00000218235HM0. 00000218236HM8. 00000218421HM6. 00000218422HM4. 00000218423HM2. 00000218424HM0. 00000218425HM7. 00000218732HM6. 00000218733HM4. 00000219586HM5. 00000219587HM3. 00000219588HM1. 00000219589HM9. 00000219590HM7. 00000219702HM8. 00000219703HM6. 00000219704HM4. 00000220155HM6. 00000220156HM4. 00000220157HM2. 00000220158HM0. 00000220159HM8. 00000220512HM8. 00000220513HM6. 00000220514HM4. 00000220515HM1. 00000221055HM7. 00000221056HM5. 00000221057HM3. 00000221058HM1. 00000221059HM9. 00000221060HM7. 00000221681HM0. 00000221710HM7. 00000221711HM5. 00000221712HM3. 00000221713HM1. 00000221714HM9. 00000222284HM2. 00000222285HM9. 00000222286HM7. 00000222287HM5. 00000222288HM3. 00000222289HM1. 00000222784HM1. 00000222785HM8. 00000222786HM6. 00000222787HM4. 00000222788HM2. 00000222789HM0. 00000222790HM8. 00000222791HM6. 00000223455HM7. 00000223456HM5. 00000223534HM9. 00000223535HM6. 00000223536HM4. 00000223537HM2. 00000223538HM0. 00000223539HM8. 00000223540HM6. 00000223541HM4. 00000223542HM2. 00000223543HM0. 00000223900HM2. 00000223901HM0. 00000223902HM8. 00000223903HM6. 00000223904HM4. 00000224529HM8. 00000224530HM6. 00000224531HM4. 00000224541HM3. 00000224967HM0. 00000225116HM3. 00000225117HM1. 00000225118HM9. 00000225276HM5. 00000225277HM3. 00000225278HM1. 00000225279HM9. 00000225280HM7. 00000225839HM0. 00000225840HM8. 00000225841HM6. 00000225842HM4. 00000226232HM7. 00000226233HM5. 00000226234HM3. 00000226235HM0. 00000226236HM8. 00000226840HM7. 00000226841HM5. 00000226842HM3. 00000226843HM1. 00000227337HM3. 00000227338HM1. 00000227339HM9. 00000227340HM7. 00000227934HM7. 00000227935HM4. 00000227936HM2. 00000227937HM0. 00000227938HM8. 00000230075HM4. 00000230076HM2. 00000230077HM0. 00000231039HM9. 00000231040HM7. 00000231049HM8. 00000231498HM7. 00000231499HM5. 00000231500HM0. 00000231501HM8. 00000231654HM5. 00000234977HM7. 00000235478HM5. 00000236624HM3. 00000236689HM6. 00000236914HM8. 00000237109HM4. 00000237608HM5. 00000237609HM3. 00000237610HM1. 00000237611HM9. 00000239250HM4. 00000239251HM2. 00000239252HM0. 00000241506HM5. 00000241620HM4. 00000242448HM9. 00000242449HM7. 00000243034HM6. 00000243945HM3. 00000244390HM1. 00000244391HM9. 00000244392HM7. 00000245923HM8. 00000245924HM6. Model 5377708-540, serial numbers: 00000017325YC2. 00000017326YC0. 00000017327YC8. 00000017470YC6. 00000017471YC4. 00000017507YC5. 00000017508YC3. 00000017509YC1. 00000017510YC9. 00000017511YC7. 00000017512YC5. 00000017621YC4. 00000017665YC1. 00000017666YC9. 00000017821YC0. 00000017822YC8. 00000017823YC6. 00000017824YC4. 00000017825YC1. 00000017826YC9. 00000017827YC7. 00000018175YC0. 00000018181YC8. 00000018182YC6. 00000018183YC4. 00000018217YC0. 00000018218YC8. 00000018298YC0. 00000018299YC8. 00000018300YC4. 00000018403YC6. 00000018404YC4. 00000018511YC6. 00000018512YC4. 00000018513YC2. 00000018601YC5. 00000018602YC3. 00000018695YC7. 