Guidant VITALITY 2 (models T165, T175), VITALITY DS (model… recallZ-1292-06 · Boston Scientific CRM Corp · initiated May 15, 2006

Terminated All recalls posted in July 2006

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Boston Scientific CRM Corp or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Guidant has identified a pattern of premature battery depletion in devices manufactured with a single lot of capacitors from a single supplier. If function of this low voltage capacitor is compromised, the battery may deplete prematurely.

FDA root cause category: Other

What the company is doing

An Urgent Medical Device Safety Information & Corrective Action letter, dated May 12, 2006, was sent to physicians. The letter gave a description of the issue, clinical implications, rate of occurrence, devices affected and recommendation. Physicians are asked to follow-up with affected patients as soon as possible because ERI to EOL time may be shortened. Physicians are asked to contact Guidant for instructions on performing a baseline 'save to disk' and provide this information to Guidant. Guidant will be able to analyze the device memory data and estimate remaining longevity and provide individualized follow-up and replacement guidelines. Subsequent follow-up visits should continue to monitor for evidence of premature battery depletion.

Recall details

Recall number
Z-1292-06
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
May 15, 2006
Posted by the FDA
Jul 27, 2006
Recalling firm
Boston Scientific CRM Corp
Quantity
672
Product code
LWP
FDA submissions
P960040
Distribution
Nationwide including Puerto Rico. Worldwide, including Argentina, Belgium, Canada, Denmark, France, Germany, Greece, Ireland, Italy, Netherlands, New Zealand, Portugal, Spain, Syrian Arab Republic, United Arab Emirates

Product description

Guidant VITALITY 2 (models T165, T175), VITALITY DS (model T125), VITALITY EL (model T135), Implantable Cardioverter Defibrillators (ICD), are designed to detect and terminate ventricular tachycardia (VT) and ventricular fibrillation (VF) and provide bradycardia therapy (atrial and ventricular pacing). Therapies include both low- and high-energy shocks using either a biphasic or monophasic waveform. Vitality 2 devices also offer a wide variety of antitachycardia pacing schemes to terminate slower, more stable ventricular tachyarrhythmias. Bradycardia pacing, including adaptive-rate features, is available to detect and treat bradyarrhythmias and to support the cardiac rhythm after defibrillation therapy. Devices denoted with DR offer dual-chamber bradycardia features (atrial and/or ventricular pacing and sensing), and the devices denoted with VR offer single-chamber bradycardia features (ventricular pacing and sensing).

Code information (models, lots, serial numbers)

Show the full code information (8k characters)

