Medical device recalls: July 2006
The FDA posted 136 medical device recall records in July 2006, from 54 recall events by 44 companies. 136 have not been classified yet.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Radiometer America Inc · 3 events, 3 records
- Smith & Nephew, Inc. Endoscopy Division · 2 events, 8 records
- Kensey Nash Corp · 2 events, 4 records
- Terumo Cardiovascular Systems Corp · 2 events, 4 records
- Abbott Laboratories MPG · 2 events, 2 records
- Beckman Coulter Inc · 2 events, 2 records
- Boston Scientific Corp · 2 events, 2 records
- International Biophysics Corp · 2 events, 2 records
FDA root cause
- Other · 131
- Package design/selection · 2
- Device Design · 2
- Component change control · 1
Not yet classified (136)
The FDA has posted these records but has not published a class for them yet.
- Z-1241-06 · Kimberly-Clark Corporation · posted Jul 28, 2006 · Terminated
The tinfoil in the triangle electrodes has the potential to develop cracks which may hinder electrical flow and product performance during defibrillation.
- Z-1242-06 · Kimberly-Clark Corporation · posted Jul 28, 2006 · Terminated
The tinfoil in the triangle electrodes has the potential to develop cracks which may hinder electrical flow and product performance during defibrillation.
- Z-1305-06 · Kensey Nash Corp · posted Jul 28, 2006 · Terminated
Alarm activiation-A priming issue involving the flow control unit due to a software problem has caused false postive Extraction Line Block (ELB) alarming. Use of this product may result in a health hazard resulting from early termination of the flush and…
- Z-1243-06 · SCC Soft Computer · posted Jul 27, 2006 · Terminated
Text from one case is overwriting another. This could cause an incorrect diagnostic report to be sent to a clinician/surgeon.
- Z-1255-06 · Brainlab AG · posted Jul 27, 2006 · Terminated
During ExacTrac Robotics installation on the Varian Exact Couch, the couch height position indication is re-calibrated to read out the correct positions. This re-calibration could cause the eventual failure of the vertical lift mechanism of the couch,…
- Z-1256-06 · AGFA Corp. · posted Jul 27, 2006 · Terminated
Possible corrupted image appearing after System Start.
- Z-1261-06 · Radiometer America Inc · posted Jul 27, 2006 · Terminated
Air bubbles may be trapped on or close to the O2 sensor in the pO2 chamber of the cuvettes in cartridges used for blood gas analysis on NPT7 analyzer. This produces too high pO2 readings without a question mark.
- Z-1262-06 · Radiometer America Inc · posted Jul 27, 2006 · Terminated
Software anomaly. Data generated by blood gas equipment did not correlate with data shown in the Radiance Data Management System.
- Z-1266-06 · Terumo Cardiovascular Systems Corp · posted Jul 27, 2006 · Terminated
The pump may stop during the autodose delivery without completion of the dose delivery to the patient.
- Z-1267-06 · Terumo Cardiovascular Systems Corp · posted Jul 27, 2006 · Terminated
The pump may stop during the autodose delivery without completion of the dose delivery to the patient.
- Z-1268-06 · Varian Medical Systems · posted Jul 27, 2006 · Terminated
Medical device software for brachytherapy may cause erroneous data to be recorded and affect patient radiation treatments. Erroneous data may be listed on the treatment history report, and the default step size may lead to a misadministration if treatment…
- Z-1269-06 · Varian Medical Systems · posted Jul 27, 2006 · Terminated
Medical device software for brachytherapy may cause erroneous data to be recorded and affect patient radiation treatments. Erroneous data may be listed on the treatment history report, and the default step size may lead to a misadministration if treatment…
- Z-1270-06 · Varian Medical Systems · posted Jul 27, 2006 · Terminated
Medical device software for brachytherapy may cause erroneous data to be recorded and affect patient radiation treatments. Erroneous data may be listed on the treatment history report, and the default step size may lead to a misadministration if treatment…
- Z-1271-06 · Zimmer Inc. · posted Jul 27, 2006 · Terminated
During provisional trialing, the instrument may become stuck in the humeral canal and require surgical removal.