00000018696YC5. 00000018697YC3. 00000018698YC1. 00000018784YC9. 00000018785YC6. 00000018786YC4. 00000018900YC1. 00000018901YC9. 00000018902YC7. 00000018903YC5. 00000018904YC3. 00000018905YC0. 00000018957YC1. 00000019048YC8. 00000019049YC6. 00000019126YC2. 00000019127YC0. 00000019255YC9. 00000019256YC7. 00000019257YC5. 00000019401YC9. 00000019402YC7. 00000019403YC5. 00000019404YC3. 000000208703HM. 00000195292HM8. 00000195293HM6. 00000195294HM4. 00000196294HM3. 00000196295HM0. 00000196296HM8. 00000196312HM3. 00000196313HM1. 00000196314HM9. 00000196638HM1. 00000197187HM8. 00000197346HM0. 00000197355HM1. 00000197356HM9. 00000197785HM9. 00000197786HM7. 00000197787HM5. 00000197788HM3. 00000197789HM1. 00000197790HM9. 00000197809HM7. 00000197810HM5. 00000197811HM3. 00000198797HM3. 00000198798HM1. 00000198799HM9. 00000198800HM5. 00000198801HM3. 00000198802HM1. 00000199556HM2. 00000199557HM0. 00000199558HM8. 00000199559HM6. 00000200149HM3. 00000200153HM5. 00000200154HM3. 00000200161HM8. 00000200690HM6. 00000200691HM4. 00000200692HM2. 00000201095HM7. 00000201099HM9. 00000201100HM5. 00000201115HM3. 00000201560HM0. 00000201561HM8. 00000201562HM6. 00000201563HM4. 00000201576HM6. 00000201577HM4. 00000201578HM2. 00000201596HM4. 00000201597HM2. 00000202433HM9. 00000202434HM7. 00000202435HM4. 00000202436HM2. 00000202620HM1. 00000202624HM3. 00000202625HM0. 00000202626HM8. 00000202627HM6. 00000202628HM4. 00000203596HM2. 00000203600HM2. 00000203601HM0. 00000203602HM8. 00000203603HM6. 00000204144HM0. 00000204145HM7. 00000204163HM0. 00000204164HM8. 00000204197HM8. 00000204198HM6. 00000204199HM4. 00000205488HM0. 00000205489HM8. 00000205496HM3. 00000205497HM1. 00000205498HM9. 00000205936HM8. 00000205937HM6. 00000205938HM4. 00000205939HM2. 00000205962HM4. 00000205963HM2. 00000205964HM0. 00000205965HM7. 00000205966HM5. 00000205967HM3. 00000205998HM8. 00000205999HM6. 00000206848HM4. 00000206849HM2. 00000207034HM0. 00000207035HM7. 00000207361HM7. 00000207362HM5. 00000207741HM0. 00000207742HM8. 00000208226HM1. 00000208227HM9. 00000208228HM7. 00000208229HM5. 00000208230HM3. 00000208253HM5. 00000208691HM6. 00000208692HM4. 00000208701HM3. 00000208702HM1. 00000208703HM9. 00000208704HM7. 00000209561HM0. 00000209578HM4. 00000209579HM2. 00000209611HM3. 00000209612HM1. 00000210016HM2. 00000210017HM0. 00000214731HM2. 00000214734HM6. Model 5377708-560, serial numbers: 00000019027YC2. 00000019577YC6. 00000019578YC4. 00000019579YC2. 00000019651YC9. 00000019652YC7. 00000019808YC5. 00000019809YC3. 00000019910YC9. 00000019911YC7. 00000020098YC0. 00000020152YC5. 00000020153YC3. 00000020283YC8. 00000020284YC6. 00000020456YC0. 00000020457YC8. 00000020458YC6. 00000020471YC9. 00000020522YC9. 00000020523YC7. 00000020711YC8. 00000020712YC6. 00000020713YC4. 00000020757YC1. 00000020758YC9. 00000020837YC1. 00000020942YC9. 00000020943YC7. 00000021158YC1. 00000021159YC9. 