Skip to content

Micrus Endovascular Microcoil Delivery System recallZ-0845-2011 · Micrus Endovascular Corp., 821 Fox Lane, San Jose, CA 95131 · initiated Oct 21, 2010

Terminated All recalls posted in January 2011

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Micrus Endovascular Corp., 821 Fox Lane, San Jose, CA 95131 or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The firm has received reports of difficulty connecting to the detachment control cable. Loss of the electrical connection will result in failure to detach the coil for deployment inside an aneurysm.

FDA root cause category: Other

What the company is doing

Micrus Endovascular Corp sent Urgent: Medical Device Recall Letters dated October 21, 2010, to all customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to examine inventory (including consignment) and quarantine the product to prevent it from being used or shipped. If the product was further distributed, customers were to notify their customers by sending a copy of the recall letter and were instructed to work with all customers to ensure that all product was returned and that pre-use protocol was followed. The response form was to be completed and returned indicating the number of devices that had been used (or shipped if distributor) and the number that would be returned. Customers were to call the Micrus Customer Service Department at 1-888-550-4129 from 8:00 am to 5:00 Pacific time for arrange for a Return Authorization number and replacement. Recall letters also stressed the importance of always following the pre-use protocol and inspecting the device immediately prior to use. Additionally, connector functionality should be verified and instructions were provided. Questions or concerns should be directed to Micrus Customer Service at 1-888-4120. Consignees who did not respond to the initial recall notification after 15 days were notified again by letter. A third letter was sent after another 15 days if there was still no response.

Recall details

Recall number
Z-0845-2011
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Oct 21, 2010
Posted by the FDA
Jan 3, 2011
Recalling firm
Micrus Endovascular Corp., 821 Fox Lane, San Jose, CA 95131
Quantity
6755
Product code
HCG
FDA submissions
K062036, K072173, K083646
Distribution
Worldwide Distribution - USA including AK, AL, AZ, CA, CO, DC, DE, FL, GA, ID, IL, IN, KY, LA, MA, MD, MI, MN, NC, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV and the countries of Australia, Austria, Belgium, Canada, Chile, Czech Republic, Denmark, France, Germany, Hungary, Ireland, Italy, Japan, Netherlands, Poland, Republic of Korea, Russia, Singapore, Slovakia, Spain, Sweden, Switzerland, Turkey, and the United Kingdom

Product description

Micrus Endovascular Microcoil Delivery System; The Micrus Microcoil Delivery System consists of three components, a Microcoil System, a Connecting Cable, and a Detachment Control Box (DCB). Each component is sold separately. Catalog Numbers: PC4180933, SSR181128, CPL100252, PC4181647, SSR181640, CPL100404, DFS100254, SSR181230, CPL100406, DFS100410, DPL100206, SRC141230, CRC140304, DFS100716, CRC140406, CPL100203, DFS100204, DPL100408, SSR100253, PC4181240, DPL100202, CDF100408, PC4181137, CSP100500, CDF100154, PC4180830, CSP100350, DFS100152, SRC140408, PC4181034, CPL100206, CDF100152, SRC141025, SRC140225, CPL100304, CRC141025, CPL 100254, PC4181447, DPL100252, DFS100154, SRC140303, DFS100720, DPL 100204, PC4181343, CSP180500, DFS1 0031 0, DPL100306, DPL100256, CPL100306, CDF100304, CDF100203, CDF100410, PC4100412, DPL 100254, DPL 100053, PC41 00626, CDF100208, CRC140306, PC4100517, DFS100203, CSP181400, CPL 100152, DFS100210, CDF100720, CPL100204, CSP180200, DFS100204, CDF100510, DFS101025, DFS100408, PC41 00730, SSR181740, CRC140615, CPL100256, CDF100206, DFS100308, CPL100151, CSP100200, CDF100616, CRC140711, SSR181334, CDF100310, SRC140445, DFS100252, CPL100408, CDF100306, DFS100250, DPL 100152, DPL 100201, CDF100258, SRC140306, CDF100156, DPL 100153, DPL100304, SRC140922, SSR100825, DFS100206, DPL 100203, PC4181550, DFS100616, CSP180900, PC4181950, DFS100406, SRC140717, PC4181750, DFS100925, CSP100400, PC4181846, SRC140711, CPL100153, PC4182050, DFS100306, SRC140406, CSP100900, SRC140615, CDF100308, CSP100300, DFS1 0051 0, CDF100406, CDF1 0021 0, DFS100156, CDF100515, DPL100151, CDF100925, SSR181028, CDF101025, SSR181540, CPL 100201, SSR180825, CPL 100202; Manufactured and distributed by Micrus Endovascular Corporation, San Jose, CA