model T165, serial numbers: 106641, 106643, 106644, 106645, 106646, 106647, 106648, 106649, 106651, 106652,  106676, 106677, 106678, 106679, 106680, 106681, 106683,  106685, 106686, 106687, 106691, 106692, 106693, 106694,  106696, 106697, 106700, 106701, 106702, 106778, 106779,  106780, 106781, 106782, 106788, 106789, 106790, 106792,  106793, 106794, 106795, 106796, 106797, 106798, 106799,  106803, 106804, 106805, 106806, 106807, 106808, 106810,  106813, 106815, 106816, 106817, 106818, 106819, 106820,  106821, 106822, 106823, 106824, 106825, 106826, 106827,  106828, 106829, 106830, 106831, 106832, 106833, 106834,  106835, 106836, 106837, 106838, 106839, 106840, 106841,  106842, 106843, 106844, 106845, 106846, 106847, 106849,  106850, 106851, 106852, 106853, 106854, 106855, 106856,  106857, 106858, 106860, 106861, 106862, 106863, 106864, 106865, 106867, 106868, 106869, 106870, 106871, 106872, 106873, 106874, 106875, 106876, 106877, 106878, 106879, 106880, 106881, 106882, 106883, 106884, 106885, 106886,  106888, 106889, 106890, 106891, 106892, 106894, 106895,  106896, 106898, 106899, 106900, 106901, 106903, 106905,  106906, 106907, 106908, 106909, 106910, 106911, 106914,  106915, 106916, 106917, 106919, 106920, 106922, 106923,  106925, 106926, 106929, 106932, 106935, 106936, 106937,  106939, 106940, 106941, 106942, 106943, 106945, 106946,  106947, 106948, 106949, 106952, 106953, 106954, 106955,  106956, 106957, 106959, 106960, 106962, 106963, 106964,  106965, 106968, 106969, 106970, 106971, 106972, 106973,  106974, 106975, 106976, 106977, 106978, 106979, 106980,  106981, 106982, 106983, 106984, 106985, 106987, 106990,  106991, 106992, 106993, 106996, 106997, 106998, 106999,  107000, 107001, 107002, 107003, 107004, 107007, 107008,  107009, 107010, 107011, 107012, 107013, 107015, 107017,  107018, 107019, 107020, 107021, 107033, 107034, 107035, 107036, 107037, 107039, 107043, 107044, 107045, 107046,  107047, 107057, 107058, 107059, 107060, 107061, 107062,  107064, 107065, 107066, 107067, 107068, 107069, 107070,  107072, 107075, 107077, 107078, 107079, 107082, 107087,  107091, 107092, 107099, 107101, 107102, 107103, 107106,  107110, 107116, 107123, 107124, 107141, 107208, 107210,  107254, 107263, 107264, 107265, 107266, 107267, 107298,  107299, 107300, 107301, 107303, 107304, 107306, 107307, 107308, 107315, 107316, 107317, 107318, 107320, 107321,  107322, 107389, 107718, 108310, 108376, 115934   model T175, serial numbers: 105118, 105119, 105120,  105122, 105126, 105130, 105131, 105140, 105141, 105156,  105157, 105158, 105161, 105162, 105165, 105166, 105167,  105169 105170, 105172, 105173, 105174, 105176, 105177,  105178, 105179, 105180, 105181, 105182, 105183, 105184,  105186, 105187, 105189, 105191, 105192, 105194, 105196,  105197, 105200, 105207, 105208, 105209, 105210, 105211,  105212, 105213, 105217, 105304, 105305, 105306, 105307,  105308, 105364, 105365, 105366, 105367, 105368, 105369,  105370, 105371, 105372, 105373, 105374, 105375, 105376,  105377, 105378, 105379, 105380, 105381, 105382, 105383,  105384, 105385, 105386, 105387, 105388, 105389, 105390,  105391, 105393, 105395, 105397, 105398, 105399, 105400,  105401, 105402, 105403, 105404, 105405, 105406, 105407,  105409, 105410, 105411, 105412, 105413, 105414, 105415,  105416, 105417, 105418, 105419, 105420, 105421, 105422,  105423, 105424, 105425, 105426, 105427, 105428, 105429,  105431, 105432, 105433, 105434 ,105435, 105436, 105437,  105438, 105439, 105440, 105442, 105443, 105444, 105445,  105446, 105447, 105448, 105449, 105450, 105451, 105452,  105453, 105454, 105455, 105456, 105458, 105459, 105460, - 105461, 105462, 105463, 105464, 105465, 105466, 105467,  105468, 105469, 105470, 105471, 105472, 105473, 105474,  105476, 105477, 105478, 105479, 105480, 105481, 105482,  105483, 105484, 105485, 105487, 105488, 105573, 105575, 105576, 105577, 105578, 105579, 105580, 105581, 105584,  105586, 105587, 105588, 105589, 105590, 105591, 105592,  