- Z-1272-06 · Zimmer Inc. · posted Jul 27, 2006 · Terminated
During provisional trialing, the instrument may become stuck in the humeral canal and require surgical removal.
- Z-1273-06 · Zimmer Inc. · posted Jul 27, 2006 · Terminated
During provisional trialing, the instrument may become stuck in the humeral canal and require surgical removal.
- Z-1274-06 · Zimmer Inc. · posted Jul 27, 2006 · Terminated
During provisional trialing, the instrument may become stuck in the humeral canal and require surgical removal.
- Z-1275-06 · Zimmer Inc. · posted Jul 27, 2006 · Terminated
During provisional trialing, the instrument may become stuck in the humeral canal and require surgical removal.
- Z-1276-06 · Zimmer Inc. · posted Jul 27, 2006 · Terminated
During provisional trialing, the instrument may become stuck in the humeral canal and require surgical removal.
- Z-1277-06 · Zimmer Inc. · posted Jul 27, 2006 · Terminated
During provisional trialing, the instrument may become stuck in the humeral canal and require surgical removal.
- Z-1278-06 · Zimmer Inc. · posted Jul 27, 2006 · Terminated
During provisional trialing, the instrument may become stuck in the humeral canal and require surgical removal.
- Z-1279-06 · Zimmer Inc. · posted Jul 27, 2006 · Terminated
During provisional trialing, the instrument may become stuck in the humeral canal and require surgical removal.
- Z-1280-06 · Zimmer Inc. · posted Jul 27, 2006 · Terminated
During provisional trialing, the instrument may become stuck in the humeral canal and require surgical removal.
- Z-1281-06 · Zimmer Inc. · posted Jul 27, 2006 · Terminated
During provisional trialing, the instrument may become stuck in the humeral canal and require surgical removal.
- Z-1282-06 · Zimmer Inc. · posted Jul 27, 2006 · Terminated
During provisional trialing, the instrument may become stuck in the humeral canal and require surgical removal.
- Z-1283-06 · Zimmer Inc. · posted Jul 27, 2006 · Terminated
During provisional trialing, the instrument may become stuck in the humeral canal and require surgical removal.
- Z-1284-06 · Zimmer Inc. · posted Jul 27, 2006 · Terminated
During provisional trialing, the instrument may become stuck in the humeral canal and require surgical removal.
- Z-1285-06 · Zimmer Inc. · posted Jul 27, 2006 · Terminated
During provisional trialing, the instrument may become stuck in the humeral canal and require surgical removal.
- Z-1286-06 · Terumo Cardiovascular Systems Corp · posted Jul 27, 2006 · Terminated
While on battery power, the system may flash a low battery warning even though the battery is adequately charged.
- Z-1287-06 · Terumo Cardiovascular Systems Corp · posted Jul 27, 2006 · Terminated
While on battery power, the system may flash a low battery warning even though the battery is adequately charged.
- Z-1288-06 · Boston Scientific CRM Corp · posted Jul 27, 2006 · Terminated
Guidant has identified a pattern of premature battery depletion in devices manufactured with a single lot of capacitors from a single supplier. If function of this low voltage capacitor is compromised, the battery may deplete prematurely.
- Z-1289-06 · Boston Scientific CRM Corp · posted Jul 27, 2006 · Terminated
Guidant has identified a pattern of premature battery depletion in devices manufactured with a single lot of capacitors from a single supplier. If function of this low voltage capacitor is compromised, the battery may deplete prematurely.
- Z-1290-06 · Boston Scientific CRM Corp · posted Jul 27, 2006 · Terminated
Guidant has identified a pattern of premature battery depletion in devices manufactured with a single lot of capacitors from a single supplier. If function of this low voltage capacitor is compromised, the battery may deplete prematurely.
- Z-1291-06 · Boston Scientific CRM Corp · posted Jul 27, 2006 · Terminated
Guidant has identified a pattern of premature battery depletion in devices manufactured with a single lot of capacitors from a single supplier. If function of this low voltage capacitor is compromised, the battery may deplete prematurely.
- Z-1292-06 · Boston Scientific CRM Corp · posted Jul 27, 2006 · Terminated
Guidant has identified a pattern of premature battery depletion in devices manufactured with a single lot of capacitors from a single supplier. If function of this low voltage capacitor is compromised, the battery may deplete prematurely.
- Z-1300-06 · Boston Scientific Corp · posted Jul 27, 2006 · Terminated
Sterility is compromised because the product is labeled as sterile, but the packages may have holes in them.
- Z-1252-06 · SCC Soft Computer · posted Jul 26, 2006 · Terminated
The client had made a decision to use the SoftPath Module in a nonstandard manner. Under specific circumstances, incorrect text may appear on a patient report.
- Z-1257-06 · Hebron International, Inc. · posted Jul 26, 2006 · Terminated
Contact lenses were distributed and sold to inappropriate retail establishments without professional eye care involvement.
- Z-1258-06 · St. Francis Medical Technologies Inc · posted Jul 26, 2006 · Terminated
Some units may be mislabeled as 10 mm X STOP implant that actually contain 14 mm X STOP implant.
- Z-1259-06 · Beckman Coulter Inc · posted Jul 26, 2006 · Terminated
Beckman Coulter has determined that under certain conditions a system error generated during a Pause and Rotate routine may result in a sample misidentification with a risk of reporting erroneous results.
- Z-1260-06 · Radiometer America Inc · posted Jul 26, 2006 · Terminated
An increased failure rate was observed with the yellow rubber tubing. The tubing was found to be leaking blood and solutions after distribution.
- Z-1253-06 · Baxter Healthcare Renal Div · posted Jul 25, 2006 · Terminated
The UF Removal Regulators were assembled incorrectly. The mis-assembly results in a failure to successfully calibrate the instrument and begin patient therapy.
- Z-1254-06 · Nicolet Biomedical Div of Viasys Healthcare · posted Jul 25, 2006 · Terminated
The Bar Electrodes do not meet performance specification for electrode impedance. A manufacturing defect was discovered in disposable bar electrodes, part number 019-435600, lot 604664, that has the potential to reduce the amplitude of a motor or sensory…
- Z-1263-06 · Microcuff Gmbh · posted Jul 25, 2006 · Terminated
Inadequate seals and/or holes in the pouch may compromise the sterility of the product.
- Z-1264-06 · Microcuff Gmbh · posted Jul 25, 2006 · Terminated
Inadequate seals and/or holes in the pouch may compromise the sterility of the product.
- Z-1265-06 · Microcuff Gmbh · posted Jul 25, 2006 · Terminated
Inadequate seals and/or holes in the pouch may compromise the sterility of the product.
- Z-1249-06 · Hamilton Medical Inc · posted Jul 22, 2006 · Terminated
ALARM FAILURE-Following an oxygen cell calibration, the user may inadvertently and unknowingly disable the alarm system.
- Z-1250-06 · Hamilton Medical Inc · posted Jul 22, 2006 · Terminated
ALARM FAILURE-Following an oxygen cell calibration, the user may inadvertently and unknowingly disable the alarm system.
- Z-1251-06 · Hamilton Medical Inc · posted Jul 22, 2006 · Terminated
ALARM FAILURE-Following an oxygen cell calibration, the user may inadvertently and unknowingly disable the alarm system.
- Z-1248-06 · Celsion Corporation · posted Jul 19, 2006 · Terminated
A medical device accessory used by healthcare practitioners in the invasive treatment of (BPH) benign prostatic hyperplasia may be defective. The catheter may fail to achieve or maintain recommended pressure.
- Z-1244-06 · Abbott Laboratories MPG · posted Jul 18, 2006 · Terminated
The CA 15-3 Controls replicates out of range high resulting in invalid runs.
- Z-1245-06 · Beckman Coulter Inc · posted Jul 18, 2006 · Terminated
Beckman has confirmed that erroneous results could be reported when the workstation database crashes.