Code information (models, lots, serial numbers)

Show the full code information (2k characters)

Lot Numbers: F62328 ,G11073 ,G11135 ,G11205 ,G11271 ,G11368  ,F62370 ,G11074 ,G11138 ,G11208 ,G11273 ,G11370  ,G10831 ,G11075 ,G11139 ,G11209 ,G11274 ,G11371  ,G10874 ,G11076 ,G11140 ,G11210 ,G11275 ,G11374  ,G10889 ,G11079 ,G11144 ,G11211 ,G11276 ,G11376  ,G10915 ,G11081 ,G11146 ,G11212 ,G11279 ,G11379  ,G10928 ,G11082 ,G11147 ,G11213 ,G11280 ,G11381  ,G10939 ,G11084 ,G11149 ,G11214 ,G11281 ,G11385  ,G10940 ,G11086 ,G11151 ,G11216 ,G11283 ,G11386  ,G10949 ,G11087 ,G11152 ,G11217 ,G11284 ,G11567  ,G10951 ,G11088 ,G11153 ,G11218 ,G11285 ,G11604  ,G10953 ,G11090 ,G11154 ,G11219 ,G11287 ,G11698  ,G10962 ,G11093 ,G11155 ,G11223 ,G11289 ,G11741  ,G10965 ,G11094 ,G11157 ,G11224 ,G11293 J10184  ,G10967 ,G11095 ,G11165 ,G11225 ,G11294 J10185  ,G10969 ,G11098 ,G11166 ,G11228 ,G11295 J10186  ,G10971 ,G11100 ,G11171 ,G11229 ,G11297 J10187  ,G10987 ,G11101 ,G11172 ,G11230 ,G11298 J10188  ,G10995 ,G11102 ,G11173 ,G11233 ,G11303 J10189  ,G10996 ,G11106 ,G11175 ,G11235 ,G11304 J10190  ,G10997 ,G11109 ,G11176 ,G11238 ,G11307 J10191  ,G11007 ,G11111 ,G11177 ,G11239 ,G11310 J10192  ,G11009 ,G11112 ,G11180 ,G11242 ,G11314 J10193  ,G11010 ,G11114 ,G11182 ,G11244 ,G11315 J10194  ,G11025 ,G11115 ,G11184 ,G11246 ,G11316 J10195  ,G11027 ,G11117 ,G11185 ,G11249 ,G11318 J10196  ,G11032 ,G11118 ,G11186 ,G11251 ,G11325 J10219  ,G11034 ,G11121 ,G11187 ,G11253 ,G11326 ,G10946  ,G11044 ,G11122 ,G11191 ,G11256 ,G11327 ,G11062  ,G11050 ,G11126 ,G11192 ,G11258 ,G11328 ,G11369  ,G11053 ,G11127 ,G11193 ,G11262 ,G11332  ,G11054 ,G11128 ,G11194 ,G11263 ,G11333  ,G11055 ,G11129 ,G11195 ,G11265 ,G11335  ,G11065 ,G11131 ,G11198 ,G11268 ,G11350  ,G11068 ,G11132 ,G11199 ,G11269 ,G11356  ,G11071 ,G11133 ,G11200 ,G11270 ,G11364

Other devices under the same FDA submission

19 devices cleared under K062036, K072173, K083646 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates