Guidant CONTAK RENEWAL 3 (models H170, H175) CONTAK RENEWAL… recallZ-1288-06 · Boston Scientific CRM Corp · initiated May 15, 2006

Terminated All recalls posted in July 2006

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Boston Scientific CRM Corp or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Guidant has identified a pattern of premature battery depletion in devices manufactured with a single lot of capacitors from a single supplier. If function of this low voltage capacitor is compromised, the battery may deplete prematurely.

FDA root cause category: Other

What the company is doing

An Urgent Medical Device Safety Information & Corrective Action letter, dated May 12, 2006, was sent to physicians. The letter gave a description of the issue, clinical implications, rate of occurrence, devices affected and recommendation. Physicians are asked to follow-up with affected patients as soon as possible because ERI to EOL time may be shortened. Physicians are asked to contact Guidant for instructions on performing a baseline 'save to disk' and provide this information to Guidant. Guidant will be able to analyze the device memory data and estimate remaining longevity and provide individualized follow-up and replacement guidelines. Subsequent follow-up visits should continue to monitor for evidence of premature battery depletion.

Recall details

Recall number
Z-1288-06
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
May 15, 2006
Posted by the FDA
Jul 27, 2006
Recalling firm
Boston Scientific CRM Corp
Quantity
59
Product code
NIK
FDA submissions
P010012
Distribution
Nationwide including Puerto Rico. Worldwide, including Argentina, Belgium, Canada, Denmark, France, Germany, Greece, Ireland, Italy, Netherlands, New Zealand, Portugal, Spain, Syrian Arab Republic, United Arab Emirates

Product description

Guidant CONTAK RENEWAL 3 (models H170, H175) CONTAK RENEWAL 3 HE (models H177, H179) Cardiac Resynchronization Therapy Defibrillator (CRT-D). Guidant Corporation, 4100 Hamline Avenue North, St. Paul, MN 55112-5798 USA. Cardiac resynchronization therapy defibrillators (CRT-D) provide ventricular tachyarrhythmia and cardiac resynchronization therapies. Ventricular tachyarrhythmia therapy is for the treatment of ventricular tachycardia (VT) and ventricular fibrillation (VF), rhythms that are associated with sudden cardiac death (SCD). Cardiac resynchronization therapy is for the treatment of heart failure (HF) and uses biventricular electrical stimulation to synchronize ventricular contractions. Cardioversion/defibrillation therapies include a range of low-and high-energy shocks using either a biphasic or monophasic waveform.

Code information (models, lots, serial numbers)

model H170, serial numbers: 509659, 509677, 509678,  509679, 509685, 509694, 509695, 509696, 509762, 509884,  509974, 510012   model H175, serial numbers: 512665, 512666, 512667, 512668, 512708, 512709, 512710, 512711, 512713, 512719,  512721, 512722, 512723, 512759, 512791, 512793, 513062,  513064, 513073, 513085, 513086, 513227   model H177, serial numbers: 506951, 506953, 506954,  506955, 506956, 506957, 506959, 506960, 506967, 506969,  506970, 506973, 506974, 507013, 507014, 507472   model H179, serial numbers: 508312, 508316, 508322,  508325, 508326, 508336, 508342, 508349, 508404

Other devices under the same FDA submission

50 devices cleared under P010012 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 35525).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page