- Z-1246-06 · Instrumentation Laboratory Co. · posted Jul 18, 2006 · Terminated
Firm received 2 complaints concerning incorrect prothrombin time (PT) results while using the reagents on samples of patients taking the antibiotic CUBICIN (Daptomycin for injection).
- Z-1247-06 · Instrumentation Laboratory Co. · posted Jul 18, 2006 · Terminated
Firm received 2 complaints concerning incorrect prothrombin time (PT) results while using the reagents on samples of patients taking the antibiotic CUBICIN (Daptomycin for injection).
- Z-1225-06 · Teleflex Medical · posted Jul 13, 2006 · Terminated
Teleflex Medical has identified that the product may malfunction, causing the handle to heat up. There is a potential for the heated handle to burn the user.
- Z-1226-06 · Smith & Nephew, Inc. Endoscopy Division · posted Jul 13, 2006 · Terminated
The sterility seal of the package trays may have incomplete seals compromising the sterility of the device
- Z-1227-06 · Smith & Nephew, Inc. Endoscopy Division · posted Jul 13, 2006 · Terminated
The sterility seal of the package trays may have incomplete seals compromising the sterility of the device
- Z-1228-06 · Smith & Nephew, Inc. Endoscopy Division · posted Jul 13, 2006 · Terminated
The sterility seal of the package trays may have incomplete seals compromising the sterility of the device
- Z-1229-06 · Smith & Nephew, Inc. Endoscopy Division · posted Jul 13, 2006 · Terminated
The sterility seal of the package trays may have incomplete seals compromising the sterility of the device
- Z-1230-06 · Smith & Nephew, Inc. Endoscopy Division · posted Jul 13, 2006 · Terminated
Sterility seal of the package tray may have gaps/voids compromising the sterility of the device.
- Z-1231-06 · Smith & Nephew, Inc. Endoscopy Division · posted Jul 13, 2006 · Terminated
Sterility seal of the package tray may have gaps/voids compromising the sterility of the device.
- Z-1232-06 · Smith & Nephew, Inc. Endoscopy Division · posted Jul 13, 2006 · Terminated
Sterility seal of the package tray may have gaps/voids compromising the sterility of the device.
- Z-1233-06 · Smith & Nephew, Inc. Endoscopy Division · posted Jul 13, 2006 · Terminated
Sterility seal of the package tray may have gaps/voids compromising the sterility of the device.
- Z-1234-06 · 3M Company / Medical Division · posted Jul 13, 2006 · Terminated
After reviewing manufacturing records it has been determined that these lots inadvertently were not sterilized after packaging in individual boxes labeled as sterile product.
- Z-1235-06 · Siemens Medical Solutions, USA, Inc · posted Jul 13, 2006 · Terminated
The cardiologist's report comments may not be retained by the system due to a software bug.
- Z-1236-06 · W L Gore & Assoc, Inc. · posted Jul 13, 2006 · Terminated
Two lots of uniquely different product were switched and mislabeled with the wrong length. 34mm x 10cm devices were mislabeled as 34mm x 15cm and vice versa.
- Z-1237-06 · W L Gore & Assoc, Inc. · posted Jul 13, 2006 · Terminated
Two lots of uniquely different product were switched and mislabeled with the wrong length. 34mm x 10cm devices were mislabeled as 34mm x 15cm and vice versa.
- Z-1238-06 · Boston Scientific Corp · posted Jul 13, 2006 · Terminated
Instructions on how to inject are not included in the package.
- Z-1239-06 · Aksys, Ltd. · posted Jul 13, 2006 · Terminated
The hemodialysis device was being marketed with a modified treatment length that exceeds the treatment length for which the device was originally designed.
- Z-1240-06 · Cordis Corporation · posted Jul 13, 2006 · Terminated
The tip of the catheter may separate from the body during operation.
- Z-1196-06 · GE Healthcare · posted Jul 12, 2006 · Terminated
There exists the possibility that the ADU Fresh Gas Control unit (A-FGC1) can shut off during normal handling of the anesthesia machine and result in no anesthetic agent output.