105593, 105594, 105595, 105596, 105598, 105599, 105600,  105602, 105603, 105604, 105605, 105606, 105607, 105611,  105613, 105624, 105625, 105627, 105628, 105629, 105630,  105632, 105633, 105634, 105635, 105636, 105638, 105639,  105640, 105642, 105643, 105644, 105645, 105646, 105647,  105648, 105649, 105650, 105651, 105652, 105654, 105655,  105656, 105657, 105658, 105659, 105660, 105661, 105662,  105663, 105664, 105665, 105666, 105667, 105668, 105669,  105670, 105671, 105672, 105673, 105674, 105675, 105676,  105677, 105678, 105679, 105680, 105681, 105682, 105683,  105687, 105688, 105689, 105690, 105691, 105694, 105705,  105706, 105707, 105712, 105713, 105714, 105715, 105716,  105717, 105718, 105720, 105721, 105722, 105723, 105724,  105725, 105726, 105727, 105728, 105739, 105740, 105741, - 105743, 105744, 105745, 105746, 105747, 105748, 105749,  105750, 105752, 105753, 105754, 105755, 105756, 105757,  105758, 105759, 105760, 105762, 105763, 105764, 105765,  105766, 105767, 105768, 105785, 105786, 105790, 105792,  105796, 105820, 105857, 105881, 105884, 105885, 105895,  105898, 105914, 105915, 105917, 105919, 105923, 105924,  105925, 105926, 105927, 105928, 105929, 105930, 105931,  105932 105933, 105934, 105935, 105936, 105937, 105940,  105944, 105945, 105946, 105947, 105948, 105949, 105950, 105951, 105952, 105953, 106074, 106155, 106156, 106157,  106171, 106172, 106173, 106174, 106177, 106178, 106186,  106187, 106189, 106191, 106199, 106200, 106201, 106212,  106213, 106235, 106282, 106361, 106503, 107047,107301,  107593, 108285   model T125, serial numbers: 113211, 113221, 113222,  113223, 113224, 113241, 113243, 113244, 113272, 113273,  113274, 113275, 113276, 113277, 113278, 113279, 113282, 113283, 113284, 113285, 113286, 113287, 113288, 113291,  113292, 113294, 113295, 113296, 113297, 113298, 113299,  113300, 113301, 113317, 113318, 113319, 113320, 113322,  113323, 113324, 113325, 113326, 113327, 113328, 113329,  113330, 113331, 113333, 113334, 113335, 113336, 113337,  113339, 113340, 113341, 113342, 113343, 113344, 113345,  113346, 113347, 113348, 113349, 113350, 113351, 113352,  113353, 113355, 113356, 113359, 113360, 113362, 113363,  113364, 113365, 113366, 113367, 113368, 113369, 113370,  113372, 113373, 113374, 113375, 113378, 113379, 113380,  113381, 113383, 113384, 113385, 113386, 113387, 113388,  113389, 113390, 113391, 113393, 113394, 113396, 113397,  113398, 113399, 113400, 113403, 113408, 113414, 113417,  113420, 113422, 113425, 113427, 113428, 113429, 113430,  113431, 113433, 113434, 113435, 113487, 113492, 113495,  113496, 113500, 113557, 113643, 113847, 113850, 114102    model T135, serial numbers: 941570, 941572, 941573, 941574, 941578, 941585, 941589, 941590, 941591, 941592,  941593, 941595, 941597, 941601, 941602, 941603, 941604,  941605, 941606, 941607, 941608, 941609, 941610, 941613,  941614, 941615, 941616, 941618, 941619, 941620, 941621, 941622, 941623, 941624, 941625, 941626, 941627, 941628,  941629, 941630, 941632, 941633, 941634, 941652, 941653,  941654, 941680, 941681, 941682, 941683, 941688, 941689,  941690, 941691 941692, 941695, 941696, 941697, 941698,  941699, 941700, 941701, 941702, 941703, 941704, 941709,  941710, 941712, 941715, 941716, 941717, 941718, 941719, 941721, 941722, 941723, 941724, 941725, 941727, 941728,  941732, 941734, 941735, 941736, 941737, 941738, 941739,  941740, 941741, 941742, 941743, 941745, 941747, 941748,  941749, 941750, 941752, 941753, 941754, 941756, 941757,  941759, 941762, 941767, 941775, 941776, 941777, 941778,  941781, 941784, 941786, 941787, 941788, 941789, 941800,  941802, 941803, 941804, 941805, 941807, 941810, 941812,  941813, 941814, 941819, 941829, 941830, 941831, 941845,  941846, 941929, 941930, 941933, 941940, 941948, 941953,  941956, 941962, 941970, 941975, 941978, 941979, 942017

Other devices under the same FDA submission

45 devices cleared under P960040 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 35525).