CARTO 3 System Interface Cable, 34-pin to 10-pin, Red, 10'… recallZ-0884-2011 · Biosense Webster, Inc. · initiated Oct 5, 2010
Terminated All recalls posted in January 2011
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Biosense Webster, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
When connecting this cable to the mapping system, loss of Body Surface ECG signals may occur. This may result in inaccurate locations appearing for catheters displayed 3D on the CARTO monitor.
FDA root cause category: Device Design
What the company is doing
Biosense Webster issued an Urgent Field Safety Notice letter dated October 5, 2010. The letter identified the product, the problem,and the action to be taken by the customer. Customers were instructed to return the affected cable, accompanied by the attached Return Authorization form. For questions regarding this recall call Biosense Webster at 909 839-8874.
Recall details
- Recall number
- Z-0884-2011
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Oct 5, 2010
- Posted by the FDA
- Jan 7, 2011
- Recalling firm
- Biosense Webster, Inc.
- Quantity
- 1692
- Product code
- DSA
- FDA submissions
- K090017
- Distribution
- Worldwide Distribution: USA, including the states of: AK, AL, AR, AZ, CA, CO, CT, FL, GA, IA, IL, KS, KY, MA, MD, MI, MN, NC, NE, NJ, NV, NY, OH, OR, PA, SC, TN, TX, UT, WA, WI, and WV, and the countries of Belgium, China, Israel, Japan, Singapore, and Thailand.
Product description
CARTO 3 System Interface Cable, 34-pin to 10-pin, Red, 10' - Part Number: D-1286-18; Catalog Number CR3410CT This product provides a signal interface between a Biosense Webster electrophysiology catheter and the appropriate equipment. This cable may be re-used subject to established cleaning and sterilization restrictions
Code information (models, lots, serial numbers)
Part Number: D-1286-18; Catalog Number CR3410CT
Other devices under the same FDA submission
6 devices cleared under K090017 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- CARTO 3 FG540000Biosense Webster (Israel) Ltd. · Computer, diagnostic, programmable1 death437 reports
- Carto 3 CR3434CTBiosense Webster Inc · COMPUTER, DIAGNOSTIC, PROGRAMMABLE2 reports
- Carto 3 CB3410CTBiosense Webster Inc · COMPUTER, DIAGNOSTIC, PROGRAMMABLE1 report
- Carto 3 CB3412CTBiosense Webster Inc · COMPUTER, DIAGNOSTIC, PROGRAMMABLE1 report
- Carto 3 CB3434CTBiosense Webster Inc · COMPUTER, DIAGNOSTIC, PROGRAMMABLE1 report
- Carto 3 CY1212CTBiosense Webster Inc · COMPUTER, DIAGNOSTIC, PROGRAMMABLE1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.