- Z-1207-06 · Sammons Preston Rolyan · posted Jul 12, 2006 · Terminated
There have been reports of structural failure on the skirt of select tables. The risk to patients presented by possible structural failure of the skirt of the table is a slow collapse of the plinth which could cause the patient to be displaced.
- Z-1208-06 · Sammons Preston Rolyan · posted Jul 12, 2006 · Terminated
There have been reports of structural failure on the skirt of select tables. The risk to patients presented by possible structural failure of the skirt of the table is a slow collapse of the plinth which could cause the patient to be displaced.
- Z-1209-06 · Sammons Preston Rolyan · posted Jul 12, 2006 · Terminated
There have been reports of structural failure on the skirt of select tables. The risk to patients presented by possible structural failure of the skirt of the table is a slow collapse of the plinth which could cause the patient to be displaced.
- Z-1210-06 · Sammons Preston Rolyan · posted Jul 12, 2006 · Terminated
There have been reports of structural failure on the skirt of select tables. The risk to patients presented by possible structural failure of the skirt of the table is a slow collapse of the plinth which could cause the patient to be displaced.
- Z-1211-06 · Sammons Preston Rolyan · posted Jul 12, 2006 · Terminated
There have been reports of structural failure on the skirt of select tables. The risk to patients presented by possible structural failure of the skirt of the table is a slow collapse of the plinth which could cause the patient to be displaced.
- Z-1212-06 · Vanguard Medical Concepts, Inc · posted Jul 12, 2006 · Terminated
Review of complaints and testing show that the distal tip of the trocars may separate while in use.
- Z-1213-06 · Boston Scientific · posted Jul 12, 2006 · Terminated
Boston Scientific discovered through an internal inspection process that the PTFE (polytetrafluoroethylene) coating is not consistently adhering to the transition area of the distal end of the guide wires which may cause PTFE to flake off.
- Z-1214-06 · Boston Scientific · posted Jul 12, 2006 · Terminated
Boston Scientific discovered through an internal inspection process that the PTFE (polytetrafluoroethylene) coating is not consistently adhering to the transition area of the distal end of the guide wires which may cause PTFE to flake off.
- Z-1215-06 · Boston Scientific · posted Jul 12, 2006 · Terminated
Boston Scientific discovered through an internal inspection process that the PTFE (polytetrafluoroethylene) coating is not consistently adhering to the transition area of the distal end of the guide wires which may cause PTFE to flake off.
- Z-1216-06 · Boston Scientific · posted Jul 12, 2006 · Terminated
Boston Scientific discovered through an internal inspection process that the PTFE (polytetrafluoroethylene) coating is not consistently adhering to the transition area of the distal end of the guide wires which may cause PTFE to flake off.
- Z-1217-06 · Boston Scientific · posted Jul 12, 2006 · Terminated
Boston Scientific discovered through an internal inspection process that the PTFE (polytetrafluoroethylene) coating is not consistently adhering to the transition area of the distal end of the guide wires which may cause PTFE to flake off.
- Z-1218-06 · Boston Scientific · posted Jul 12, 2006 · Terminated
Boston Scientific discovered through an internal inspection process that the PTFE (polytetrafluoroethylene) coating is not consistently adhering to the transition area of the distal end of the guide wires which may cause PTFE to flake off.
- Z-1219-06 · Kensey Nash Corp · posted Jul 12, 2006 · Terminated
Volume control knob not able to go up to larger diameters
- Z-1220-06 · Kensey Nash Corp · posted Jul 12, 2006 · Terminated
volume control knob not able to go up to larger diameters
- Z-1221-06 · Kensey Nash Corp · posted Jul 12, 2006 · Terminated
volume control knob not able to go up to larger diameters
- Z-1222-06 · International Biophysics Corp · posted Jul 12, 2006 · Terminated
Tyvek pouches containing device labeled as sterile have defective seals.
- Z-1223-06 · International Biophysics Corp · posted Jul 12, 2006 · Terminated
Complaints received regarding the device's failure to zero out.