00000021160YC7. 00000021264YC7. 00000021270YC4. 00000021271YC2. 00000021272YC0. 00000021666YC3. 00000021667YC1. 00000021668YC9. 00000021669YC7. 00000021670YC5. 00000021767YC9. 00000021768YC7. 00000021769YC5. 00000021880YC0. 00000021881YC8. 00000022160YC6. 00000022161YC4. 00000022162YC2. 00000022163YC0. 00000022259YC6. 00000022260YC4. 00000022261YC2. 00000022352YC9. 00000022353YC7. 00000022457YC6. 00000022458YC4. 00000022459YC2. 00000022460YC0. 00000022705YC8. 00000022706YC6. 00000023030YC0. 00000023031YC8. 00000023245YC4. 00000023246YC2. 00000023474YC0. 00000023475YC7. 00000023862YC6. 00000023863YC4. 00000023864YC2. 00000024039YC0. 00000024040YC8. 00000024153YC9. 00000024154YC7. 00000024354YC3. 00000024355YC0. 00000024356YC8. 00000024357YC6. 00000024470YC7. 00000024471YC5. 00000024704YC9. 00000024705YC6. 00000024706YC4. 00000024707YC2. 00000024807YC0. 00000024808YC8. 00000024809YC6. 00000024810YC4. 00000024811YC2. 00000025053YC0. 00000025054YC8. 00000025206YC4. 00000025207YC2. 00000210086HM5. 00000210090HM7. 00000210091HM5. 00000210111HM1. 00000210112HM9. 00000210113HM7. 00000210723HM3. 00000210724HM1. 00000210725HM8. 00000211321HM5. 00000211322HM3. 00000211323HM1. 00000211334HM8. 00000211335HM5. 00000211336HM3. 00000211337HM1. 00000211458HM5. 00000211459HM3. 00000211460HM1. 00000212434HM5. 00000212438HM6. 00000212439HM4. 00000212440HM2. 00000212441HM0. 00000213223HM1. 00000213224HM9. 00000213225HM6. 00000213226HM4. 00000213227HM2. 00000213252HM0. 00000213800HM6. 00000213801HM4. 00000213945HM9. 00000213946HM7. 00000213947HM5. 00000213948HM3. 00000213982HM2. 00000213986HM3. 00000213987HM1. 00000213988HM9. 00000213989HM7. 00000213990HM5. 00000214143HM0. 00000214144HM8. 00000214145HM5. 00000214412HM9. 00000214416HM0. 00000214417HM8. 00000214418HM6. 00000214428HM5. 00000214429HM3. 00000214430HM1. 00000214761HM9. 00000214762HM7. 00000214844HM3. 00000214845HM0. 00000215384HM9. 00000215385HM6. 00000215386HM4. 00000215468HM0. 00000215469HM8. 00000215655HM2. 00000215656HM0. 00000215657HM8. 00000216140HM4. 00000216141HM2. 00000216142HM0. 00000216160HM2. 00000216161HM0. 00000216162HM8. 00000216516HM5. 00000216517HM3. 00000216518HM1. 00000216519HM9. 00000216520HM7. 00000217031HM4. 00000217032HM2. 00000217039HM7. 00000217040HM5. 00000217041HM3. 00000217370HM6. 00000217371HM4. 00000217372HM2. 00000217382HM1. 00000217603HM0. 00000217604HM8. 00000217611HM3. 00000217612HM1. 00000217613HM9. 00000217614HM7. 00000217769HM9. 00000217770HM7. 00000217788HM9. 00000217789HM7. 00000217804HM4. 00000218226HM9. 00000218230HM1. 00000218248HM3. 00000218261HM6. 00000218262HM4. 00000218408HM3. 00000218409HM1. 00000218410HM9. 00000218446HM3. 00000218450HM5. 00000218755HM7. 00000218756HM5. 00000218757HM3. 00000218767HM2. 00000218768HM0. 00000218769HM8. 00000218770HM6. 00000218771HM4. 00000218791HM2. 00000218792HM0. 00000219088HM2. 00000219089HM0. 00000219096HM5. 00000219100HM5. 00000219606HM1. 00000219610HM3. 00000219656HM6. 00000219657HM4. 00000219664HM0. 00000219665HM7. 00000219666HM5. 00000219667HM3. 00000219667HM7. 00000220127HM5. 00000220143HM2. 00000220144HM0. 00000220145HM7. 00000220496HM4. 00000220497HM2. 00000220498HM0. 00000220499HM8. 00000220528HM4. 00000220529HM2. 00000221079HM7. 00000221080HM5. 00000221081HM3. 00000221082HM1. 00000221095HM3. 00000221730HM5. 00000221731HM3. 00000221732HM1. 00000221742HM0. 00000222227HM1. 00000222228HM9. 00000222229HM7. 00000222239HM6. 00000222240HM4. 00000222241HM2. 00000222242HM0. 00000222243HM8. 00000222244HM6. 00000222245HM3. 00000222267HM7. 00000222268HM5. 00000222269HM3. 00000222821HM1. 00000222825HM2. 00000222826HM0. 00000222827HM8. 00000222828HM6. 00000223506HM7. 00000223507HM5. 00000223508HM3. 00000223509HM1. 00000223510HM9. 00000223511HM7. 00000223530HM7. 00000223876HM4. 00000223877HM2. 00000223878HM0. 00000223879HM8. 00000223880HM6. 00000223881HM4. 00000224143HM8. 00000224545HM4. 00000224546HM2. 00000224547HM0. 00000224978HM7. 00000224982HM9. 00000225007HM4. 00000225008HM2. 00000225248HM4. 00000225249HM2. 00000225250HM0. 00000225251HM8. 00000225264HM1. 00000225265HM8. 00000225266HM6. 00000225316HM9. 00000225862HM2. 00000225863HM0. 00000225864HM8. 00000225865HM5. 00000226063HM6. 00000226064HM4. 00000226065HM1. 00000226066HM9. 00000226079HM2. 00000226080HM0. 00000226718HM5. 00000226719HM3. 00000226720HM1. 00000226721HM9. 00000226785HM4. 00000226856HM3. 00000226860HM5. 00000226861HM3. 00000226862HM1. 00000226863HM9. 00000227325HM8. 00000227326HM6. 00000227327HM4. 00000227353HM0. 00000227862HM0. 00000227863HM8. 00000227864HM6. 00000227865HM3. 00000227866HM1. 00000227867HM9. 00000227868HM7. 00000227869HM5. 00000227870HM3. 00000227871HM1. 00000227922HM2. 00000227923HM0. 00000227924HM8. 00000228625HM0. 00000228626HM8. 00000228627HM6. 00000228628HM4. 00000228629HM2. 00000228665HM6. 00000228666HM4. 00000228673HM0. 00000228827HM2. 00000228831HM4. 00000228832HM2. 00000228833HM0. 00000228834HM8. 00000228835HM5. 00000229315HM7. 00000229477HM5. 00000229478HM3. 00000229479HM1. 00000229489HM0. 00000229490HM8. 00000229497HM3. 00000229499HM9. 00000229500HM4. 00000230027HM5. 00000230028HM3. 00000230043HM2. 00000230044HM0. 00000230045HM7. 00000230055HM6. 00000230056HM4. 00000230057HM2. 00000230059HM8. 00000230989HM6. 00000230990HM4. 00000230999HM5. 00000231000HM1. 00000231001HM9. 00000231002HM7. 00000231003HM5. 00000231004HM3. 00000231029HM0. 00000231030HM8. 00000231443HM3. 00000231444HM1. 00000231445HM8. 00000231446HM6. 00000231447HM4. 00000231448HM2. 00000231449HM0. 00000231478HM9. 00000231479HM7. 00000231488HM8. 00000231489HM6. 00000231659HM4. 00000232027HM3. 00000232028HM1. 00000233011HM6. 00000233012HM4. 00000233382HM1. 00000233383HM9. 00000233392HM0. 00000233409HM2. 00000233410HM0. 00000233411HM8. 00000233412HM6. 00000233893HM7. 