Medical device recalls: May 2008
The FDA posted 78 medical device recall records in May 2008, from 31 recall events by 30 companies. 78 have not been classified yet.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Abbott Laboratories · 2 events, 2 records
- Ev3, Inc · 1 event, 12 records
- Covidien LP · 1 event, 9 records
- Beckman Coulter Inc · 1 event, 8 records
- Pointe Scientific, Inc. · 1 event, 8 records
- Boston Scientific Corporation · 1 event, 6 records
- Viasys Respiratory Care, Inc.dba Bird Products · 1 event, 5 records
- Medefil Incorporated · 1 event, 3 records
FDA root cause
- Software design · 13
- Material/Component Contamination · 13
- Labeling Change Control · 13
- Process control · 11
- Labeling mix-ups · 10
- Component design/selection · 6
- Device Design · 4
- Nonconforming Material/Component · 3
Not yet classified (78)
The FDA has posted these records but has not published a class for them yet.
- Z-1137-2008 · Philips Medical Systems North America Co. Phillips · posted May 28, 2008 · Terminated
Delay in Imaging: MultiDiagnost Eleva with Flat Detector x-ray system cannot perform fluoroscopy procedures while certain image file housekeeping operations are performed. Fluoroscopy cannot be performed and new exposures are not able to be acquired until…
- Z-1546-2008 · Lorad, A Hologic, Inc. · posted May 28, 2008 · Terminated
Marker Misalignment: The new software version contained a magnification factor that shows CAD markers misaligned with the identified indications. The marked area is misaligned up to 2cm from the indication.
- Z-1644-2008 · Emageon · posted May 28, 2008 · Terminated
Miscalculation: The calculation of the standard uptake value does not include the required calibration factor/decay correction factor. The resulting value is incorrectly calculated, which may lead to incorrect interpretation of tissue density. The…
- Z-1094-2008 · Beckman Coulter Inc · posted May 24, 2008 · Terminated
Intermittent Failure of Stirrer Motor-Beckman Coulter has determined that incorrect Glucose results on LX20 or DxC instruments have been related to an intermittent failure of the stirrer motor. -Stirrer motors can stall without any flags or motion errors.…
- Z-1095-2008 · Beckman Coulter Inc · posted May 24, 2008 · Terminated
Intermittent Failure of Stirrer Motor-Beckman Coulter has determined that incorrect Glucose results on LX20 or DxC instruments have been related to an intermittent failure of the stirrer motor. -Stirrer motors can stall without any flags or motion errors.…
- Z-1096-2008 · Beckman Coulter Inc · posted May 24, 2008 · Terminated
Intermittent Failure of Stirrer Motor-Beckman Coulter has determined that incorrect Glucose results on LX20 or DxC instruments have been related to an intermittent failure of the stirrer motor. -Stirrer motors can stall without any flags or motion errors.…
- Z-1097-2008 · Beckman Coulter Inc · posted May 24, 2008 · Terminated
Intermittent Failure of Stirrer Motor-Beckman Coulter has determined that incorrect Glucose results on LX20 or DxC instruments have been related to an intermittent failure of the stirrer motor. -Stirrer motors can stall without any flags or motion errors.…
- Z-1098-2008 · Beckman Coulter Inc · posted May 24, 2008 · Terminated
Intermittent Failure of Stirrer Motor-Beckman Coulter has determined that incorrect Glucose results on LX20 or DxC instruments have been related to an intermittent failure of the stirrer motor. -Stirrer motors can stall without any flags or motion errors.…
- Z-1099-2008 · Beckman Coulter Inc · posted May 24, 2008 · Terminated
Intermittent Failure of Stirrer Motor-Beckman Coulter has determined that incorrect Glucose results on LX20 or DxC instruments have been related to an intermittent failure of the stirrer motor. -Stirrer motors can stall without any flags or motion errors.…
- Z-1100-2008 · Beckman Coulter Inc · posted May 24, 2008 · Terminated
Intermittent Failure of Stirrer Motor-Beckman Coulter has determined that incorrect Glucose results on LX20 or DxC instruments have been related to an intermittent failure of the stirrer motor. -Stirrer motors can stall without any flags or motion errors.…
- Z-1101-2008 · Beckman Coulter Inc · posted May 24, 2008 · Terminated
Intermittent Failure of Stirrer Motor-Beckman Coulter has determined that incorrect Glucose results on LX20 or DxC instruments have been related to an intermittent failure of the stirrer motor. -Stirrer motors can stall without any flags or motion errors.…
- Z-1166-2008 · Abbott Laboratories, Inc · posted May 24, 2008 · Terminated
Quality Control (Out of Range)- ICT Calibrator lot number: 0505017, may contain an interferent. Not all vials within a kit are affected. The investigation determined that the interferent affects Potassium (K+) only. Use of the affected calibrator vials…
- Z-1110-2008 · Pointe Scientific, Inc. · posted May 22, 2008 · Terminated
Failure of the reagent to produce test results. The R1 reagent may be contaminated with microorganisms.