- Z-1116-06 · Disetronic Medical Systems, Inc. · posted Jul 8, 2006 · Terminated
The luer tube may break at the lock-tubing connection to the pump, causing an interruption of insulin delivery, which can and has contributed to hyperglycemia
- Z-1117-06 · Disetronic Medical Systems, Inc. · posted Jul 8, 2006 · Terminated
The luer tube may break at the lock-tubing connection to the pump, causing an interruption of insulin delivery, which can and has contributed to hyperglycemia
- Z-1118-06 · Disetronic Medical Systems, Inc. · posted Jul 8, 2006 · Terminated
The luer tube may break at the lock-tubing connection to the pump, causing an interruption of insulin delivery, which can and has contributed to hyperglycemia
- Z-1119-06 · Disetronic Medical Systems, Inc. · posted Jul 8, 2006 · Terminated
The luer tube may break at the lock-tubing connection to the pump, causing an interruption of insulin delivery, which can and has contributed to hyperglycemia
- Z-1120-06 · Disetronic Medical Systems, Inc. · posted Jul 8, 2006 · Terminated
The luer tube may break at the lock-tubing connection to the pump, causing an interruption of insulin delivery, which can and has contributed to hyperglycemia
- Z-1121-06 · Disetronic Medical Systems, Inc. · posted Jul 8, 2006 · Terminated
The luer tube may break at the lock-tubing connection to the pump, causing an interruption of insulin delivery, which can and has contributed to hyperglycemia
- Z-1122-06 · Disetronic Medical Systems, Inc. · posted Jul 8, 2006 · Terminated
The luer tube may break at the lock-tubing connection to the pump, causing an interruption of insulin delivery, which can and has contributed to hyperglycemia
- Z-1123-06 · Disetronic Medical Systems, Inc. · posted Jul 8, 2006 · Terminated
The luer tube may break at the lock-tubing connection to the pump, causing an interruption of insulin delivery, which can and has contributed to hyperglycemia
- Z-1124-06 · Disetronic Medical Systems, Inc. · posted Jul 8, 2006 · Terminated
The luer tube may break at the connection to the pump and, if this is not noticed by the patient, the patient may receive an under delivery of insulin.
- Z-1125-06 · Disetronic Medical Systems, Inc. · posted Jul 8, 2006 · Terminated
The luer tube may break at the lock-tubing connection to the pump, causing an interruption of insulin delivery, which can and has contributed to hyperglycemia
- Z-1126-06 · Disetronic Medical Systems, Inc. · posted Jul 8, 2006 · Terminated
The luer tube may break at the lock-tubing connection to the pump, causing an interruption of insulin delivery, which can and has contributed to hyperglycemia
- Z-1127-06 · Disetronic Medical Systems, Inc. · posted Jul 8, 2006 · Terminated
The luer tube may break at the lock-tubing connection to the pump, causing an interruption of insulin delivery, which can and has contributed to hyperglycemia
- Z-1128-06 · Disetronic Medical Systems, Inc. · posted Jul 8, 2006 · Terminated
The luer tube may break at the lock-tubing connection to the pump, causing an interruption of insulin delivery, which can and has contributed to hyperglycemia
- Z-1129-06 · Disetronic Medical Systems, Inc. · posted Jul 8, 2006 · Terminated
The luer tube may break at the lock-tubing connection to the pump, causing an interruption of insulin delivery, which can and has contributed to hyperglycemia
- Z-1130-06 · Disetronic Medical Systems, Inc. · posted Jul 8, 2006 · Terminated
The luer tube may break at the lock-tubing connection to the pump, causing an interruption of insulin delivery, which can and has contributed to hyperglycemia
- Z-1131-06 · Disetronic Medical Systems, Inc. · posted Jul 8, 2006 · Terminated
The luer tube may break at the lock-tubing connection to the pump, causing an interruption of insulin delivery, which can and has contributed to hyperglycemia
- Z-1132-06 · Disetronic Medical Systems, Inc. · posted Jul 8, 2006 · Terminated
The luer tube may break at the lock-tubing connection to the pump, causing an interruption of insulin delivery, which can and has contributed to hyperglycemia
