Defibtech Sentry Semi-Automatic External Defibrilator (AED)… recallZ-0580-2007 · Defibtech, LLC · initiated Feb 17, 2007
Terminated All recalls posted in March 2007
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Defibtech, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Device Malfunction-The self-test software may allow a self-test to clear a previously detected low battery condition. If this occurs, the operator may be unaware of the low battery and the device may be "unable" to deliver a defibrillation shock, which could result in failure to resuscitate a patient.
FDA root cause category: Software design
What the company is doing
Defibtech LLC initiated notification to customers/distributors on 2/17/07 by email. Hard copy letters followed on 2/22/07 with acknowledgement card. Users provided with instructions for 'Immediate Recommendations' until the unit''s software is upgraded. Distributors, their customers, and direct accounts instructed to locate affected units and upgrade these units in the field. Distributors, their customers, and direct accounts shall return to Defibtech documentation confirming unit software upgrade was completed..Press was issued by the firm on March 6, 2007
Recall details
- Recall number
- Z-0580-2007
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Feb 17, 2007
- Posted by the FDA
- Mar 8, 2007
- Recalling firm
- Defibtech, LLC
- Quantity
- 22,197 units
- Product code
- MKJ
- FDA submissions
- K013896
- Distribution
- Worldwide distribution ---- USA and countries of Canada, Uruguay, Jordon, Turkey United Kingdom, Saudi Arabia, Sweden France, South Africa, Cyprus, Spain, Germany India, Japan, Israel, Australia, Norway, Philppines Singapore, Poland, Germany, Italy, Switzerland Netherlands, Indonesia, United Arab Emirates Malaysia, Austria, Ireland, Kingdom of Bahrain, and Korea.
Product description
Defibtech Sentry Semi-Automatic External Defibrilator (AED) sold as 'Lifeline AED' brand -Model DDU-100A (version 2.002 and earlier)
Code information (models, lots, serial numbers)
Model DDU 100A-Software versions earlier than version 2.002
Other devices under the same FDA submission
10 devices cleared under K013896 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Lifeline Configuration DDU-100ADefibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)1 death1,285 reports
- Lifeline DDU-100ADefibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)1 recall3 deaths640 reports
- Lifeline Configuration DDU-100EDefibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)1 recall469 reports
- Lifeline DDU-100EDefibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)1 recall177 reports
- ReviveR Configuration DDU-100BDefibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)185 reports
- ReviveR DDU-100BDefibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)71 reports
- DDU-1XX Series Adult Rescue Pads DDP-100Defibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)1 recall1 report
- DDU-1XX Series Battery Pack DBP-1400Defibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)3 reports
- Lifeforce Configuration DDU-100CDefibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)2 reports
- DDU-1XX series Battery Pack Configuration DBP-1400Defibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)2 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page