- Z-1111-2008 · Pointe Scientific, Inc. · posted May 22, 2008 · Terminated
Failure of the reagent to produce test results. The R1 reagent may be contaminated with microorganisms.
- Z-1112-2008 · Pointe Scientific, Inc. · posted May 22, 2008 · Terminated
Failure of the reagent to produce test results. The R1 reagent may be contaminated with microorganisms.
- Z-1113-2008 · Pointe Scientific, Inc. · posted May 22, 2008 · Terminated
Failure of the reagent to produce test results. The R1 reagent may be contaminated with microorganisms.
- Z-1114-2008 · Pointe Scientific, Inc. · posted May 22, 2008 · Terminated
Failure of the reagent to produce test results. The R1 reagent may be contaminated with microorganisms.
- Z-1115-2008 · Pointe Scientific, Inc. · posted May 22, 2008 · Terminated
Failure of the reagent to produce test results. The R1 reagent may be contaminated with microorganisms.
- Z-1116-2008 · Pointe Scientific, Inc. · posted May 22, 2008 · Terminated
Failure of the reagent to produce test results. The R1 reagent may be contaminated with microorganisms.
- Z-1117-2008 · Pointe Scientific, Inc. · posted May 22, 2008 · Terminated
Failure of the reagent to produce test results. The R1 reagent may be contaminated with microorganisms.
- Z-1132-2008 · Keeler Instruments Inc · posted May 22, 2008 · Terminated
Good Manufacturing Practices (GMP) deficiencies may compromise the safety and effectiveness of the device. Firm is on import alert 89-04 and released product.
- Z-1608-2008 · Civco Medical Instruments Inc · posted May 21, 2008 · Terminated
Mislabeled: The mask is mislabeled with the incorrect part number but the outer packaging is correct.
- Z-1613-2008 · Med Tec Inc dba CIVCO Medical Solutions · posted May 21, 2008 · Terminated
Mislabeled: Fiducial markers were incorrectly labeled. They are marked (2mm spheres) Bone gold markers, but contain (0.9x3mm cylindrical) Soft tissue gold markers.
- Z-1614-2008 · Med Tec Inc dba CIVCO Medical Solutions · posted May 21, 2008 · Terminated
Mislabeled: Fiducial markers were incorrectly labeled. They are marked (0.9x3mm cylindrical) Soft tissue gold markers, but contain (2mm spheres) Bone gold markers.
- Z-1616-2008 · Cardiac Science Corporation · posted May 21, 2008 · Terminated
Adaptor Cable Incorrectly Manufactured: The red and white connector ends of the adapter are switched. Therefore, the cable will not connect to the electrodes as indicated in the instructions.
- Z-1138-2008 · Siemens Medical Solutions USA, Inc. · posted May 20, 2008 · Terminated
Inaccurate results: A software bug may cause inaccurate wall motion abnormality scoring results to be displayed.
- Z-0982-2008 · Viasys Respiratory Care, Inc.dba Bird Products · posted May 14, 2008 · Terminated
Alarm Failure -- There is a potential of alarms not being activated subsequent to specific electronic faults that may cause INOP condition.
- Z-0983-2008 · Viasys Respiratory Care, Inc.dba Bird Products · posted May 14, 2008 · Terminated
Alarm Failure -- There is a potential of alarms not being activated subsequent to specific electronic faults that may cause INOP condition.