- Z-1133-06 · Disetronic Medical Systems, Inc. · posted Jul 8, 2006 · Terminated
The luer tube may break at the lock-tubing connection to the pump, causing an interruption of insulin delivery, which can and has contributed to hyperglycemia
- Z-1134-06 · Disetronic Medical Systems, Inc. · posted Jul 8, 2006 · Terminated
The luer tube may break at the lock-tubing connection to the pump, causing an interruption of insulin delivery, which can and has contributed to hyperglycemia
- Z-1135-06 · Disetronic Medical Systems, Inc. · posted Jul 8, 2006 · Terminated
The luer tube may break at the lock-tubing connection to the pump, causing an interruption of insulin delivery, which can and has contributed to hyperglycemia
- Z-1136-06 · Disetronic Medical Systems, Inc. · posted Jul 8, 2006 · Terminated
The luer tube may break at the lock-tubing connection to the pump, causing an interruption of insulin delivery, which can and has contributed to hyperglycemia
- Z-1137-06 · Disetronic Medical Systems, Inc. · posted Jul 8, 2006 · Terminated
The luer tube may break at the lock-tubing connection to the pump, causing an interruption of insulin delivery, which can and has contributed to hyperglycemia
- Z-1138-06 · Disetronic Medical Systems, Inc. · posted Jul 8, 2006 · Terminated
The luer tube may break at the lock-tubing connection to the pump, causing an interruption of insulin delivery, which can and has contributed to hyperglycemia
- Z-1139-06 · Disetronic Medical Systems, Inc. · posted Jul 8, 2006 · Terminated
The luer tube may break at the lock-tubing connection to the pump, causing an interruption of insulin delivery, which can and has contributed to hyperglycemia
- Z-1140-06 · Disetronic Medical Systems, Inc. · posted Jul 8, 2006 · Terminated
The luer tube may break at the lock-tubing connection to the pump, causing an interruption of insulin delivery, which can and has contributed to hyperglycemia
- Z-1141-06 · Disetronic Medical Systems, Inc. · posted Jul 8, 2006 · Terminated
The luer tube may break at the lock-tubing connection to the pump, causing an interruption of insulin delivery, which can and has contributed to hyperglycemia
- Z-1142-06 · Disetronic Medical Systems, Inc. · posted Jul 8, 2006 · Terminated
The luer tube may break at the lock-tubing connection to the pump, causing an interruption of insulin delivery, which can and has contributed to hyperglycemia
- Z-1143-06 · Disetronic Medical Systems, Inc. · posted Jul 8, 2006 · Terminated
The luer tube may break at the lock-tubing connection to the pump, causing an interruption of insulin delivery, which can and has contributed to hyperglycemia
- Z-1144-06 · Disetronic Medical Systems, Inc. · posted Jul 8, 2006 · Terminated
The luer tube may break at the lock-tubing connection to the pump, causing an interruption of insulin delivery, which can and has contributed to hyperglycemia
- Z-1145-06 · Disetronic Medical Systems, Inc. · posted Jul 8, 2006 · Terminated
The luer tube may break at the lock-tubing connection to the pump, causing an interruption of insulin delivery, which can and has contributed to hyperglycemia
- Z-1146-06 · Disetronic Medical Systems, Inc. · posted Jul 8, 2006 · Terminated
The luer tube may break at the lock-tubing connection to the pump, causing an interruption of insulin delivery, which can and has contributed to hyperglycemia
- Z-1195-06 · MRL, Inc., A Welch Allyn Company · posted Jul 8, 2006 · Terminated
The Welch Allyn PIC 50 Defibrillators may display a 'Defib Comm' or 'Pace Comm' error message on the device display during use which may result in a terminal failure of the device to analyze the patient's ECG and deliver the appropriate therapy.
- Z-1202-06 · Spacelabs Medical Incorporated · posted Jul 6, 2006 · Terminated
Potential for patient waveforms to move to an open receiver module or to an occupied receiver causing the intermittent inappropriate monitoring of both patients for several minutes.