00000234560HM1. 00000234561HM9. 00000234562HM7. 00000234576HM7. 00000234577HM5. 00000234648HM4. 00000234653HM4. 00000234654HM2. 00000234692HM2. 00000234693HM0. 00000235045HM2. 00000235050HM2. 00000235251HM6. 00000235252HM4. 00000235253HM2. 00000235254HM0. 00000235255HM7. 00000235256HM5. 00000235291HM2. 00000235292HM0. 00000235293HM8. 00000235490HM0. 00000235491HM8. 00000235492HM6. 00000235493HM4. 00000235891HM9. 00000235892HM7. 00000235893HM5. 00000235894HM3. 00000235895HM0. 00000236431HM3. 00000236432HM1. 00000236563HM3. 00000236564HM1. 00000236565HM8. 00000236566HM6. 00000236567HM4. 00000236609HM4. 00000236610HM2. 00000236611HM0. 00000236674HM8. 00000236675HM5. 00000236676HM3. 00000237092HM2. 00000237094HM8. 00000237114HM4. 00000237655HM6. 00000237656HM4. 00000237657HM2. 00000237658HM0. 00000237659HM8. 00000237660HM6. 00000237661HM4. 00000237662HM2. 00000237698HM6. 00000237699HM4. 00000237700HM0. 00000238529HM2. 00000238531HM8. 00000238532HM6. 00000238549HM0. 00000239129HM0. 00000239147HM2. 00000239148HM0. 00000239230HM6. 00000239231HM4. 00000239232HM2. 00000239233HM0. 00000239265HM2. 00000239266HM0. 00000239267HM8. 00000239268HM6. 00000239269HM4. 00000239270HM2. 00000239271HM0. 00000239272HM8. 00000239305HM6. 00000239306HM4. 00000239307HM2. 00000240040HM6. 00000240041HM4. 00000240042HM2. 00000240305HM3. 00000240306HM1. 00000240307HM9. 00000240308HM7. 00000240309HM5. 00000241013HM2. 00000241022HM3. 00000241023HM1. 00000241024HM9. 00000241025HM6. 00000241131HM2. 00000241636HM0. 00000241649HM3. 00000241650HM1. 00000241651HM9. 00000241652HM7. 00000241669HM1. 00000241828HM3. 00000241898HM6. 00000241903HM4. 00000242393HM7. 00000242394HM5. 00000242395HM2. 00000242396HM0. 00000242468HM7. 00000242503HM1. 00000242942HM1. 00000243029HM6. 00000243040HM3. 00000243066HM8. 00000243633HM5. 00000243634HM3. 00000243635HM0. 00000244411HM5. 00000244412HM3. 00000244413HM1. 00000244414HM9. 00000244415HM6. 00000245385HM0. 00000245396HM7. 00000246483HM2. 00000247233HM0. 00000247234HM8. 00000247263HM7. 00000247775HM0.
Other devices under the same FDA submission
3 devices cleared under K052855, K082761, K082816 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Operator Console for BrightSpeed Elite and BrightSpeed Elite SelectGE Hangwei Medical Systems Company Ltd. · System, X-Ray, Tomography, Computed1 recall0 reports
- Optima CT540GE Hangwei Medical Systems Company Ltd. · System, x-ray, tomography, computed1 recall0 reports
- Optima CT540 NGE Hangwei Medical Systems Company Ltd. · System, x-ray, tomography, computed1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 56814).
- Z-0368-2011 · TerminatedGE LightSpeed VCT Scanner System, models 5212920-300, 5212920-310, 5311595-10. Composed of a gantry, patient table, operator console, computer, and PDU and includes image acquisit…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page