- Z-0984-2008 · Viasys Respiratory Care, Inc.dba Bird Products · posted May 14, 2008 · Terminated
Alarm Failure -- There is a potential of alarms not being activated subsequent to specific electronic faults that may cause INOP condition.
- Z-0985-2008 · Viasys Respiratory Care, Inc.dba Bird Products · posted May 14, 2008 · Terminated
Alarm Failure -- There is a potential of alarms not being activated subsequent to specific electronic faults that may cause INOP condition.
- Z-0986-2008 · Viasys Respiratory Care, Inc.dba Bird Products · posted May 14, 2008 · Terminated
Alarm Failure -- There is a potential of alarms not being activated subsequent to specific electronic faults that may cause INOP condition.
- Z-0987-2008 · Cardinal Health · posted May 14, 2008 · Terminated
Oxygen fluctuations - The AirLife Infant Nasal CPAP System exhibited O2 fluctuations when used with low O2 flow rates when the PTO/Auxillary port on the driver is used as a blender in administering O2 therapy via nasal cannula at flow rates of less than…
- Z-1103-2008 · Abbott Laboratories · posted May 14, 2008 · Terminated
Microbial Contamination: The WBC (White Blood Cell) Reagent Part A was confirmed positive for Pseudomonas contamination.
- Z-1105-2008 · Stryker Howmedica Osteonics Corp. · posted May 14, 2008 · Terminated
Insertion/Removal Difficulties: The Baseplate Impactor/Extractor may not assemble/disassemble easily to the baseplate.
- Z-1461-2008 · Alcon Research, Ltd · posted May 14, 2008 · Terminated
Device corrosion: Forceps are prone to corrosion and possibly premature fracture and/or malfunction.
- Z-1462-2008 · Abbott Laboratories · posted May 14, 2008 · Terminated
Error Message: If the Positive and Negative Interpretation Cutoff parameters (116/117) are edited after installation of the revised assay file, VRTX error #0002 in Task 40 is generated and the instrument locks up when it attempts to report an AxSYM…
- Z-1624-2008 · Remel Inc · posted May 14, 2008 · Terminated
Failure to produce reaction: Use of the product with various RapID Identification systems panels may exhibit poor or no reactions.
- Z-1625-2008 · Remel Inc · posted May 14, 2008 · Terminated
Failure to produce reaction: Use of the product with various RapID Identification systems panels may exhibit poor or no reactions.
- Z-1543-2008 · Medefil Incorporated · posted May 9, 2008 · Terminated
The heparin lock flush solution was manufactured from contaminated Heparin Sodium USP Active Pharmaceutical Ingredient (API). The FDA has received reports of serious injuries and/or deaths in patients administered finished heparin injectable products…
- Z-1544-2008 · Medefil Incorporated · posted May 9, 2008 · Terminated
The heparin lock flush solution was manufactured from contaminated Heparin Sodium USP Active Pharmaceutical Ingredient (API). The FDA has received reports of serious injuries and/or deaths in patients administered finished heparin injectable products…
- Z-1545-2008 · Medefil Incorporated · posted May 9, 2008 · Terminated
The heparin lock flush solution was manufactured from contaminated Heparin Sodium USP Active Pharmaceutical Ingredient (API). The FDA has received reports of serious injuries and/or deaths in patients administered finished heparin injectable products…
- Z-1594-2008 · Covidien LP · posted May 9, 2008 · Terminated
Scientific Protein Laboratories (SPL), disclosed that two lots of Heparin Sodium, USP Active Pharmaceutical Ingredient has a heparin-like contaminant.