- Z-1205-06 · Siemens Medical Solutions USA, Inc · posted Jul 6, 2006 · Terminated
Firm became aware that a potential pinch point can exist between the stop level and the safety catch when rotating the x-ray tube assembly. Risk of injury to operators fingers.
- Z-1206-06 · Varian Medical Systems Inc · posted Jul 6, 2006 · Terminated
Machine may produce unexpected motions from the couch, collimator, and gantry rotation at maximum speeds.
- Z-1189-06 · Southmedic Inc · posted Jul 4, 2006 · Terminated
Southmedic has initiated this recall because their Japanese distributor received field reports of leaking vapofils. Leakage may cause non-toxic spills in the hospital environment, distraction of practitioners during conduct of anesthesia and surgery, and…
- Z-1190-06 · Southmedic Inc · posted Jul 4, 2006 · Terminated
Southmedic has initiated this recall because their Japanese distributor received field reports of leaking vapofils. Leakage may cause non-toxic spills in the hospital environment, distraction of practitioners during conduct of anesthesia and surgery, and…
- Z-1192-06 · Guidant Corporation · posted Jul 4, 2006 · Terminated
Guidant has received 2 reports of device malfunction associated with subpectoral implantation in an uncommon orientation (serial number facing ribs). Repetitive mechanical stress in this orientation can result in Loss of shock therapy, Loss of pacing…
- Z-1193-06 · Guidant Corporation · posted Jul 4, 2006 · Terminated
Guidant has received 2 reports of device malfunction associated with subpectoral implantation in an uncommon orientation (serial number facing ribs). Repetitive mechanical stress in this orientation can result in Loss of shock therapy, Loss of pacing…
- Z-1194-06 · Guidant Corporation · posted Jul 4, 2006 · Terminated
Guidant has received 2 reports of device malfunction associated with subpectoral implantation in an uncommon orientation (serial number facing ribs). Repetitive mechanical stress in this orientation can result in Loss of shock therapy, Loss of pacing…
- Z-1197-06 · Ossur North America Inc · posted Jul 4, 2006 · Terminated
The company's initiation of the recall is based on a finding that some units of the Total Knee device may contain faulty pins based in the axis of the knee. Small cracks propagate through the pin cross section until the pin breaks, severely compromising…
- Z-1198-06 · Ossur North America Inc · posted Jul 4, 2006 · Terminated
The company's initiation of the recall is based on a finding that some units of the Total Knee device may contain faulty pins based in the axis of the knee. Small cracks propagate through the pin cross section until the pin breaks, severely compromising…
- Z-1199-06 · Ossur North America Inc · posted Jul 4, 2006 · Terminated
The company's initiation of the recall is based on a finding that some units of the Total Knee device may contain faulty pins based in the axis of the knee. Small cracks propagate through the pin cross section until the pin breaks, severely compromising…
- Z-1200-06 · Ossur North America Inc · posted Jul 4, 2006 · Terminated
The company's initiation of the recall is based on a finding that some units of the Total Knee device may contain faulty pins based in the axis of the knee. Small cracks propagate through the pin cross section until the pin breaks, severely compromising…
- Z-1203-06 · Meridian Bioscience Inc · posted Jul 4, 2006 · Terminated
The product may produce non-specific reactions with negative specimens and/or the negative control reagent. These non-specific interactions may result in increased invalid test rates and/or positivity rates.
- Z-1204-06 · Abbott Laboratories MPG · posted Jul 4, 2006 · Terminated
Some ARCHITECT Ferritin Reagent lots may not be meeting the accuracy by correlation to AxSYM claims as they are listed in the package insert (slope of 1.0 +/- 0.2).
- Z-1188-06 · Sorenson Medical, Inc. · posted Jul 1, 2006 · Terminated
Intermittent Infusion pump may continually dispense medication under certain conditions.
- Z-1191-06 · Respironics, Inc. · posted Jul 1, 2006 · Terminated
Gas input pressures over 62 psi may cause CPAP pressure oscillation