- Z-1595-2008 · Covidien LP · posted May 9, 2008 · Terminated
Scientific Protein Laboratories (SPL), disclosed that two lots of Heparin Sodium, USP Active Pharmaceutical Ingredient has a heparin-like contaminant
- Z-1596-2008 · Covidien LP · posted May 9, 2008 · Terminated
Scientific Protein Laboratories (SPL), disclosed that two lots of Heparin Sodium, USP Active Pharmaceutical Ingredient has a heparin-like contaminant
- Z-1597-2008 · Covidien LP · posted May 9, 2008 · Terminated
Scientific Protein Laboratories (SPL), disclosed that two lots of Heparin Sodium, USP Active Pharmaceutical Ingredient has a heparin-like contaminant Tyco Healthcare, Mansfield, MA 02048
- Z-1598-2008 · Covidien LP · posted May 9, 2008 · Terminated
Scientific Protein Laboratories (SPL), disclosed that two lots of Heparin Sodium, USP Active Pharmaceutical Ingredient has a heparin-like contaminant
- Z-1599-2008 · Covidien LP · posted May 9, 2008 · Terminated
Scientific Protein Laboratories (SPL), disclosed that two lots of Heparin Sodium, USP Active Pharmaceutical Ingredient has a heparin-like contaminant
- Z-1600-2008 · Covidien LP · posted May 9, 2008 · Terminated
Scientific Protein Laboratories (SPL), disclosed that two lots of Heparin Sodium, USP Active Pharmaceutical Ingredient has a heparin-like contaminant
- Z-1601-2008 · Covidien LP · posted May 9, 2008 · Terminated
Scientific Protein Laboratories (SPL), disclosed that two lots of Heparin Sodium, USP Active Pharmaceutical Ingredient has a heparin-like contaminant
- Z-1602-2008 · Covidien LP · posted May 9, 2008 · Terminated
Scientific Protein Laboratories (SPL), disclosed that two lots of Heparin Sodium, USP Active Pharmaceutical Ingredient has a heparin-like contaminant
- Z-1141-2008 · Philips Healthcare Informatics, Inc. · posted May 8, 2008 · Terminated
Incorrect display of one of the patient's images. This malfunction could cause an image to be incorrectly displayed, for example flipped left to right and anterior to posterior. The original image is still presented in the correct orientation next to the…
- Z-0517-2008 · Medtronic Neuromodulation · posted May 7, 2008 · Terminated
Unable to detach from source: The capsule may not detach from the delivery system following attachment to the esophageal wall.
- Z-1035-2008 · B. Braun Medical, Inc. · posted May 7, 2008 · Terminated
Faulty tubing does not prime machine as intended.
- Z-1036-2008 · B. Braun Medical, Inc. · posted May 7, 2008 · Terminated
Faulty tubing does not prime machine as intended.
- Z-1179-2008 · ConvaTec · posted May 7, 2008 · Terminated
Mislabeled: The market unit carton label reads 5/8 - 1 inch, and it should read 5/16 - 1 inch. The metric units (8-25mm) on the label are correct.
- Z-1240-2008 · Diasorin Inc. · posted May 7, 2008 · Terminated
Inaccurate measurements: Kits may sporadically recover kit and external control values out of range high.
- Z-1300-2008 · Boston Scientific Corporation · posted May 7, 2008 · Terminated
Mislabeled: Seven lots/ batches of Boston Scientific Mach 1 Guide Catheter may be mislabeled with the incorrect device length.
- Z-1301-2008 · Boston Scientific Corporation · posted May 7, 2008 · Terminated
Mislabeled: Seven lots/ batches of Boston Scientific Mach 1 Guide Catheter may be mislabeled with the incorrect device length.
- Z-1302-2008 · Boston Scientific Corporation · posted May 7, 2008 · Terminated
Mislabeled: Seven lots/ batches of Boston Scientific Mach 1 Guide Catheter may be mislabeled with the incorrect device length.
- Z-1303-2008 · Boston Scientific Corporation · posted May 7, 2008 · Terminated
Mislabeled: Seven lots/ batches of Boston Scientific Mach 1 Guide Catheter may be mislabeled with the incorrect device length.
- Z-1304-2008 · Boston Scientific Corporation · posted May 7, 2008 · Terminated
Mislabeled: Seven lots/ batches of Boston Scientific Mach 1 Guide Catheter may be mislabeled with the incorrect device length.
- Z-1305-2008 · Boston Scientific Corporation · posted May 7, 2008 · Terminated
Mislabeled: Seven lots/ batches of Boston Scientific Mach 1 Guide Catheter may be mislabeled with the incorrect device length.
- Z-1118-2008 · Innovacon Inc · posted May 6, 2008 · Terminated
Incorrect results: The Sure-Vue Serum/Urine hCG-STAT test, when interpreted at extended read times, may exhibit sensitivity to patient samples containing hCG at levels well below the cut-off, potentially resulting in a false positive interpretation by the…
- Z-1119-2008 · Siemens Medical Solutions Diagnostics · posted May 6, 2008 · Terminated
Incorrect results: False reactive specimens on the ADVIA Centaur HAV IgM assay.
- Z-1120-2008 · Ev3, Inc · posted May 6, 2008 · Terminated
Incorrect information on label: Specific lots of ev3 IntraStent Unmounted Balloon Expandable Stents have the two symbols for length and diameter on the side and end flaps reversed.
- Z-1121-2008 · Ev3, Inc · posted May 6, 2008 · Terminated
Incorrect information on label: Specific lots of ev3 IntraStent Unmounted Balloon Expandable Stents have the two symbols for length and diameter on the side and end flaps reversed.
- Z-1122-2008 · Ev3, Inc · posted May 6, 2008 · Terminated
Incorrect information on label: Specific lots of ev3 IntraStent Unmounted Balloon Expandable Stents have the two symbols for length and diameter on the side and end flaps reversed.
- Z-1123-2008 · Ev3, Inc · posted May 6, 2008 · Terminated
Incorrect information on label: Specific lots of ev3 IntraStent Unmounted Balloon Expandable Stents have the two symbols for length and diameter on the side and end flaps reversed.
- Z-1124-2008 · Ev3, Inc · posted May 6, 2008 · Terminated
Incorrect information on label: Specific lots of ev3 IntraStent Unmounted Balloon Expandable Stents have the two symbols for length and diameter on the side and end flaps reversed.
- Z-1125-2008 · Ev3, Inc · posted May 6, 2008 · Terminated
Incorrect information on label: Specific lots of ev3 IntraStent Unmounted Balloon Expandable Stents have the two symbols for length and diameter on the side and end flaps reversed.
- Z-1126-2008 · Ev3, Inc · posted May 6, 2008 · Terminated
Incorrect information on label: Specific lots of ev3 IntraStent Unmounted Balloon Expandable Stents have the two symbols for length and diameter on the side and end flaps reversed.
- Z-1127-2008 · Ev3, Inc · posted May 6, 2008 · Terminated
Incorrect information on label: Specific lots of ev3 IntraStent Unmounted Balloon Expandable Stents have the two symbols for length and diameter on the side and end flaps reversed.
- Z-1128-2008 · Ev3, Inc · posted May 6, 2008 · Terminated
Incorrect information on label: Specific lots of ev3 IntraStent Unmounted Balloon Expandable Stents have the two symbols for length and diameter on the side and end flaps reversed.
- Z-1129-2008 · Ev3, Inc · posted May 6, 2008 · Terminated
Incorrect information on label: Specific lots of ev3 IntraStent Unmounted Balloon Expandable Stents have the two symbols for length and diameter on the side and end flaps reversed.
- Z-1130-2008 · Ev3, Inc · posted May 6, 2008 · Terminated
Incorrect information on label: Specific lots of ev3 IntraStent Unmounted Balloon Expandable Stents have the two symbols for length and diameter on the side and end flaps reversed.
- Z-1131-2008 · Ev3, Inc · posted May 6, 2008 · Terminated
Incorrect information on label: Specific lots of ev3 IntraStent Unmounted Balloon Expandable Stents have the two symbols for length and diameter on the side and end flaps reversed.
- Z-1104-2008 · GE Healthcare Integrated IT Solutions · posted May 1, 2008 · Terminated
Report missing characters: Four special characters, when entered into the exam notes, are not transferred to the preview panel or hard copy printout (greater than [>], less than [<], quotation mark ["], and apostrophe [']). This may not be evident to the…
- Z-1106-2008 · Stryker Instruments Div. of Stryker Corporation · posted May 1, 2008 · Terminated
Sterility may be compromised, as the packaging may have channels in the packaging seal.